Peter Huber Kältemaschinenbau Minichiller 800w-H OLÉ Kompakter
Peter Huber Kältemaschinenbau Minichiller 800w-H OLÉ Kompakter — Recirculating chillers.
The Minichiller 800w-H OLÉ Kompakter is a water-cooled recirculating chiller with a pressure-and-suction pump and OLÉ controller, covering -20 to 100 °C. Its G1/2 cooling-water connections distinguish it from the 1000-series siblings. It serves as a compact process cooling source for closed loops requiring both pressure and suction operation.
Peter Huber Kältemaschinenbau Minichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Safety class | Temperature range, °C | Interface | Power supply |
|---|---|---|---|---|---|---|
| Minichiller 800w-H OLÉ Kompakter Current | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1000w Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 1200w-H Kompakter | — | Pilot ONE | III / FL | -20–100 | — | — |
| Minichiller 1200w-H OLÉ Kompakter | — | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 280 OLÉ Kompakter | 53 | OLÉ | I / NFL | -5–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300 OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w-H OLÉ Kompakter | 51 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 300w OLÉ Kompakter | 53 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 220-240V 1~/2~ 50/60Hz |
| Minichiller 600-H OLÉ Kompakter | 56 | OLÉ | III / FL | -20–100 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 600 Kompakter | — | Pilot ONE | I / NFL | -20–40 | Ethernet, USB (Host u. D | 110-120V 1~ 60Hz |
| Minichiller 600 OLÉ Kompakter | 56 | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 110-120V 1~ 60Hz |
| Minichiller 600w OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
| Minichiller 800 OLÉ Kompakter | — | OLÉ | I / NFL | -20–40 | USB (Device), RS232 Inte | 208-240V 1~/2~ 50/60Hz |
Full specifications
| Pump connection | M16x1 AG male |
| Pump speed adjustable | no |
| Pump type | Pressure- and suction pump |
| Temperature range | -20...100 °C |
| Power supply | 208-240V 1~/2~ 50/60Hz |
| Controller | OLÉ |
| Display resolution | 0, 1 K |
| Digital interface | USB (Device), RS232 Interface |
| Temperature display | OLED Display |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 24 l/h (at 15°C, flow -20°C) at 15°C, flow -10°C: 30 l/h · at 15°C, flow 0°C: 36 l/h · at 15°C, flow 15°C: 39 l/h |
| Cooling water differential pressure | 3 bar (min.) |
| Cooling Water Pressure | 6 bar (max.) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled refrigeration circuit using R-290 (A3, flammable) refrigerant; cooling water connections are G1/2 male, process pump connection is M16x1 male.
- Pressure-and-suction pump with fixed speed (not adjustable); plastic pump body limits media compatibility.
- OLÉ controller with OLED display, 0.1 K display resolution, USB (Device) and RS232 interfaces.
- Safety class III/FL with adjustable overtemperature protection per DIN 12876; IP20 enclosure.
- Within the Minichiller series, this 800w-H OLÉ variant sits below the 1000-series siblings (Minichiller 1000-H, 1000-H OLÉ, 1000 OLÉ, 1000w-H) in cooling capacity class; the 'w' denotes water cooling, 'H' the pressure-and-suction pump configuration, and 'OLÉ' the controller variant.
Behavior under load and media
- Temperature range -20 to 100 °C; cooling water inlet must remain between 5 and 20 °C, with minimum differential pressure 3 bar and maximum pressure 6 bar.
- Cooling water consumption varies with process temperature at 15 °C cooling water: 39 l/h at 15 °C process, 36 l/h at 0 °C, 30 l/h at -10 °C, 24 l/h at -20 °C.
- Cooling capacity decreases as cooling water temperature rises, and required cooling water flow rate may increase; sizing must use actual site water temperature, not only the 15 °C reference.
- The plastic pump is not resistant to acetone or polyglycols; these media require a different pump specification.
- For sub-zero operation, water should be mixed with glysantine or ethylene glycol for freeze protection.
Installation, control and operation
- Ambient temperature 5–40 °C; unit must not be installed in an ATEX zone; a drip tray for condensation water and thermal fluid must be provided below the unit.
- Electrical lines directly below the unit must be protected against liquid ingress due to condensation.
- Power supply 208–240 V 1~/2~ 50/60 Hz; voltage tolerance ±5% only with simultaneous frequency tolerance ±2%, and the combination -5% voltage with +2% frequency is not permitted.
- Three-phase devices below 63 A are supplied with cable but without plug.
- Do not install valves or quick-release couplings in feed/discharge lines to the application.
- Do not move the unit while operating due to burn, freezing and thermal fluid spill hazards.
- Only drinking water or filtered river/sea water may be used as cooling water; sea water must not be used. Anti-corrosion agents are mandatory if salts (chlorides, bromide) are added to the water circuit.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water should be used.
Integration & Operating Notes
Operating outside this range or below the dew point requires insulation of the cooling water pipes inside the unit; otherwise condensation and reduced performance occur. The datasheet explicitly designs the unit for 5–20 °C cooling water intake.
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed. Site power quality must be checked against this envelope before connection.
The unit requires 208–240 V single/two-phase 50/60 Hz supply; three-phase devices with current consumption less than 63 A are supplied with cable but without plug, so a suitable plug must be provided on site.
Operation in a hazardous area can cause explosion. The unit must not be installed or started up within an ATEX zone, and the refrigerant R-290 (A3, flammable) requires attention to ignition sources and ventilation in the installation area.
The responsible body must install a condensation water / thermal fluid drip tray below the temperature control unit to contain leaks. Electrical lines directly below the unit must also be protected against liquid ingress because condensation can form.
Do not install valves or quick-release couplings in the feed/discharge lines from the temperature control unit to the application and from the application to the temperature control unit, to avoid overpressure damage, especially to glass apparatus.
Only drinking water or filtered river/sea water may be used as cooling water; sea water must not be used. Anti-corrosion agents are mandatory if salts such as chlorides or bromide are added to the water circuit.
The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer), so these media must be avoided or a more resistant pump specified. For freeze protection below 0 °C, water should be mixed with glysantine or ethylene glycol.
Observe the amount of refrigerant contained (see data sheet/rating plate) and the room size when installing the unit to avoid exceeding the refrigerant limit value per m³ of ambient air; additional safety measures may be required.
Selection & Integration Q&A
Choose the 800w-H OLÉ when the required cooling capacity fits the 800 W class and you need the pressure-and-suction pump with OLÉ controller. The 1000-series siblings offer a higher capacity class; select them if your load exceeds the 800-class capability. The 'w' denotes water cooling, so a cooling water supply is mandatory for all these variants.
Provide cooling water at 5–20 °C with at least 3 bar differential pressure (max 6 bar); connections are G1/2 male, so mating hoses/fittings must be G1/2 female. The process pump connection is M16x1 male, so process hoses must match this thread. Only drinking water or filtered river water may be used; sea water is prohibited.
Install indoors at 5–40 °C ambient, not in an ATEX zone, with a drip tray below the unit and protection of electrical lines against condensation. The unit is IP20 and must not be exposed to water spray. Do not move it while operating. Power supply is 208–240 V 1~/2~ 50/60 Hz with interdependent voltage/frequency tolerances.
The base delivery includes cable but no plug for three-phase units below 63 A. A drip tray must be provided on site. Anti-corrosion agents are mandatory if salts are added to the water circuit. For sub-zero operation, glysantine or ethylene glycol must be added for freeze protection. Check refrigerant charge against room volume per the manual.