Induztr
Engineering Digest Document Viewer Related Models

Chili Immersion Circulators

Contents

  1. Overview p. 1
    Product introduction and model lineup of Chili circulators
  2. Spec p. 2
    Specifications according to DIN 12876 for the models
  3. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Operating temperature range65... 300 °C
Temperature stability at 70°C0,02 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
External sensorPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V3 kW
Heating power at 230V3 kW
Heating power at 220V3 kW
Heating power at 208V3 kW
Heating power at 200V2,7 kW
Circulation pump: max. delivery45 l/min
max. delivery pressure0,9 bar
Pump connectionM24x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity1,45 l
Filling capacity expansion tank2 l
Overall dimensions WxDxH **240x427x393 mm
Net weight23 kg
Power supply requirement200-240V 1~/2~ 50/60Hz
max. current15,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S369994

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a heating circulator intended for closed systems, so it cannot be used for open bath applications.

Determines whether the circulator matches the intended closed-loop process or requires a different open-bath model.

Source
“Heating Circulator for closed systems.”
Installation

The pump connection is M24x1.5 male, so mating hoses or adapters must match this thread.

Prevents ordering incompatible hoses or fittings and ensures leak-free connection at commissioning.

Source
“Pump connection M24x1,5 male”
Installation

The electrical supply must be 200-240 V single or two-phase at 50/60 Hz with a 16 A fuse and 15.5 A maximum current.

Defines the branch circuit, protection device, and supply configuration needed before installation.

Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 15,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are limited to ±5% voltage with simultaneous ±2% frequency, and some combinations are explicitly not allowed.

Site power quality must be checked; out-of-tolerance supply can prevent safe and reliable operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

External branch circuit protection according to UL 489 is required.

The installer must provide a compliant external breaker or fuse; the unit does not include branch protection.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Single- and two-phase devices are supplied with a power cable and country-specific plug, which must be specified at order.

Avoids delivery of the wrong plug type and delays during commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The ambient temperature must remain between 5 °C and 40 °C for proper operation.

The installation location must be assessed for room temperature before commissioning.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s.

Selecting a fluid with too high viscosity can overload the pump and degrade circulation.

Source
“max. permissible kin. viscosity 50 mm²/s”
Operation

The minimum filling capacity is 1.45 l and the expansion tank holds 2 l.

Determines the minimum system volume and expansion allowance needed for safe thermal expansion.

Source
“min. filling capacity 1,45 l Filling capacity expansion tank 2 l”
Performance

The circulation pump delivers up to 45 l/min at a maximum delivery pressure of 0.9 bar.

Used to check whether the pump can overcome the pressure drop of the external circuit at the required flow.

Source
“max. delivery 45 l/min max. delivery pressure 0,9 bar”
Compliance

The equipment is classified as pressure equipment category Art. 4.3 PED.

Affects documentation and conformity requirements for the pressure-containing circuit.

Source
“Pressure equipment category Art. 4.3 PED”
Compliance

The disturbance classification is EN55011 Class A, Group 1.

Class A equipment may require additional EMC measures in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The pump connection bore shape is Y (60°) per DIN 3863, with ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures the correct hose end form and nut are used to achieve a proper seal.

Source
“Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety

The unit has adjustable overtemperature protection according to DIN 12876.

The overtemperature limit must be set appropriately during commissioning to protect the process and fluid.

Source
“With adjustable overtemperature protection according to DIN 12876.”