Petite Fleur w — Technical Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of the water bath
  2. Spec pp. 1-2
    Technical data, dimensions, and performance parameters per DIN 12876
  3. Spec p. 2
    Detailed operating specifications and compliance values
  4. p. 2 table, 17 rows
  5. p. 1 table, 32 rows

Technical data according to DIN 12876

ParameterValue
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C34 l/h
Consumption at water 15°C, flow -20°C26 l/h
min. cooling water differential pressure2 bar
max. cooling water pressure6 bar
min. filling capacity1,5 l
Filling capacity expansion tank2 l
Overall dimensions WxDxH **260x504x450 mm
Net weight47 kg
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current11 A
Fuse16 A
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S544847

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-40...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power at 240V2 kW
Heating power at 230V1,8 kW
Heating power at 220V1,6 kW
Cooling power with Thermooil at 200°C0,48 kW
Cooling power with Thermooil at 150°C0,48 kW
Cooling power with Thermooil at 100°C0,48 kW
Cooling power with Thermooil at 50°C0,48 kW
Cooling power with Thermooil at 20°C0,48 kW
Cooling power with Ethanol at 0°C0,45 kW
Cooling power with Ethanol at -20°C0,27 kW
Cooling power with Ethanol at -30°C0,16 kW
Cooling power with Ethanol at -40°C0,04 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,06 kg
CO2 equivalent0,0 t
UN-numberUN 2857
Circulation pump: max. delivery25 l/min
Circulation pump: max. delivery pressure0,9 bar
Pump connectionM16x1 male

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires cooling water with a differential pressure of at least 2 bar and no more than 6 bar at the connection.

To ensure adequate cooling water flow and avoid overpressure damage to the water-cooled condenser.

Source
“min. cooling water differential pressure 2 bar”
Installation

The cooling water intake temperature must be between 5°C and 20°C; below the dew point, cooling water pipes inside the unit must be insulated.

To prevent condensation, corrosion, and ensure rated cooling performance.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

The power supply must be 220-240 V, 1~/2~, 50/60 Hz with a maximum current of 11 A and a 16 A fuse.

To size the electrical supply, protection, and cabling correctly.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but a -5% voltage combined with +2% frequency is not allowed.

To avoid operating the compressor and pump outside safe limits and voiding warranty.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-290 (propane, A3, H220) with a charge of 0.06 kg, requiring appropriate explosion protection measures.

To assess flammability risk, ventilation requirements, and compliance with local refrigerant regulations.

Source
“Refrigerant (ASHRAE, GHS) R-290 (A3, H220)”
Operation

The unit is a hydraulically sealed circulator with an atmospheric open expansion tank, suitable only for externally closed applications.

To avoid using the unit with open baths or unsealed loops that would cause fluid loss or contamination.

Source
“For externally closed applications.”
Operation

The expansion tank is thermally passive, preventing HTF vapour formation and moisture absorption.

To ensure stable heat transfer fluid properties and reduce maintenance.

Source
“No HTF vapour and no moisture absorption because the expansion tank is thermally passive.”
Performance

Cooling power with ethanol drops from 0.45 kW at 0°C to 0.04 kW at -40°C, so low-temperature operation has very limited heat removal capacity.

To correctly size the cooling load at the lowest required operating temperature.

Source
“at -40°C 0,04 kW”
Performance

The circulation pump delivers a maximum of 25 l/min at 0.9 bar, which limits the achievable flow in high-pressure-drop loops.

To verify that the external circuit pressure drop is compatible with the pump curve.

Source
“max. delivery pressure 0,9 bar”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

To select a suitable heat transfer fluid that remains pumpable at the lowest operating temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The minimum filling capacity is 1.5 l and the expansion tank capacity is 2 l.

To determine the required fluid volume and expansion volume for the external loop.

Source
“min. filling capacity 1,5 l”
Installation

The ambient temperature must be between 5°C and 40°C and the degree of protection is IP20.

To ensure adequate ventilation and avoid overheating or ingress of contaminants.

Source
“max. ambient temperature 40 °C”
Compliance

The equipment is classified as EN55011 Class A, Group 1, so it may cause radio interference in residential areas.

To plan EMC measures and avoid use in sensitive residential or laboratory environments without filtering.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Three-phase devices with current consumption less than 63 A are supplied with cable but without a plug.

To arrange for a suitable plug and electrical connection on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”