The unit must not be installed or started in an ATEX zone because of explosion risk.
Determines whether the installation site is permissible at all.
Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
| Раздел | Название | Страница |
|---|---|---|
| 4.2.8 | Sub-category: “Configuration auto” | 47 |
| 4.2.8.1 | Sub-category: “Find parameters” | 47 |
| 4.2.8.2 | Sub-category: “Control Dynamics” | 49 |
| 4.2.8.3 | Sub-category: “Fluid Properties” | 50 |
| 4.2.8.4 | Sub-category: “Display parameters” | 51 |
| 4.2.9 | Sub-category: “Configuration manual” | 52 |
| 4.2.9.1 | Sub-category: “Change parameters” | 52 |
| 4.2.9.2 | Sub-category: “Display parameters” | 53 |
| 4.2.9.3 | Sub-category: “Control structure” | 53 |
| 4.2.10 | Sub-category: “Reset parameters” | 53 |
| 4.2.11 | Sub-category: “Display parameters” | 54 |
| 4.2.12 | Setting the setpoint thresholds | 54 |
| 4.2.13 | Setting the setpoint | 55 |
| 4.3 | Filling, venting, degassing and draining | 55 |
| 4.3.1 | Externally closed application | 55 |
| 4.3.1.1 | Filling and venting externally closed application | 55 |
| 4.3.1.2 | Degassing externally closed applications | 57 |
| 4.3.1.3 | Draining externally closed applications | 59 |
| 5 | Normal operation | 61 |
| 5.1 | Automatic operation | 61 |
| 5.1.1 | Temperature control | 61 |
| 5.1.1.1 | Starting the temperature control process | 61 |
| 5.1.1.2 | Ending the temperature control process | 61 |
| 5.1.2 | Temperature control via a created temperature control program | 62 |
| 5.1.2.1 | Starting the temperature control program | 62 |
| 5.1.2.2 | Ending/cancelling the temperature control program | 62 |
| 6 | Interfaces and software update | 63 |
| 6.1 | Interfaces at the “Pilot ONE®” controller | 63 |
| 6.1.1 | 10/100 Mbps Ethernet for RJ45 network sockets | 63 |
| 6.1.2 | USB-2.0 interface | 63 |
| 6.1.2.1 | USB-2.0 interface, host | 64 |
| 6.1.2.2 | USB-2.0 interface, device | 64 |
| 6.2 | Interfaces on the temperature control unit | 64 |
| 6.2.1 | Interfaces at the back | 64 |
| 6.2.1.1 | Service interface | 64 |
| 6.2.1.2 | Female RS232 serial | 64 |
| 6.2.2 | Interfaces at the top | 65 |
| 6.2.2.1 | Connection jack for Pt100 process controller sensor | 65 |
| 6.3 | Interfaces at the Com.G@te® (optional) | 65 |
| 6.3.1 | Jack LEVEL (Com.G@te® external only) | 66 |
| Description | (A) | (B) | (C) | (D) | (E) | (F) |
|---|---|---|---|---|---|---|
| Acoustics | X | X | X | – | – | – |
| Display functions Warnings; Messages; Inactive Menu Items; Resolution Temperature | X | X | X | – | – | – |
| Display modes | X | X | X | – | – | – |
| AutoStart response | X | X | X | – | – | – |
| Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units) | X | X | X | – | – | – |
| Com.G@te (only if Com.G@te is connected) Analog Interface (configuration input) Meaning of AIF input; response on cable break; measurement range setting | X | X | X | – | – | X |
| Calibration | – | – | – | – | – | X |
| In case of analog error; current / voltage switching; filter constant | X | X | X | – | – | X |
| Configuration Output AIF output variable; Output range setting | X | X | X | – | – | X |
| Calibration | – | – | – | – | – | X |
| RS232/RS485 (Hardware RS, Baud rate, device address); ECS Standby; POKO Alarm | X | X | X | – | – | – |
| Properties (other) Change thermal fluid (thermal fluid, liter indicator, bypass usage); cooling bath / cooling (CC-E only) | X | X | X | – | – | – |
| Function | E-grade Basic | E-grade Exclusive | E-grade Professional |
|---|---|---|---|
| Temperature control | |||
| Controller parameterization: predefined1 / TAC2 | X/– | –/X | –/X |
| Find parameters: Fast Identification / With Preliminary Test / Estimate Control Parameters | –/–/X | X/X/– | X/X/– |
| Sensor calibration for control sensor3: x -point | 2 | 5 | 5 |
| Monitoring: Low level and overtemperature4 | X | X | X |
| Adjustable alarm thresholds | X | X | X |
| VPC5 (Variable Pressure Control) | X | X | X |
| Ventilation program | X | X | X |
| Automated compressor | X | X | X |
| Setpoint limitation | X | X | X |
| Program controller: x programs / max x steps | –/– | 3/15 | 10/100 |
| Ramp function: linear / non-linear | –/– | X/– | X/X |
1 TAC function available as a 30 day evaluation version. 2 True Adaptive Control. 3 Internal Pt100 and External Pt100. 4 For temperature control units with integrated overtemperature protection. 5 For units with variable-speed pump or external bypass.
