Peter Huber Kältemaschinenbau CS 25 by Van der Heijden
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau CS 25 by Van der Heijden

Peter Huber Kältemaschinenbau CS 25 by Van der Heijden — Recirculating chillers with Noise level 60 dB.

The CS 25 is a 37-litre circulating bath chiller with a -20…15 °C working range, positioned as the smallest capacity model in the Cs Chiller series. It uses R-449A refrigerant and a 50:50 water-glycol mixture to hold process temperature in closed loops. Its 230 V single-phase supply and G1/2 male pump connection suit pilot-scale or laboratory cooling duties.

Noise level
60 dB

Peter Huber Kältemaschinenbau Cs Chiller Series — Model Comparison & Selection

Model Noise level, dBAmbient temperature limit, °CMinimum fill volume, m³Power supplyPump connectionTemperature range, °C
CS 25 by Van der Heijden Current 605–320.024230V 1~ 50HzG1/2 AG male-20–15
CS 100 by Van der Heijden 705–320.16400V 3~ 50HzG3/4 AG male-20–15
CS 100-H by Van der Heijden 715–320.175400V 3~ 50HzG3/4 AG male-20–85
CS 200 by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–15
CS 200-H by Van der Heijden 785–320.225400V 3~ 50HzG1 AG male-20–85
CS 20-H by Van der Heijden 575–320.024230V 1~ 50HzG3/8 AG male-20–85
CS 50 by Van der Heijden 635–320.1400V 3~ 50HzG3/4 AG male-20–15
CS 75 by Van der Heijden 665–320.16400V 3~ 50HzG3/4 AG male-20–15
Huber CS 200-H 785–320.225460V 3~ 60HzG1 AG male-20–85
Huber CS 25 605–320.024208V 2~ 60HzG1/2 AG male-20–15
Huber CS 250HT OD 78-20–500.16460V 3~ 60HzG3/4 AG male10–22
Huber CS 330-H OD 78-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550-H OD 79-20–500.16460V 3~ 60HzG1 AG male-20–85
Huber CS 550 OD 79-20–500.16460V 3~ 60HzG1 AG male-20–15
Huber LCS 80 77-20–500.04460V 3~ 60HzG1 AG male-45–15

Full specifications

Core Performance & Functional Specifications
Noise level 60 dB(A)
Bath volume 37 l
Minimum fill volume 24 l
Pump connection G1/2 AG male
Temperature range -20...15 °C
Dimensions, Weight & Mechanical Design
Weight 110 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 230V 1~ 50Hz
Operating Conditions, Environment & Protection
Ambient temperature limit 5...32 °C
Protection rating IP52
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behaviour in operation

Installation and control

Integration & Operating Notes

Electrical
Supply must tolerate 12.5 A continuous draw on a 16 A fuse

The 230 V single-phase 50 Hz supply must be sized for a maximum current of 12.5 A and protected by a 16 A fuse. Voltage and frequency tolerances are interdependent: a -5% voltage deviation is permitted only with -2% frequency deviation, not with +2%, so the site supply quality must be checked before commissioning.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Process Limits
Fill volume must stay between 24 and 27 litres

The 37-litre tank accepts a working fill of 24–27 l. The external loop, hoses and expansion volume must be counted so the unit is neither underfilled (pump cavitation, poor temperature stability) nor overfilled (overflow from the atmospheric tank). The digital level indicator should be checked against these limits.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Cooling & Media
Use 50:50 water–monoethylene glycol to reach -20 °C

The recommended heat-transfer fluid is a 50:50 water–monoethylene glycol mixture. Using plain water or a different ratio risks freezing, corrosion or reduced heat transfer at the -20 °C lower limit, so the fluid must be specified before the loop is filled.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Mounting
Install only where ambient stays between 5 and 32 °C

Operation outside 5–32 °C ambient can prevent proper cooling and may damage the refrigeration circuit. Because cooling capacity is rated at 20 °C ambient and drops as ambient rises, a hot plant room also reduces available capacity below the datasheet values, so the site must be assessed for both limits.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Control & I/O
Rear access needed to set flow via bypass valve

Flow rate is adjusted by a manual bypass valve on the backside of the chiller. The installation layout must leave clear rear access for commissioning and later adjustment, otherwise the loop flow cannot be balanced to the process requirement.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Electrical
Mains plug only supplied if specified at order

Single-phase devices are delivered with a power cable and a country-specific plug only when this is specified at ordering. If the plug type is not confirmed with the order, the unit may arrive without a usable mains connection for the site, delaying commissioning.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications
Process Limits
R-449A leak response must account for H280 hazard

The refrigerant is R-449A with GHS classification H280 (pressurized gas). Room ventilation and any leak-detection provisions must be designed for a pressurized-gas hazard rather than a simple asphyxiation or flammability scenario.

CS 25 Circulating Bath — Technical Datasheet, Technical Specifications

Selection & Integration Q&A

When should I choose the CS 25 rather than a larger Cs Chiller sibling such as the CS 100 or CS 125?

Pick the CS 25 when the duty fits a 37 l bath, a 24–27 l fill and a -20…15 °C range at 60 dB(A). The CS 100/125 and -H variants are larger, noisier units (70–75 dB(A)) for higher capacity; the CS 25 is the smallest modification in the series.

What supply and loop connections must be ready before the CS 25 is installed?

A 230 V single-phase 50 Hz supply protected by a 16 A fuse, sized for 12.5 A maximum current, with the country-specific plug confirmed at ordering. The process loop connects to the G1/2 AG male pump connection, and the fill must land between 24 and 27 l.

What ambient and installation limits apply to the CS 25, and why do they matter for capacity?

The site must stay between 5 and 32 °C ambient; outside this the refrigeration circuit can be damaged. Cooling capacity is rated at 20 °C ambient and drops as ambient rises, so a hot plant room reduces available cooling below datasheet values. IP52 also rules out washdown and heavy dust.

What is not included in the base delivery and what must be planned for commissioning and upkeep?

The country-specific mains plug is supplied only if specified at ordering. The 50:50 water–monoethylene glycol fluid must be provided on site, and rear access is needed to set flow via the manual bypass valve. The digital level indicator should be checked to keep fill within 24–27 l.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
CS 25 Circulating Bath — Technical Datasheet
Datasheet ·140,3 КБ ·2 pp.
View PDF