Peter Huber Kältemaschinenbau CC-906w Kälte-Badumwälzthermostat
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CC-906w Kälte-Badumwälzthermostat

Peter Huber Kältemaschinenbau CC-906w Kälte-Badumwälzthermostat — Thermostatic baths.

The CC-906w is a water-cooled refrigerated bath circulator with a 30 l bath and a -90…200 °C working range, using a two-stage cascade (R-507/R-23) and a variable-speed pressure/suction pump. It is the large-bath, deep-temperature modification of the CC series, distinguished by its 460 V 3~ 60 Hz supply and 3 kW heating capacity. It serves as a precision temperature source for external circuits and bath immersion duties requiring ±0.02 K stability.

Direct drop-in replacements

Eppendorf
MixMate ®
45%
match
⚖️ Compare

Peter Huber Kältemaschinenbau Cc Modelle Bis 90C Series — Model Comparison & Selection

Model Noise level, dBPower supplyPump connectionTemperature range, °CWeight, kgFilling volume, m³
CC-906w Kälte-Badumwälzthermostat Current —460V 3~ 60HzM16x1 AG male-90–2002340.025
CC-410wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-45–200710.017
CC-415wl Kälte-Badumwälzthermostat —220-240V 1~/2~ 50/60HzM16x1 AG male-40–200610.004
CC-508w Kälte-Badumwälzthermostat 57208-230V 1~/2~ 60HzM16x1 AG male-55–200—0.004
CC-510w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-50–2001060.016
CC-515w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001050.016
CC-520w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001610.015
CC-525w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-55–2001630.015
CC-905 Kälte-Badumwälzthermostat 64400V 3~N 50HzM16x1 AG male-90–2001710.019
CC-905w Kälte-Badumwälzthermostat —400V 3~N 50HzM16x1 AG male-90–2001750.019
CC-K12 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–200290.007
CC-K15 Kälte-Badthermostat —220-240V 1~/2~ 50/60Hz(Option, #19607)-20–20028—
CC-K20 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.012
CC-K25 Kälte-Badthermostat 50220-240V 1~/2~ 50/60Hz(Option, #19607)-30–200410.017
CC-K6 Kälte-Badumwälzthermostat 57208-240V 1~/2~ 50/60HzM16x1 AG male-25–20025—
Variostat Kältethermostat für variable Badgefäße 54220-240V 1~/2~ 50/60HzM16x1 AG male-30–15025—

Full specifications

Core Performance & Functional Specifications
Bath volume 30 l
Filling volume 25 l
Heating capacity 3 kW
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure and suction pump, variable speed
External sensor connection Pt100
Temperature range -90...200 °C
Temperature stability mfr ±0, 02 K
Dimensions (W x D x H) mfr 605 x 706 x 1216 mm
Delivery capacity pressure pump max. mfr 25 l/min; 0, 7 bar
Delivery capacity suction pump max. mfr 18, 5 l/min; 0, 4 bar
Bath opening WxD/BD mfr 260 x 260 / 200
Dimensions, Weight & Mechanical Design
Weight 234 kg
Electrical, Power & Drive Specifications
Current consumption 14 A (max.)
Power supply 460V 3~ 60Hz
Fuse mfr 3x16A
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 180 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 3 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

The CC-906w is a floor-standing refrigerated bath circulator (605 x 706 x 1216 mm, 234 kg) with a 30 l bath, 25 l minimum filling, and a 260 x 260 mm opening 200 mm deep. Within the CC series, it is the largest and coldest variant: siblings such as the CC-405w, CC-410wl, CC-415wl and CC-505wl are smaller-bath models, while this unit combines a 30 l working volume with a -90 °C lower limit via a two-stage cascade refrigerant circuit (R-507 first stage, R-23 second stage). The Pilot ONE controller provides the setpoint and safety layer, with safety classification III / FL and adjustable overtemperature protection per DIN 12876.

