Peter Huber Kältemaschinenbau KISS 108A
Peter Huber Kältemaschinenbau KISS 108A — Thermostatic baths.
The KISS 108A is a 6-litre immersion circulator with a pressure/suction pump and 1.5–2.1 kW heater, operating from 25 to 100 °C (down to 15 °C only with an external cooling unit). It is the 6-litre member of the KISS series, distinguished by its fixed-speed pump and 200–240 V single- or two-phase supply. It serves as a temperature-control source for external circuits or open baths in laboratory processes.
Direct drop-in replacements
Peter Huber Kältemaschinenbau Mit Polycarbonatbaedern Bis 100C Series — Model Comparison & Selection
| Model | Controller | Interface | Max delivery flow rate, liter/minute | Pump connection | Weight, kg | Ambient temperature limit, °C |
|---|---|---|---|---|---|---|
| KISS 108A Current | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-104A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-106A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 5 | 5–40 |
| CC-108A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 6 | 5–40 |
| CC-110A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male | 6 | 5–40 |
| CC-112A | Pilot ONE | Ethernet, USB (Host u. D | 27 | M16x1 AG male (Option, # | 8 | 5–40 |
| CC-118A | Pilot ONE | Ethernet, USB (Host u. D | 22 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 104A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male | 5 | 5–40 |
| KISS 106A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 5 | 5–40 |
| KISS 110A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 6 | 5–40 |
| KISS 112A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
| KISS 118A | KISS | USB (Device), RS232 Inte | 14 | M16x1 AG male (Option, # | 8 | 5–40 |
Full specifications
| Alarm output | optic, acoustic |
| Bath volume | 6 l |
| Maximum delivery flow rate | 14 l/min (pressure pump max.; 0, 25 bar) |
| Pump connection | M16x1 AG male (Option, #19606) (Option, #19606) |
| Pump speed adjustable | no |
| Pump type | Pressure and suction pump, speed not variable |
| Temperature range | 25...100 °C with cooling unit: 15...100 °C · with water cooling: 20...100 °C |
| Temperature stability mfr | ±0, 05 K |
| Dimensions (W x D x H) mfr | 147 x 407 x 330 mm |
| Heating capacity mfr | 1, 5 - 2, 1 kW |
| Delivery capacity suction pump max. mfr | 10, 5 l/min (suction pump max.; 0, 17 bar) |
| Bath opening WxD/BD mfr | 130 x 210 / 150 mm |
| Filling volume min. mfr | 3, 5 l (min.) |
| Weight | 6 kg |
| Current consumption | 10 A (max.) |
| Power supply | 200-240V 1~/2~ 50/60Hz V, Hz |
| Controller | KISS |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Immersion circulator with a 6 l bath volume (minimum filling 3.5 l) and a bath opening of 130 × 210 mm, 150 mm deep.
- Fixed-speed pressure and suction pump: pressure side up to 14 l/min at 0.25 bar; suction side up to 10.5 l/min at 0.17 bar.
- Controller: KISS, with OLED display; interfaces USB (Device) and RS232; alarm output optical and acoustic.
- Safety class III / FL with adjustable overtemperature protection per DIN 12876.
- Within the KISS series, this is the 6-litre model; sibling CC-104A/106A/108A/110A differ in capacity and should be selected when a different bath volume is required.
Thermal behaviour and media
- Base temperature range 25–100 °C; 20–100 °C with water cooling; 15–100 °C with a refrigerator. The required low temperature dictates which cooling option must be ordered.
- Heating power is voltage-dependent: 2.1 kW at 240 V, 2.0 kW at 230 V, 1.8 kW at 220 V, 1.6 kW at 208 V, 1.5 kW at 200 V. Temperature stability is ±0.05 K.
- The plastic pump is not resistant to acetone or polyglycols; for freeze protection, water should be mixed with glysantine or ethylene glycol.
Installation and operation
- Supply: 200–240 V, 1~ or 2~, 50/60 Hz, max. current 10 A. Voltage and frequency tolerances are ±10% only if they do not act in opposite directions.
- External branch circuit protection according to UL 489 is mandatory. Single- and two-phase units are delivered with a power cable and country-specific plug that must be specified at order.
- Ambient temperature limit 5–40 °C; protection rating IP20; weight 6 kg; dimensions 147 × 407 × 330 mm.
- Autostart function restarts the unit automatically after a power failure; assess whether unattended restart is acceptable for the process.
- EMC classification EN 55011 Class A, Group 1 (industrial environment).
Integration & Operating Notes
The datasheet requires external branch circuit protection according to UL 489. The installer must provide this breaker; without it the installation is non-compliant. Size the circuit for the maximum current of 10 A at 200–240 V.
The ±10% voltage tolerance applies only as long as the frequency tolerance does not run in the opposite direction. Verify the site supply envelope before commissioning; combined deviations in opposite directions exceed the allowed operating range.
Single- and two-phase devices (100 V to 240 V) are delivered with a power cable and a country-specific plug. The plug type must be specified when ordering, otherwise the unit may not connect to the site outlet.
The base range is 25–100 °C. Reaching 20 °C requires water cooling; reaching 15 °C requires a refrigerator. The cooling method must be selected to match the required process temperature before ordering.
The pump is not resistant to acetone or polyglycols (depending on the manufacturer). Avoid these media or order a more resistant plastic to prevent premature pump failure.
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than used pure, to avoid freezing and damage at low temperatures.
The bath has a 6 l volume with a minimum filling capacity of 3.5 l. Underfilling risks pump cavitation and poor temperature control; overfilling risks overflow.
The unit must only be operated within an ambient temperature of 5 °C to 40 °C. Operation outside this range affects reliability and warranty compliance.
The autostart function restarts the unit automatically after a power failure. Engineers must assess whether unattended restart is acceptable for their process, particularly where a hot bath restart could affect samples or safety.
The pressure pump delivers up to 14 l/min at 0.25 bar; the suction side delivers up to 10.5 l/min at 0.17 bar. Check that the external circuit flow and head requirements fall within these limits.
Heating power ranges from 1.5 kW at 200 V to 2.1 kW at 240 V. Heat-up time must be matched to the actual site voltage.
The unit is classified as EN 55011 Class A, Group 1 for electromagnetic disturbance. Class A equipment is intended for industrial environments and may not be suitable for residential EMC environments.
Selection & Integration Q&A
Choose the KISS 108A when a 6 l bath volume with a fixed-speed pressure/suction pump and 25–100 °C base range fits the process. Select a sibling CC model when a different bath capacity is required; the 108A is the 6-litre member of the series.
Provide a 200–240 V, 1~ or 2~, 50/60 Hz supply with a UL 489 listed branch breaker, sized for 10 A max. The pressure pump supplies up to 14 l/min at 0.25 bar and suction up to 10.5 l/min at 0.17 bar; match external circuit head and flow to these limits.
The base delivery does not include the country-specific plug (must be specified at order), external cooling water or refrigerator for sub-25 °C operation, or freeze-protection additives (glysantine or ethylene glycol). The M16x1 AG male pump connection is an option (#19606).