Installation
The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and protection device needed before installation.
Source
“Power supply requirement (3 phase) 400V 3~N 50Hz”
Installation
The cooling water circuit needs a differential pressure of at least 3 bar and must not exceed 6 bar.
Ensures the available cooling water supply is adequate and safe for the water-cooled condenser.
Source
“min. cooling water differential pressure 3 bar”
Installation
Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Prevents condensation damage and loss of cooling performance at the installation site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected process performance.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.
Guides water treatment and material compatibility checks to avoid corrosion or fouling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable site power quality for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the plug.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The pump connection is M16x1 male and the maximum permissible kinematic viscosity is 50 mm²/s.
Ensures the external circuit and heat transfer fluid are compatible with the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms the safety concept required for flammable or sensitive bath fluids.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The refrigerant is R-452A with GWP 2141 and a CO2 equivalent of 3.5 t, which matters for leak monitoring and environmental reporting.
Relevant for refrigerant regulations, leak checks and environmental compliance.
Source
“CO2 equivalent 3,5 t”
Installation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.
Defines the permissible room conditions and protection against ingress.
Source
“max. ambient temperature 40 °C”
Selection
The bath holds 17 l, with a minimum filling of 15 l and 10 l when a displacement rack is used.
Helps select the right bath volume and displacement accessory for the application.
Source
“Bath capacity with displacement rack 10 l”