Contents

  1. Overview p. 1
    Product identification and general description of the CS 25 unit
  2. Spec p. 2
    Technical data according to DIN 12876 for the CS 25
  3. p. 1 table, 31 rows

Technical data according to DIN 12876 — Huber CS 25 (VDH1100664)

ParameterValue
Operating temperature range-20...15 °C
temperature set point / displaycolour LED Touchscreen
Internal temperature sensorPt100
Temperature stability at -10°C1,5 K
Interface digitalRS232
Safety classificationI / NFL
Cooling power at ambient temperature 20°C
at 15°C3,5 kW
at 10°C2,5 kW
at 0°C1,6 kW
at -10°C0,98 kW
at -20°C0,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Refrigerant quantity2 kg
CO2 equivalent2,79 t
Circulation pump: max. delivery50 l/min
max. delivery pressure4,5 bar
Pump connectionG1/2 male
min. filling capacity24 l
max. filling capacity37 l
Overall dimensions WxDxH **580x660x820 mm
Net weight121 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement208V 2~ 60Hz
max. current12,5 A
Fuse2x20 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature32 °C
from Serial-No.:1.0/20

** Overall dimensions WxDxH. Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller's cooling capacity is rated at 20°C ambient and will drop if ambient temperature rises, so sizing must account for actual room conditions.

Cooling capacity must be derated for higher ambient temperatures to avoid undersizing.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 208V two-phase 60Hz supply with 12.5A max current and 2x20A fuses, so the electrical supply must match this configuration.

Electrical supply and protection must be provisioned accordingly before installation.

Source
“Power supply requirement 208V 2~ 60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.

Supply quality must be verified to avoid invalid tolerance combinations that could damage the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices under 63A are supplied with cable but without plug, so a suitable plug must be provided on site.

Planning must include procurement and wiring of a compatible plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The chiller must be installed in an ambient temperature between 5°C and 32°C.

Site conditions outside this range will affect performance and may void warranty.

Source
“min. ambient temperature 5 °C”
Operation

The recommended thermofluid is a 50:50 water-monoethylene glycol mixture.

Using the correct fluid ensures proper heat transfer and freeze protection.

Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation

The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.

Operators need to know how to regulate flow to match process requirements.

Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurized gas), requiring appropriate handling precautions.

Personnel must be aware of the refrigerant hazard for safe installation and maintenance.

Source
“R-449A (A1, H280)”
Safety

The unit has safety classification I / NFL according to DIN 12876.

This classification determines the required safety measures and permissible fluids.

Source
“Safety classification I / NFL”
Compliance

The equipment emits Class A, Group 1 disturbances per EN55011, so it may cause radio interference in residential areas.

Installation in residential environments may require additional EMC mitigation.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Compliance

The equipment is compliant with US-SNAP for industrial process refrigeration and all applicable EU laws.

Regulatory compliance must be confirmed for the intended market and application.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws.”
Maintenance

The refrigerant charge is 2 kg with a CO2 equivalent of 2.79 t, which must be recorded for leak checks and F-gas regulations.

Refrigerant quantity determines leak inspection intervals and environmental reporting obligations.

Source
“Refrigerant quantity 2 kg”
Performance

The circulation pump delivers up to 50 l/min at 4.5 bar, which must be sufficient for the process loop pressure drop.

Pump performance must match the external circuit resistance to ensure adequate flow.

Source
“max. delivery pressure 4,5 bar”