Peter Huber Kältemaschinenbau Petite Fleur w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Petite Fleur w Dynamisches

Peter Huber Kältemaschinenbau Petite Fleur w Dynamisches — Temperature control units.

Hydraulically sealed circulator with a thermally passive expansion tank, rated for externally closed loops from -40 to 200 °C. The water-cooled 'w' variant uses R-290 (0.06 kg) and requires 2–6 bar cooling-water differential pressure. It suits precise temperature control where low-temperature cooling capacity is modest (0.04 kW at -40 °C).

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Petite Fleur w Dynamisches Current —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Control output POKO ONE
Expansion volume 2 l
Pump connection M16x1 AG male
Pump speed adjustable Yes
Pump type Pressure pump, variable speed
External sensor connection Pt100
Temperature range -40...200 °C
Dimensions, Weight & Mechanical Design
Weight 47 kg
Electrical, Power & Drive Specifications
Current consumption 11 A (max)
Fuse rating 16 A
Power supply 220-240V 1~/2~ 50/60Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Refrigerant R-290 (A3, H220) (1st stage)
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water consumption 26 л/час (at 15°C, flow -20°C)
at 15°C, flow 0°C: 34 l/h
Cooling water differential pressure 2 bar (min)
Cooling Water Pressure 6 bar (max)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, connection and operation

Integration & Operating Notes

Cooling & Media
Cooling-water differential pressure must stay within 2–6 bar

Below 2 bar the condenser receives insufficient flow and cooling performance collapses; above 6 bar the water-cooled circuit and fittings are overstressed. Verify the site supply at the G1/2 male connection before commissioning.

Petite Fleur w — Technical Datasheet, Specifications
Cooling & Media
Cooling-water intake temperature limited to 5–20 °C

Operation outside this window risks condensation inside the unit or loss of rated cooling capacity. If the cooling water is below the dew point, pipes inside the unit must be insulated to prevent condensation.

Petite Fleur w — Technical Datasheet, Specifications
Electrical
Do not combine -5 % voltage with +2 % frequency

The supply tolerances are ±5 % voltage and ±2 % frequency, but the combination of -5 % voltage and +2 % frequency is explicitly not allowed because it can overload the compressor and pump. Check the site supply against both limits simultaneously.

Petite Fleur w — Technical Datasheet, Specifications
Electrical
Three-phase units arrive with cable but no plug

Devices with current consumption below 63 A are supplied with a cable only, so the site must provide a suitable plug and a 16 A fused supply. Power cables longer than 3 m must not be used so the unit can always be disconnected from the mains.

Petite Fleur w — Technical Datasheet, Specifications; Grande Fleur Operation Manual, Installation
Mounting
Install a drip tray below the unit before commissioning

The responsible body must install a condensation-water / thermal-fluid drip tray below the temperature control unit. Electrical lines directly below a water-cooled unit must be protected against liquid ingress from condensation.

Grande Fleur Operation Manual, Installation
Mounting
Do not install or start in an ATEX zone without an Ex px cabinet

The unit must not be installed or started in an ATEX zone without an Ex px cabinet. For natural-refrigerant units, if the charge exceeds 8 g per m³ of room air, a calibrated gas warning sensor is mandatory; units with more than 1 kg refrigerant must not be installed below ground level.

Grande Fleur Operation Manual, Safety and Installation
Process Limits
Connect circulation flow to the lower application port

The circulation flow from the unit must attach to the lower connection point of the application and the return to the higher connection point, so that air bubbles can vent. Reversing these connections traps air and prevents proper circulation.

Grande Fleur Operation Manual, Installation
Cooling & Media
Shut off cooling water even for overnight downtime

The cooling water supply must be shut off even for short downtimes such as overnight to avoid water damage and condensation risks. Electrical lines directly below the unit must be protected against liquid ingress.

Grande Fleur Operation Manual, Operation

Selection & Integration Q&A

When should I choose the Petite Fleur w over the larger Grande Fleur variants?

Pick the Petite Fleur w for compact, externally closed loops needing -40…200 °C with modest cooling load (0.04 kW at -40 °C). Choose a Grande Fleur sibling when higher cooling capacity or a different cooling configuration is required; the Petite Fleur w is the water-cooled compact unit with 0.06 kg R-290.

What cooling-water supply and electrical connection must be provided on site?

Provide cooling water at 5–20 °C with 2–6 bar differential pressure at the G1/2 male connections (34 l/h at 0 °C flow, 26 l/h at -20 °C). Electrical: 220–240 V, 1~/2~, 50/60 Hz, max. 11 A, 16 A fuse; three-phase units come with cable but no plug.

What mounting and ambient limits apply to this unit?

Ambient 5–40 °C, IP20, EN55011 Class A Group 1. Install a drip tray below the unit, protect cables below against condensation, do not install in an ATEX zone without an Ex px cabinet, and keep power cables ≤3 m for disconnection.

What maintenance and consumables are not included in the base delivery?

The base delivery excludes the plug (three-phase units ship with cable only), the drip tray, and the heat-transfer fluid. The fluid must be selected for ≤50 mm²/s at the lowest temperature; never mix fluid types and rinse the circuit when changing fluid.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Grande Fleur Operation Manual
Manual ·3,1 МБ ·88 pp.
View PDF
PDF
Petite Fleur w — Technical Datasheet
Datasheet ·102,3 КБ ·2 pp.
View PDF