TC100E-F Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the TC100E-F circulator
  2. Spec p. 2
    Technical data according to DIN 12876 for the unit
  3. p. 1 table, 39 rows

Technical data according to DIN 12876

ParameterValue
Order-No.3005.0115.99
ModelTC100E-F
Operating temperature range-100...40 °C
Temperature stability at -10°C0,5 K
temperature set point / displaydigital
Temperature indicationdigital
Sensor external connectionPt100
Cooling power
at 0°C0,16 kW
at -20°C0,15 kW
at -30°C0,14 kW
at -50°C0,13 kW
at -60°C0,12 kW
at -80°C0,12 kW
at -90°C0,07 kW
at -100°C0,01 kW
Safety classificationI / NFL
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,26 kg
CO2 equivalent0,6 t
Refrigerant 2nd stage (ASHRAE, GHS)R-1150 (A3, H220)
Global Warming Potential (GWP)4
Refrigerant quantity 2nd stage0,056 kg
CO2-Äquivalent 2.Stufe0,0 t
Nominal diameter probe13 mm
Length of probe900 mm
Length flexible connection1150 mm
Overall dimensions WxDxH **295x500x570 mm
Net weight61 kg
sound pressure level +/- 4 dB(A)55 dB(A)
Power supply requirement110-120V 1~ 60Hz
max. current11 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.S433255

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit's operating temperature range is -100 to 40 °C, so it must be matched to the process temperature window.

Defines whether the cooler can cover the required process temperatures before selection.

Source
“Operating temperature range -100...40 °C”
Performance

Cooling capacity drops sharply at the cold end (0.01 kW at -100 °C), so low-temperature duty needs careful capacity sizing.

Prevents oversizing assumptions and ensures the process heat load is removable at the coldest set point.

Source
“at -100°C 0,01 kW”
Installation

The unit requires a 110-120 V 1~ 60 Hz supply with up to 11 A and a 16 A fuse.

Determines site power wiring, breaker sizing and plug/circuit compatibility.

Source
“Power supply requirement 110-120V 1~ 60Hz max. current 11 A Fuse 16 A”
Installation

Ambient temperature must stay between 5 °C and 40 °C for correct operation.

Room conditions outside this band can degrade cooling performance or prevent proper operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may fall if ambient rises.

Capacity margins must account for actual installation room temperature, not just nominal data.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.

Site power quality must be checked; some combined deviations are explicitly prohibited.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The refrigerant is R-452A (A1, H280) with GWP 2141 and 0.26 kg charge, giving 0.6 t CO2 equivalent.

Affects refrigerant handling, leak response and environmental reporting obligations.

Source
“R-452A (A1, H280)”
Safety

The second-stage refrigerant is R-1150 (A3, H220), a flammable refrigerant requiring appropriate precautions.

Flammable refrigerant classification influences installation location, ventilation and service safety.

Source
“R-1150 (A3, H220)”
Installation

The probe is 13 mm diameter and 900 mm long with a 1150 mm flexible connection, constraining vessel geometry.

Must be verified against the vessel opening, immersion depth and routing space before mounting.

Source
“Nominal diameter probe 13 mm Length of probe 900 mm Length flexible connection 1150 mm”
Installation

The unit is 295x500x570 mm and weighs 61 kg, requiring adequate floor space and handling provision.

Needed for layout planning, access and safe lifting/positioning during installation.

Source
“Overall dimensions WxDxH ** 295x500x570 mm Net weight 61 kg”
Operation

The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit.

Explains control behavior and why the compressor is not cycled on/off during normal operation.

Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Compliance

The equipment is EMC Class A, Group 1, which is intended for non-domestic environments.

Class A equipment may cause interference in residential areas and affects installation environment choice.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.

The correct plug type must be ordered up front to avoid site rework.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Compliance

The equipment complies with US-SNAP and applicable EU laws, with US-SNAP end-use being industrial process refrigeration.

Confirms regulatory suitability for industrial process refrigeration applications.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”