Product introduction and key features of KISS E circulator
Specp. 2
Technical specifications according to DIN 12876 standard
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
KISS E
Operating temperature range
25...200 °C
with water cooling
20...200 °C
with refrigerator
-30...200 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
digital
Internal temperature sensor
Pt100
Interface digital
USB (Device), RS232
Interface
Alarm message
optic, acoustic
Safety classification
III / FL
Heating power at 240V
2,1 kW
Heating power at 230V
2 kW
Heating power at 220V
1,8 kW
Heating power at 208V
1,6 kW
Heating power at 200V
1,5 kW
max. delivery
14 l/min
max. delivery pressure
0,25 bar
max. delivery (suction)
10,5 l/min
max. delivery pressure (suction)
0,17 bar
Pump connenction (optional)
M16x1 male
Immersion depth
150 mm
Overall dimensions WxDxH **
132x163x312 mm
Net weight
4 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
Power input
2,1 kW
max. current
10 A
min. Fuse
10A
max. Fuse
16A
Pressure equipment category
N/A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
S384165
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Selection
The operating temperature range depends on the cooling method, so the required range must be matched to whether water cooling or a refrigerator is used.
Determines whether the unit can cover the process temperature window without additional cooling equipment.
Source
“Operating temperature range 25...200 °C with water cooling 20...200 °C with refrigerator -30...200 °C”
Installation
The unit requires a 200-240 V single- or two-phase supply at 50/60 Hz and draws up to 10 A, so the site must provide a suitable circuit.
Needed to confirm site power availability and correct branch circuit design before installation.
Source
“Power supply requirement 200-240V 1~/2~ 50/60Hz”
Installation
The minimum fuse must be 10 A and the maximum fuse 16 A, which constrains the upstream protection device selection.
Prevents nuisance tripping or inadequate protection of the supply circuit.
Source
“min. Fuse 10A max. Fuse 16A”
Installation
External branch circuit protection according to UL 489 is required, so the installer must provide a listed branch breaker.
Affects electrical installation compliance and hardware selection for the branch circuit.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Voltage and frequency tolerances are ±10% only if they do not run in opposite directions, so supply quality must be checked before commissioning.
Avoids operating the unit outside its allowed supply envelope and prevents damage or malfunction.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.
Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides correct fluid preparation to avoid freezing and maintain safe operation.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety
The unit has adjustable overtemperature protection and safety class III/FL, which is relevant for unattended operation and risk assessment.
Supports safety classification and configuration of over-temperature limits for the process.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
An autostart function for power failure is included, so the behavior after a power interruption must be considered for the process.
Affects restart strategy and risk of unexpected heating or pumping after power returns.
Source
“Autostart function for power failure”
Installation
The degree of protection is IP20, so the unit must be installed in a dry, protected environment.
Restricts installation location and protects against dust and water ingress.
Source
“Degree of Protection IP20”
Installation
The ambient temperature must remain between 5 °C and 40 °C for proper operation.
Ensures the unit is not installed in an environment outside its rated operating conditions.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
The pump delivers up to 14 l/min with a maximum delivery pressure of 0.25 bar, which limits the external circuit resistance.
Used to check whether the pump can overcome the pressure drop of the external loop.