Peter Huber Kältemaschinenbau CC-225B
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-225B

Peter Huber Kältemaschinenbau CC-225B — Thermostatic baths.

The CC-225B is a 23.5-litre stainless-steel bath with an integrated variable-speed pressure/suction pump and Pilot ONE controller, rated for 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. It is the largest bath in the CC-20xB series, distinguished by its 17-litre minimum fill and 27 l/min pressure-pump delivery, and serves as a temperature-control source for external circuits and immersed samples.

Direct drop-in replacements

Eppendorf
MixMate ®
58%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Mit Edelstahlbaedern Bis 200C Series — Model Comparison & Selection

Model ControllerInterfacePump connectionTemperature range, °CWeight, kgMax delivery flow rate, liter/minute
CC-225B Current Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001627
CC-202C Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200922
CC-205B Pilot ONEEthernet, USB (Host u. DM16x1 AG male45–200927
CC-208B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–2001027
CC-212B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001122
CC-215B Pilot ONEEthernet, USB (Host u. DM16x1 AG male25–20012—
CC-220B Pilot ONEEthernet, USB (Host u. DM16x1 AG male (Option, #25–2001427
KISS 202C KISSUSB (Device), RS232 InteM16x1 AG male45–200814
KISS 205B Badumwälzthermostat mit Edelstahlbad KISSUSB (Device), RS232 InteM16x1 AG male45–200914
KISS 208B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001014
KISS 212B KISSUSB (Device), RS232 InteM16x1 AG male25–2001114
KISS 215B KISSUSB (Device), RS232 InteM16x1 AG male25–2001214
KISS 220B KISSUSB (Device), RS232 InteM16x1 AG male (Option, #25–2001414
KISS 225B KISSUSB (Device), RS232 InteM16x1 AG male25–2001614

Full specifications

Core Performance & Functional Specifications
Filling volume 17 l (min)
Maximum delivery flow rate 27 l/min (pressure pump; also 0, 7 bar)
suction pump; also 0, 4 bar: 22 l/min
Pump connection M16x1 AG male (Option, #19607) (Option, #19607)
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range 25...200 °C
with cooling unit: -30...200 °C · with water cooling: 20...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 350 x 555 x 425 mm
Heating capacity mfr 1, 5 - 2, 1 kW
Bath opening WxD/BD mfr 290 x 329 / 200
Bath volume mfr 23, 5 l
Dimensions, Weight & Mechanical Design
Weight 16 kg
Electrical, Power & Drive Specifications
Current consumption 10 A (max)
Power supply 200-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-225B is an immersion circulator/heating bath built around a 23.5-litre stainless-steel bath (bath opening 290 × 329 mm, depth 200 mm) with a bath bridge that includes a hole for an immersion cooling probe (TC45–TC100E). Within the CC-20xB series it is the largest-capacity variant; the CC-202C, CC-205B, CC-208B and CC-212B are smaller siblings. The unit weighs 16 kg and measures 350 × 555 × 425 mm (W × D × H). Heating capacity is 1.5–2.1 kW, and temperature stability is ±0.02 K. The controller is the Pilot ONE, with Ethernet, USB (host and device) and RS232 interfaces and a Pt100 external sensor connection.

Thermal behavior and media

Achievable temperature depends on the cooling method: 25–200 °C without cooling, 20–200 °C with water cooling, and -30–200 °C with a refrigerator. The variable-speed pressure and suction pump delivers up to 27 l/min at 0.7 bar (pressure side) and 22 l/min at 0.4 bar (suction side), bounding the external circuit resistance that can be served. Minimum filling is 17 l of the 23.5-litre bath volume. When a Huber pump adapter is used, polyglycol must not be employed as the heat transfer fluid. Adjustable overtemperature protection according to DIN 12876 is provided and must be set for the application.

Installation, electrical and control

Supply is 200–240 V, 1~ or 2~, 50/60 Hz, with maximum current 10 A; branch protection must be sized between 10 A and 16 A, and external branch circuit protection according to UL 489 is required. Voltage and frequency tolerances of ±10% are acceptable only if they do not act in opposite directions. Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering. The enclosure is IP20, so the unit must be kept dry and not exposed to water; ambient temperature must remain between 5 °C and 40 °C. The equipment is classified Class A, Group 1 to EN55011 and may cause interference in residential environments.

