Product name and cover page for the Minichiller unit
Specp. 2
Technical data table according to DIN 12876
p. 2table, 10 rows
Technical data according to DIN 12876
Parameter
Value
max. cooling water pressure
6 bar
min. filling capacity
2,8 l
expansion tank
2,2 l
Overall dimensions WxDxH **
280x490x424 mm
Power supply requirement
208-240V 1~/2~ 50/60Hz
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/23
** Please respect space requirements. See operating conditions at www.huber-online.com
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is water-cooled, so a cooling water supply with at least 3 bar differential pressure and 5–20 °C inlet temperature must be available at the installation site.
Without adequate cooling water pressure and temperature the specified cooling capacity cannot be achieved.
Source
“min. cooling water differential pressure”
Installation
Cooling water connections are G1/2 male, so mating hoses/fittings must be G1/2 female.
Ensures correct mechanical connection and avoids adapters or leaks.
Source
“Cooling water connection”
Installation
The process pump connection is M16x1 male, so process hoses must match this thread.
Prevents mismatch with process loop fittings and ensures leak-free circulation.
Source
“Pump connection”
Operation
The refrigerant is R-290 (A3, flammable), so ignition sources and ventilation must be considered in the installation area.
A3 refrigerant requires attention to safety classification and local flammable-refrigerant regulations.
Source
“R-290 (A3, H220)”
Operation
Cooling water inlet must stay between 5 °C and 20 °C; below the dew point the cooling water pipes inside the unit must be insulated.
Operating outside this range or without insulation causes condensation and reduced performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.
Sizing must account for actual site cooling water temperature, not only the 15 °C reference.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection
The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant pump specified.
Prevents chemical attack and premature pump failure when selecting the heat transfer fluid.
Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than using pure water below 0 °C.
Avoids freezing and damage in the cooling circuit when operating at sub-zero temperatures.
Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.
Site power quality must be checked; some voltage/frequency combinations are outside the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is IP20, so it must be installed in a dry, protected location and not exposed to water spray.
IP20 limits installation to indoor, non-wet environments.
Source
“Degree of Protection”
Installation
Ambient temperature must be between 5 °C and 40 °C for proper operation.
Installation room conditions affect cooling performance and reliability.
Source
“max. ambient temperature”
Installation
The unit requires 208–240 V single/two-phase 50/60 Hz supply; three-phase devices below 63 A are supplied with cable but without plug.
Electrical preparation must include a suitable plug and supply matching the delivered cable configuration.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance
Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water should be used.
Guides water quality and treatment to avoid corrosion and scaling.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms built-in safety limiting for unattended operation and selection against process risk.
Source
“With adjustable overtemperature protection according to DIN 12876.”