The unit uses Peltier technology and is refrigerant-free, which matters for environmental compliance and where refrigerant handling is restricted.
Refrigerant-free design avoids refrigerant logistics, leak checks and disposal obligations during selection and operation.
Source
“Refrigerated Heating Circulator with the state of the art Peltier technology, i.e. Free of refrigerants and therefore completely environmentally friendly.”
Performance
Cooling capacity drops sharply at low temperatures, so the 4 °C point delivers only 0.03 kW and may not hold a load.
Engineers must verify heat load versus available cooling capacity at the lowest required set point.
Source
“at 4°C 0,03 kW”
Performance
Cooling capacity is rated at 20 °C ambient and falls if ambient rises, so enclosure ventilation and room temperature must be controlled.
Installation location and ambient conditions directly affect achievable cooling performance.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The device accepts 100–240 V single- or two-phase supply at 50/60 Hz, so it can be connected to standard mains without a transformer.
Confirms mains compatibility and that no site-specific voltage conversion is needed.
Source
“100-240V 1~/2~ 50/60Hz”
Installation
The minimum fuse is 10 A and maximum 16 A, so the upstream circuit breaker must be sized within this window.
Protective device coordination must respect the manufacturer's fuse limits to avoid nuisance trips or inadequate protection.
Source
“min. Fuse 10A”
Installation
Single- and two-phase units ship with a power cable and country-specific plug that must be specified at order time.
Ordering the correct plug avoids delays and rework at commissioning.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
The unit is IP20, so it must be installed in a dry, protected location away from water spray and dust.
IP20 limits where the circulator can be safely placed in the lab.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C for correct operation.
Operating outside this range risks performance loss or equipment damage.
Source
“min. ambient temperature 5 °C”
Operation
The autostart function restarts the unit after a power failure, which may be undesirable for unattended or sensitive processes.
Engineers must decide whether automatic restart is acceptable for their process safety concept.
Source
“Autostart function for power failure”
Operation
The pump delivers up to 1.85 l/min at up to 0.95 bar, which constrains external loop pressure drop and hose sizing.
External circuit resistance must be checked against available pump head and flow.
Source
“max. delivery pressure 0,95 bar”
Operation
The internal cooling circuit holds only 0.1–0.2 l, so fluid volume and thermal inertia are small.
Small internal volume affects fluid selection, expansion and response time.