Peter Huber Kältemaschinenbau TC100E-F
Peter Huber Kältemaschinenbau TC100E-F — Immersion coolers with Noise level 55 dB.
The TC100E-F is an air-cooled two-stage immersion cooler with a 13 mm × 900 mm probe and 1150 mm flexible connection, reaching -100 to 40 °C. Within the TC100 series it is the external-sensor (E) variant with the coldest end point, distinguished from the TC100-F by its Pt100 external sensor connection. It is used to condense volatiles in cold traps and to hold low-temperature baths in industrial process refrigeration.
Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection
| Model | Noise level, dB | Controller | Power supply | Refrigerant | Safety class | Temperature range, °C |
|---|---|---|---|---|---|---|
| TC100E-F Current | 55 | MPC | 110-120V 1~ 60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| CT50 Single OLÉ | 59 | OLÉ | 220-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | III / FL | -50–50 |
| TC100 | 55 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC100-F | 58 | ON-OFF-Regler | 220-240V 1~/2~ 50/60Hz | R-1150 (A3, H220) | I / NFL | -100–40 |
| TC45 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45E-F | 54 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC45-F | 59 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -45–100 |
| TC50 | — | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E | — | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50E-F | 60 | MPC | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
| TC50-F | 60 | ON-OFF-Regler | 208-240V 1~/2~ 50/60Hz | R-290 (A3, H220) | I / NFL | -50–50 |
Full specifications
| Noise level | 55 dB(A) |
| Pump speed adjustable | no |
| Pump type | no pump |
| External sensor connection | Pt100 |
| Temperature range | -100...40 °C |
| Temperature stability mfr | ±0, 5 K |
| Alarm message mfr | optic, acoustic |
| Dimensions (W x D x H) mfr | 295 x 500 x 570 mm |
| Refrigeration machine mfr | air-cooled, CFC- and HCFC-free |
| Weight | 61 kg |
| Current consumption | 11 A (max.) |
| Fuse rating | 16 A |
| Power supply | 110-120V 1~ 60Hz |
| Controller | MPC |
| Temperature display | LED Display |
| Refrigerant | R-452A (A1, H280) (1st stage) 2nd stage: R-1150 (A3, H220) |
| Refrigerant quantity 1st stage mfr | 0, 26 kg (1st stage) |
| Refrigerant quantity 2nd stage mfr | 0, 056 kg (2nd stage) |
| Ambient temperature limit | 5...40 °C |
| Cooling capacity mfr | 0, 01 kW (at -100 °C) |
| Safety class | I / NFL |
Description
Construction and positioning
- Two-stage cascade refrigeration: R-452A (A1, H280) in the first stage, R-1150 (A3, H220) in the second stage; air-cooled, CFC- and HCFC-free machine.
- Probe geometry: 13 mm nominal diameter, 900 mm length, 1150 mm flexible connection; overall unit 295 × 500 × 570 mm, 61 kg.
- Series position: the TC100E-F shares the 55 dB(A) noise level of the TC100/TC100E but adds the external Pt100 sensor connection; the TC100-F is the same cold-end class without that external sensor interface, and the CT50 Single OLÉ is a smaller single-stage sibling.
- Control: MPC controller with LED display, external Pt100 input, optical and acoustic alarm; no pump and no pump-speed adjustment.
Behavior under load
- Working range -100 to 40 °C with stated temperature stability of ±0.5 K.
- Cooling capacity is 0.01 kW at -100 °C, rated at 20 °C room temperature; capacity falls as ambient rises, so the cold-end duty must be sized against the actual installation room temperature.
- The refrigeration unit runs continuously; temperature is regulated by a solenoid valve in the cooling circuit rather than by compressor cycling.
Installation, connection and operation
- Supply: 110-120 V 1~ 60 Hz, max. current 11 A, 16 A fuse; voltage tolerance ±5% with simultaneous frequency tolerance ±2%, not all combinations permitted.
- Ambient temperature 5...40 °C; EMC Class A, Group 1 (EN55011), intended for non-domestic environments.
- Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at order.
- Refrigerant charges: 0.26 kg R-452A (GWP 2141, 0.6 t CO2e) first stage; 0.056 kg R-1150 second stage; the A3 classification of R-1150 governs ventilation and service precautions.
- Compliance: US-SNAP and applicable EU laws; US-SNAP end-use is industrial process refrigeration.
Integration & Operating Notes
The process heat load at the coldest set point must be below the stated 0.01 kW, and the figure is valid at 20 °C room temperature; higher ambient reduces it further, so size the cold-trap or bath duty with margin or accept a warmer operating point.
The circuit must be wired for the 11 A maximum current with a 16 A fuse; voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, and not all combinations of these deviations are permitted, so supply quality should be checked before commissioning.
The 13 mm diameter, 900 mm long probe and 1150 mm flexible connection constrain vessel opening, immersion depth and cable routing; the 61 kg unit also needs floor space and handling provision for its 295 × 500 × 570 mm footprint.
Single/two-phase devices in the 100-240 V class are supplied with a power cable and a country-specific plug that has to be stated when ordering; without that specification the site may need to rework the connection.
The second-stage refrigerant R-1150 is classified A3/H220, so installation location, ventilation and service access must account for a flammable refrigerant; the first stage R-452A (A1, H280) carries 0.6 t CO2 equivalent at 0.26 kg charge.
Operation outside the 5...40 °C ambient band degrades or prevents correct operation, and since cooling capacity is rated at 20 °C room temperature, a hot plant room effectively derates the cold-end performance.
The unit is classified Class A, Group 1 to EN55011, intended for non-domestic environments; in residential or light-industrial settings it may cause interference, which affects where the unit can be placed relative to sensitive electronics.
Selection & Integration Q&A
Pick the TC100E-F when you need the -100 °C cold end plus an external Pt100 sensor input for measuring at the cold trap or bath rather than at the unit. The TC100-F offers the same cold-end class without that external sensor interface; the CT50 Single OLÉ is a smaller single-stage sibling at 59 dB(A).
A 110-120 V 1~ 60 Hz supply rated for 11 A maximum with a 16 A fuse. Voltage tolerance is ±5% with simultaneous frequency tolerance ±2%, and not all combinations are permitted. The country-specific plug must be specified at order, as the cable is supplied with the unit.
Ambient must remain between 5 and 40 °C. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises, so the cold-end load must be sized against actual room conditions. The refrigeration unit runs continuously, with temperature regulated by a solenoid valve.
The power cable is supplied but the country-specific plug must be specified at order. No pump is fitted, so any circulation must be provided externally. Service planning must account for the flammable R-1150 second-stage charge (0.056 kg) and the R-452A first-stage charge (0.26 kg, 0.6 t CO2e).