Working behavior and media

Installation, connection and control

The unit requires a 460 V 3~ 60 Hz supply, 14 A maximum current and 3x16 A fusing, with external UL 489 branch protection; three-phase devices below 63 A are delivered with a cable but no plug. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not permitted. Cooling water connects via G1/2 male fittings at 3 bar minimum differential pressure and 6 bar maximum; the wetted circuit includes copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible. Control and data interfaces are Ethernet, USB (host and device) and RS232, with a Pt100 external sensor connection. Ambient temperature must remain 5…40 °C, and IP20 protection means the unit must be kept away from water ingress.

Integration & Operating Notes

Cooling & Media
Cooling water must be 5–20 °C at 3–6 bar differential

The water-cooled condenser requires a supply within this window; warmer water reduces cooling power and raises the required flow rate, so the site supply must be verified before commissioning. Pipes must be insulated where condensation can form.

CC-906w Immersion Circulator Datasheet, page 2
Electrical
External UL 489 branch protection is mandatory

The 460 V 3~ 60 Hz supply must be protected by a UL 489 branch circuit device sized for the 14 A maximum current and 3x16 A fusing; the unit is delivered with a cable but without a plug, so the customer must provide the final connection.

CC-906w Immersion Circulator Datasheet, page 2
Electrical
Voltage and frequency tolerances are coupled

A ±5% voltage tolerance is permitted only with a simultaneous ±2% frequency tolerance; the combination of -5% voltage with +2% frequency is explicitly not allowed, so site power quality must be checked against both together.

CC-906w Immersion Circulator Datasheet, page 2
Process Limits
Bath fluid viscosity must stay below 50 mm²/s

The heat transfer fluid must be selected so that its kinematic viscosity at the lowest operating temperature does not exceed 50 mm²/s; otherwise pump delivery and temperature uniformity are compromised.

CC-906w Immersion Circulator Datasheet, page 2
Process Limits
Displacement rack cuts usable volume to 19 l

With a displacement rack installed, the bath capacity falls from 30 l to 19 l, so sample loads and filling volume must be planned against the reduced figure; minimum filling without the rack is 25 l.

CC-906w Immersion Circulator Datasheet, page 2
Maintenance
Refrigerant leak checks are required for R-507/R-23

The cascade circuit uses R-507 (4.5 t CO2 equivalent) and R-23 (12.4 t CO2 equivalent); these charge sizes trigger periodic leak-check obligations and service access planning for the refrigerant circuit.

CC-906w Immersion Circulator Datasheet, page 1
Mounting
IP20 enclosure requires a dry, 5–40 °C location

The unit must be installed where ambient temperature stays between 5 °C and 40 °C and where water cannot enter the IP20 enclosure; the floor-standing 234 kg frame also requires a level, load-bearing surface.

CC-906w Immersion Circulator Datasheet, page 2

Selection & Integration Q&A

When should I choose the CC-906w over the smaller CC-405w or CC-505wl siblings?

Choose the CC-906w when you need a 30 l bath and the -90 °C lower limit from the two-stage cascade; the smaller siblings suit lower volume or less deep temperature duties. The 906w also brings a 3 kW heater and 460 V 3~ supply, so confirm site power before selecting it.

What cooling water supply and piping must be provided for the CC-906w?

Provide cooling water at 5–20 °C inlet with 3–6 bar differential pressure and 180 l/h flow at 15 °C; connect via G1/2 male fittings. The wetted circuit includes copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so water quality must be compatible, and pipes need insulation where condensation can occur.

What are the electrical and environmental limits for installing this unit?

It requires 460 V 3~ 60 Hz, 14 A max, 3x16 A fusing and external UL 489 branch protection; a cable is supplied but no plug. Voltage tolerance ±5% is allowed only with ±2% frequency tolerance, and the -5% voltage/+2% frequency combination is prohibited. Ambient must stay 5–40 °C and IP20 requires a dry location.

What maintenance obligations and delivery accessories should I plan for?

The R-507/R-23 cascade carries 4.5 t and 12.4 t CO2 equivalent, so periodic refrigerant leak checks are required. The unit ships with a cable but no plug, and no external Pt100 sensor or bath fluid is included; the displacement rack is an accessory that reduces bath capacity to 19 l.