Integration & Operating Notes

Electrical
Branch protection must be 10–16 A with UL 489 breaker

The unit draws up to 10 A at 200–240 V; the site must provide external branch circuit protection according to UL 489 sized between 10 A and 16 A. Undersizing causes nuisance trips at full heating load, while oversizing leaves the supply cable unprotected.

Technical Data Sheet, Specifications (installation
Electrical
Verify supply quality before commissioning

Voltage and frequency tolerances of ±10% are only acceptable if they do not act in opposite directions. A supply where voltage and frequency deviate in opposite senses exceeds the permitted envelope and must be corrected before commissioning.

Technical Data Sheet, Specifications (installation
Process Limits
Cooling method sets the achievable temperature range

Without cooling the bath operates from 25 °C; water cooling extends the lower limit to 20 °C; only a refrigerator reaches -30 °C. The process temperature must be chosen together with the cooling configuration, and an immersion cooler TC45–TC100E can be fitted in the bath-bridge hole.

Technical Data Sheet, Specifications (selection
Process Limits
Do not use polyglycol with a Huber pump adapter

When a Huber pump adapter is fitted, polyglycol is not permissible as the heat transfer fluid. Fluid selection must be fixed before ordering the adapter, since the wrong combination cannot be corrected by a setting.

Technical Data Sheet, Specifications (operation
Process Limits
Charge at least 17 l before starting the pump

The bath holds 23.5 l but requires a minimum filling of 17 l. Operating below this level risks pump cavitation and invalid temperature control; the heat transfer fluid charge must be planned for the full external loop plus this minimum.

Technical Data Sheet, Specifications (operation
Mounting
Keep the IP20 enclosure dry and within 5–40 °C ambient

The IP20 rating excludes water exposure, so the unit must be installed in a dry, protected location. Ambient temperature must stay between 5 °C and 40 °C; outside this band the controller and pump may not meet their specified performance.

Technical Data Sheet, Specifications (installation
Electrical
Specify the country-specific plug at ordering

Single- and two-phase devices are delivered with a power cable and a country-specific plug that must be stated on the order. Three-phase devices below 63 A arrive with cable but without plug, so termination hardware must be planned by the site.

Technical Data Sheet, Specifications (installation
Control & I/O
Set the adjustable overtemperature cut-out for the application

The unit provides adjustable overtemperature protection according to DIN 12876. The cut-out must be configured for the fluid and process so that it protects both the heat transfer fluid and the immersed sample.

Technical Data Sheet, Specifications (safety

Selection & Integration Q&A

When should I choose the CC-225B rather than a smaller CC-20xB sibling?

Pick the CC-225B when the process needs the 23.5-litre bath and 17-litre minimum fill, or the 27 l/min pressure-pump delivery for a large external loop. Smaller siblings suit lower volume demands; the CC-225B is the largest bath in the series.

What must the external circuit and fluid be compatible with on the CC-225B?

The external loop must be servable by 27 l/min at 0.7 bar (pressure) and 22 l/min at 0.4 bar (suction). If a Huber pump adapter is used, polyglycol is not permissible as the heat transfer fluid.

What are the electrical and environmental limits for installing the CC-225B?

Supply 200–240 V, 1~ or 2~, 50/60 Hz, max 10 A, with a UL 489 breaker of 10–16 A. Ambient must stay 5–40 °C, the IP20 enclosure must be kept dry, and ±10% voltage/frequency tolerances must not act in opposite directions.

What cooling accessory is needed for sub-ambient operation, and what is not included in base delivery?

Without cooling the bath starts at 25 °C; water cooling gives 20 °C, and a refrigerator is required for -30 °C. The bath bridge has a hole for an immersion cooler TC45–TC100E. The country-specific plug must be specified at ordering.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CC-225B Immersion Circulator — Technical Data Sheet
Datasheet ·64,5 КБ ·2 pp.