  1. Overview p. 1
    Product name and cover page of the datasheet
  2. Spec p. 2
    Technical data according to DIN 12876 standard

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -50 to 50 °C, so it must be selected only for applications within this span.

Defines the process temperature window the cold trap can cover.

Source
“Operating temperature range -50...50 °C”
Safety

The refrigerant is R-290 (propane, A3, H220), a flammable natural refrigerant, which affects site safety and ventilation requirements.

Flammable refrigerant classification drives installation location and safety measures.

Source
“R-290 (A3, H220)”
Installation

The unit requires a 220-240 V single/two-phase 50/60 Hz supply with max current 5 A and a fuse between 10 A and 16 A.

Determines electrical supply and protective device sizing at the installation site.

Source
“220-240V 1~/2~ 50/60Hz max. current 5 A min. Fuse 10A max. Fuse 16A”
Installation

The enclosure is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

IP20 limits where the unit can be safely placed.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C for proper operation.

Room conditions outside this range may impair cooling and reliability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is specified at 20 °C room temperature and may drop if ambient temperature rises.

Performance must be derated for warmer installation rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but simultaneous opposite-sign deviations are not allowed.

Supply quality must be checked to avoid invalid operating conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

The correct plug type must be requested to avoid rework at installation.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment is EMC Class A Group 1, which may require additional measures in residential environments.

Class A equipment is intended for industrial environments and may cause interference elsewhere.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Space requirements around the unit must be respected as stated in the operating conditions.

Clearances are needed for air-cooled operation and service access.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation

The controller supports autostart after power failure, which affects restart behavior of the process.

Engineers must know whether the unit restarts automatically after mains interruption.

Source
“Autostart function for power failure”
Selection

An optional Pt100 sensor connection can display (not control) process temperature and is only available factory-fitted.

If process temperature monitoring is needed, it must be ordered with the unit.

Source
“Option: Pt100 sensor connection to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”