RotaCool Umwälzkühler für Rotationsverdampfer

Contents

  1. Overview p. 1
    Product introduction and general description of the RotaCool unit
  2. Spec p. 2
    Technical data according to DIN 12876

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is designed for conventional rotary evaporators in terms of cooling capacity and circulation quantity, so it should be matched to that application.

Confirms the intended application and helps decide if it fits a rotary evaporator setup.

Source
“The Rotacool, in space-saving L-design, is especially adjusted to the requirements of conventional rotary evaporators concerning cooling capacity and circulation quantity.”
Selection

Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient temperatures, so selection must account for actual ambient conditions.

Prevents oversizing assumptions when the installation room is warmer than 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 208–240 V single/two-phase 50/60 Hz supply with a minimum 10 A fuse and maximum 16 A fuse.

Determines electrical supply and protection sizing before installation.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz”
Installation

Voltage and frequency tolerances must be met simultaneously; for example, -5% voltage with +2% frequency is not allowed.

Avoids installing on a supply that appears within individual tolerances but is outside the combined allowed envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified when ordering.

Ensures the correct plug is ordered and avoids delays or rework at commissioning.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water ingress.

Restricts installation environment and prevents placing the unit where splashes or dust are present.

Source
“Degree of Protection IP20”
Operation

Ambient temperature must stay between 5 °C and 40 °C for safe and rated operation.

Defines the allowable room temperature range for reliable operation.

Source
“min. ambient temperature 5 °C”
Operation

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which affects handling and room requirements.

Alerts the engineer to flammability classification and related safety measures.

Source
“R-290 (A3, H220)”
Operation

The refrigeration machine is air-cooled and natural, so adequate ventilation around the unit is required.

Prevents overheating and loss of cooling capacity due to restricted airflow.

Source
“air-cooled, natural”
Performance

The pressure pump delivers up to 14 l/min at 0.25 bar head, which must match the external circuit resistance.

Allows checking whether the pump can overcome the pressure drop of the connected loop.

Source
“max. delivery 14 l/min”
Performance

The suction pump delivers up to 10.5 l/min at 0.17 bar suction head, relevant for closed-loop suction operation.

Helps verify suction-side capability when the application uses the suction pump.

Source
“max. delivery (suction) 10,5 l/min”
Installation

Pump connections are M16x1 male, so mating fittings must be M16x1 female.

Ensures correct hose or pipe adapters are procured before installation.

Source
“Pump connection M16x1 male”
Selection

The minimum filling capacity is 1.5 l and expansion volume is 1 l, which sets the minimum bath/loop volume.

Prevents underfilling and ensures enough expansion volume for the thermal fluid.

Source
“min. filling capacity 1,5 l”
Compliance

The equipment is EMC Class A, Group 1, which may require additional measures in residential environments.

Informs EMC planning and possible restrictions for use in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”