CC-906w Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of the CC-906w circulator
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits per DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C180 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
Bath volume30 l
min. filling capacity25 l
Bath capacity with displacement rack19 l
Width bath opening WxD / bath depth260 x 260 / 200 mm
Height of bath opening967 mm
Overall dimensions WxDxH **605x706x1216 mm
Net weight234 kg
Power supply requirement (3 phase)460V 3~ 60Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
max. ambient temperature40 °C
min. ambient temperature5 °C
from Serial-No.:S609285

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-90...200 °C
Temperature stability at -10°C0,02 K
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power3 kW
Cooling power at 200°C3 kW
Cooling power at 100°C3 kW
Cooling power at 20°C3 kW
Cooling power at 0°C3 kW
Cooling power at -20°C2,8 kW
Cooling power at -40°C2,4 kW
Cooling power at -60°C1,6 kW
Cooling power at -80°C0,55 kW
Cooling power at -90°C0,15 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-507 (A1, H280)
Global Warming Potential (GWP)3985
Refrigerant quantity1,12 kg
CO2 equivalent4,5 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,84 kg
CO2 equivalent 2nd stage12,4 t
Pressure pumpyes
max. delivery25 l/min
max. delivery pressure0,7 bar
Suction pumpyes
max. delivery (suction)18,5 l/min
max. delivery pressure (suction)0,4 bar
Pump connectionM16x1 male

Order-No.: CC-906w; 2036.0008.01

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit is water-cooled and requires a cooling water supply with a differential pressure between 3 and 6 bar and a flow of 180 l/h at 15 °C inlet.

Cooling water supply capacity and pressure must be verified on site before installation.

Source
“Consumption at water 15°C, flow 0°C 180 l/h”
Installation

Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the permissible cooling water temperature window and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Performance must be derated if the available cooling water is warmer than the reference 15 °C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Water quality and treatment must be checked for compatibility with these wetted materials.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Site power quality must be checked against these combined tolerances before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

External branch circuit protection according to UL 489 is required.

The electrical installation must include UL 489 branch protection; this affects panel and breaker selection.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug.

The customer must provide and install a suitable plug or hard-wired connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.

Confirms the safety concept and the need to set the overtemperature cut-out correctly for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The pressure pump delivers up to 25 l/min at 0.7 bar and the suction pump up to 18.5 l/min at 0.4 bar, with M16x1 male connections.

External circuit pressure drop and hose/connection sizing must be checked against these pump limits.

Source
“max. delivery 25 l/min”
Operation

The maximum permissible kinematic viscosity of the bath fluid is 50 mm²/s.

The heat transfer fluid must be selected so its viscosity at the lowest operating temperature stays within this limit.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit requires a 460 V 3~ 60 Hz supply with a maximum current of 14 A and 3x16 A fusing.

The site power supply and protection devices must match this three-phase requirement.

Source
“460V 3~ 60Hz”
Operation

The ambient temperature must remain between 5 °C and 40 °C and the unit has IP20 protection.

Installation location must respect ambient limits and avoid water ingress given IP20.

Source
“max. ambient temperature 40 °C”
Maintenance

The refrigerant is R-507 in the first stage and R-23 in the second stage, with CO2 equivalents of 4.5 t and 12.4 t respectively.

Refrigerant type and CO2 equivalent affect leak-check obligations and service requirements.

Source
“CO2 equivalent 2nd stage 12,4 t”
Selection

The bath holds 30 l, has a minimum filling of 25 l, and only 19 l with a displacement rack; the bath opening is 260 x 260 mm and 200 mm deep.

Useful bath volume and opening dimensions determine whether the application's sample load fits.

Source
“Bath capacity with displacement rack 19 l”