| Description | (A) | (B) | (C) | (D) | (E) | (F) |
|---|---|---|---|---|---|---|
| Sensor calibration | – | – | – | – | – | – |
| Automated compressor | X | X | X | – | – | – |
| Enter program | X | – | – | X | – | – |
| Pump settings Pump speed setpoint; Pump pressure setpoint; Control Mode | X | X | X | – | – | – |
| Fluid properties Thermal fluid; liter indicator | X | X | X | – | – | – |
| Protective function High Limit Alarm Internal; Low Limit Alarm Internal; High Limit Alarm Process; Low Limit Alarm Process; Hydrostatic Correction; Warning Time Limit (CC-E only); Min. Level (for temperature control units with analogue level sensor); Max. Level (for temperature controllers with analogue level sensor); Freeze Protection (optional) | X | X | X | – | – | – |
| Setpoint | X | X | X | – | – | – |
| Setpoint limitation Min. Setpoint; Max. Setpoint | X | X | X | – | – | – |
| Language | X | X | X | – | – | – |
| Temperature format | X | X | X | – | – | – |
| Temperature control mode | X | X | X | – | – | – |
| Favourites menu | X | – | – | – | X | – |
| Pilot ONE Ethernet IP address; Subnet Mask; Remote Access | X | X | X | – | – | – |
| Time format | X | X | X | – | – | – |
| 2. Setpoint | X | X | X | – | – | – |
| Designation | Specification |
|---|---|
| Calcium carbonate per liter | ≤ 1.5 mmol/l; corresponds to a water hardness of: ≤ 8.4 °dH (soft) |
| PH value | between 6.0 and 8.5 |
| Ultrapure water, distillates | Add 0.1 g of sodium carbonate (Na2CO3) per liter |
| Not approved water | Distilled, deionized, demineralized, chloric, ferruginous, ammoniacal, or contaminated river water or sea water |
| Thermal fluid: Water without ethylene glycol | Use (+15) +65 °C to +90 °C |
| Thermal fluid: Water-ethylene glycol mixture | do not use |
| Function | E-grade Basic | E-grade Exclusive | E-grade Professional |
|---|---|---|---|
| Communication Watchdog | – | – | X |
| Process data recording directly to a USB stick: Setpoint, Actual Value Internal and Actual Value Process / Heating Capacity %, Cooling Capacity %, and Pump Pressure / Pump Speed, and VPC Pressure | –/–/– | X/X/– | X/X/X |
| Calendar start | – | – | X |
| Pin | Signal | Description |
|---|---|---|
| 2 | RxD | Receive Data |
| 3 | TxD | Transmit Data |
| 5 | GND | Signal GND |
Pin assignment (front view)
| Pin | Description | Signal |
|---|---|---|
| 1 | Current output, T external | 0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor. |
| 2 | Current output, setpoint | 0 a)/4 - 20 mA or 0 - 10 V |
| 3 | GND for analog outputs | GND |
| 4 | Analog input (programmable) | 0 a)/4 - 20 mA or 0 - 10 V Power input: 200 Ω working resistance Voltage input: 100 kΩ input resistance |
| 5 | Current output, freely programmable | 0 a)/4 - 20 mA or 0 - 10 V If using 0 - 10 V, integrate a 500 Ω resistor. |
| 6 | GND for analog input | GND |
a)Consult with our Customer Support.
| Pin | Signal |
|---|---|
| 1.3 | E2 |
| 2 | E1 |
The unit must not be installed or started in an ATEX zone because of explosion risk.
Determines whether the installation site is permissible at all.
“Do NOT install or start up the temperature control unit within an ATEX zone.”
A condensation water / thermal fluid drip tray must be installed below the unit by the responsible body.
Affects site preparation and containment of leaks.
“The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit.”
The unit must be allowed to acclimate about 2 hours when moving from cold to warm environment before switching on.
Prevents damage and malfunction after transport or storage.
“Allow the temperature control unit to acclimate for about 2 hours when changing from a cold to a warm environment (or vice versa). Do not turn on the temperature control unit before!”
Casters must be locked after installation to prevent crushing of limbs.
Safety requirement for units supplied with casters.
“Wheels, if installed, must be locked after installation!”
The building supply voltage and frequency must match the rating plate data.
Prevents damage to the unit and ensures correct electrical connection.
“Your building's existing power supply voltage and frequency must match the data provided on the rating plate of the temperature control unit.”
Hard-wired power connection must be carried out by an electrician.
Mandatory qualification requirement for hard-wired installations.
“Have the connection/adjustment to the power supply carried out by an electrician.”
Power cable must not be longer than 3 m.
Limits cable routing and placement of the unit relative to the supply.
“Do not use a power cable that is longer than 3 m.”
Maximum current at 230 V operation is just below 16 A and the user can select 16 A, 13 A or 10 A power consumption.
Determines fuse protection and whether the supply is adequate.
“The maximum current consumption at 230 V operation is just below 16 A.”
Maximum flow pressure must be considered for pressure-sensitive applications such as glass reactors.
Prevents overpressure damage to glass apparatus.
“The maximum flow pressure of a temperature control unit must be taken into account in case of pressure-sensitive applications, such as glass reactors.”
Thermal fluid must be adjusted to room temperature before switching off to prevent overpressure.
Protects the unit and application from pressure damage during shutdown.
“In order to prevent the risk of overpressure in the system, the thermal fluid must always be adjusted to room temperature before switching off.”
Temperature control hoses should be kept as short as possible and their inside diameter must at least match the pump connections.
Affects pressure drop and achievable temperature control performance.
“The inside diameters of the temperature control hoses must correspond at least to the pump connections.”
After pressing Stop the pump continues to run for about 30 seconds.
Operators must wait for the pump to stop before servicing or draining.
“temperature control will stop immediately and the pump with overrun for about 30 seconds.”