Product introduction and main features of KISS 108A
Specp. 2
Specifications according to DIN 12876 standard
p. 1table, 37 rows
Technical data according to DIN 12876 — KISS 108A
Parameter
Value
Order-No.
2050.0003.98
Operating temperature range
25...100 °C
with water cooling
20...100 °C
with refrigerator
15...100 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Absolute accuracy
setup for calibration
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Bath volume
6 l
min. filling capacity
3,5 l
Height of bath opening
160 mm
Width bath opening WxD/ bath depth
130x210/ 150 mm
Overall dimensions WxDxH **
147x407x330 mm
Net weight
6 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
402345
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method, so the choice of water cooling or refrigerator must match the required process temperature.
Determines whether the bath can reach the required low temperature and which cooling option must be ordered.
Source
“Operating temperature range 25...100 °C with water cooling 20...100 °C with refrigerator 15...100 °C”
Installation
The unit requires a 200–240 V single- or two-phase supply at 50/60 Hz with a maximum current of 10 A.
Needed to confirm site power availability and correct branch circuit sizing before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz max. current 10 A”
Installation
External branch circuit protection according to UL 489 is mandatory.
The installer must provide a UL 489 listed breaker; otherwise the installation is non-compliant.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not act in opposite directions.
Prevents commissioning on a supply where combined voltage and frequency deviations exceed the allowed envelope.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single- and two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.
The correct plug type must be stated on the order, otherwise the unit may not connect to the site outlet.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must be avoided or a more resistant plastic ordered.
Prevents premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure.
Guides the correct coolant preparation to avoid freezing and damage at low temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Operation
The bath needs a minimum filling capacity of 3.5 l within its 6 l volume.
Underfilling risks pump cavitation and poor temperature control; overfilling risks overflow.
Source
“Bath volume 6 l min. filling capacity 3,5 l”
Operation
The unit must only be operated within an ambient temperature of 5 °C to 40 °C.
Defines the allowable room conditions for reliable operation and warranty compliance.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety
The device is safety class III / FL with adjustable overtemperature protection per DIN 12876.
Confirms the safety classification required for flammable liquids and for risk assessment.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The autostart function restarts the unit automatically after a power failure.
Engineers must assess whether unattended restart after power loss is acceptable for their process.
Source
“Autostart function for power failure”
Performance
The pressure pump delivers up to 14 l/min at 0.25 bar, while the suction side delivers up to 10.5 l/min at 0.17 bar.
Used to check whether the pump can supply the external circuit flow and head required by the application.