Product introduction and model identification for CS 25 bath
Specp. 2
Performance and physical data according to DIN 12876 standard
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...15 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at ambient temperature 20°C
at 15°C
3,5 kW
at 10°C
2,5 kW
at 0°C
1,6 kW
at -10°C
0,98 kW
at -20°C
0,7 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
2 kg
CO2 equivalent
2,79 t
Circulation pump:
max. delivery
40 l/min
max. delivery pressure
4,5 bar
Pump connection
G1/2 male
min. filling capacity
24 l
max. filling capacity
27 l
Overall dimensions WxDxH **
580x660x820 mm
Net weight
110 kg
sound pressure level +/- 4 dB(A)
60 dB(A)
Power supply requirement
230V 1~ 50Hz
max. current
12,5 A
Fuse
16 A
Degree of Protection
IP52
min. ambient temperature
5 °C
max. ambient temperature
32 °C
from Serial-No.:
1.0/24
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
12 facts · every claim linked to a page
Selection
The chiller is only suitable for operation with ambient temperatures between 5 °C and 32 °C, so the installation site must be kept within this range.
Ambient temperature outside this range can prevent proper cooling and may damage the refrigeration circuit.
Source
“min. ambient temperature
5 °C
max. ambient temperature
32 °C”
Installation
The unit requires a 230 V single-phase 50 Hz supply with a 16 A fuse and draws up to 12.5 A, so the electrical supply must be sized accordingly.
Undersized wiring or fusing will cause nuisance tripping or overheating during startup and full-load operation.
Source
“Power supply requirement
230V 1~ 50Hz
max. current
12,5 A
Fuse
16 A”
Installation
Voltage and frequency tolerances are interdependent: a -5% voltage deviation is only allowed with -2% frequency deviation, not with +2%.
This prevents motor overheating and ensures the compressor and pump operate within EN60034-1 limits.
Source
“-5% voltage and +2% frequency -> not allowed!
-5% voltage and - 2% frequency -> allowed”
Installation
The chiller has IP52 protection, so it must be installed where it is not exposed to water jets or heavy dust.
IP52 limits the environments where the unit can be safely installed and cleaned.
Source
“Degree of Protection
IP52”
Operation
The recommended heat transfer fluid is a 50:50 water-monoethylene glycol mixture, which must be used to avoid freezing and corrosion.
Using the wrong fluid can cause freezing, corrosion, or reduced heat transfer at the -20 °C lower limit.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation
The tank holds between 24 l and 27 l of fluid, so the external loop and expansion volume must be planned within this fill range.
Underfilling causes pump cavitation and poor temperature stability; overfilling causes overflow from the atmospheric tank.
Source
“min. filling capacity
24 l
max. filling capacity
27 l”
Performance
Cooling capacity is rated at 20 °C ambient and drops if ambient rises, so actual capacity at the installation site may be lower than the table values.
Sizing the chiller without derating for a hot plant room can lead to insufficient cooling at peak load.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The refrigerant is R-449A with a GHS hazard classification H280, so leak detection and ventilation must account for a pressurized gas hazard.
H280 indicates a pressurized gas hazard, which affects handling, leak response, and room ventilation requirements.
Source
“R-449A (A1, H280)”
Compliance
The unit is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Class A equipment may not be suitable for residential use without additional filtering or separation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
Single-phase devices are supplied with a country-specific plug only if specified at ordering, so the plug type must be confirmed with the order.
Without specifying the plug, the unit may arrive without a usable mains connection for the site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The flow rate is adjusted via a manual bypass valve on the backside, so access to the rear of the unit is needed for commissioning.
Rear clearance must be planned so the bypass valve can be reached during setup and operation.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Maintenance
The unit has a digital level indicator, which should be checked to keep the fluid between the minimum and maximum fill limits.
Level monitoring prevents low-level pump damage and overflow from the open tank.