Peter Huber Kältemaschinenbau TC100E-F
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau TC100E-F

Peter Huber Kältemaschinenbau TC100E-F — Immersion coolers with Noise level 55 dB.

The TC100E-F is an air-cooled two-stage immersion cooler with a 13 mm × 900 mm probe and 1150 mm flexible connection, reaching -100 to 40 °C. Within the TC100 series it is the external-sensor (E) variant with the coldest end point, distinguished from the TC100-F by its Pt100 external sensor connection. It is used to condense volatiles in cold traps and to hold low-temperature baths in industrial process refrigeration.

Noise level
55 dB

Peter Huber Kältemaschinenbau Eintauchkuehler Kuehlfallen Series — Model Comparison & Selection

Model Noise level, dBControllerPower supplyRefrigerantSafety classTemperature range, °C
TC100E-F Current 55MPC110-120V 1~ 60HzR-1150 (A3, H220)I / NFL-100–40
CT50 Single OLÉ 59OLÉ220-240V 1~/2~ 50/60HzR-290 (A3, H220)III / FL-50–50
TC100 55ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC100-F 58ON-OFF-Regler220-240V 1~/2~ 50/60HzR-1150 (A3, H220)I / NFL-100–40
TC45 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45E-F 54MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC45-F 59ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-45–100
TC50 —ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E —MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50E-F 60MPC208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50
TC50-F 60ON-OFF-Regler208-240V 1~/2~ 50/60HzR-290 (A3, H220)I / NFL-50–50

Full specifications

Core Performance & Functional Specifications
Noise level 55 dB(A)
Pump speed adjustable no
Pump type no pump
External sensor connection Pt100
Temperature range -100...40 °C
Temperature stability mfr ±0, 5 K
Alarm message mfr optic, acoustic
Dimensions (W x D x H) mfr 295 x 500 x 570 mm
Refrigeration machine mfr air-cooled, CFC- and HCFC-free
Dimensions, Weight & Mechanical Design
Weight 61 kg
Electrical, Power & Drive Specifications
Current consumption 11 A (max.)
Fuse rating 16 A
Power supply 110-120V 1~ 60Hz
Interfaces, Connections & Control
Controller MPC
Temperature display LED Display
Materials, Media & Process Compatibility
Refrigerant R-452A (A1, H280) (1st stage)
2nd stage: R-1150 (A3, H220)
Refrigerant quantity 1st stage mfr 0, 26 kg (1st stage)
Refrigerant quantity 2nd stage mfr 0, 056 kg (2nd stage)
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling capacity mfr 0, 01 kW (at -100 °C)
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and positioning

Behavior under load

Installation, connection and operation

Integration & Operating Notes

Process Limits
Cold-end capacity is only 0.01 kW at -100 °C

The process heat load at the coldest set point must be below the stated 0.01 kW, and the figure is valid at 20 °C room temperature; higher ambient reduces it further, so size the cold-trap or bath duty with margin or accept a warmer operating point.

TC100E-F Immersion Circulator Datasheet, Specifications
Electrical
16 A fuse and 11 A max current on 110-120 V 60 Hz

The circuit must be wired for the 11 A maximum current with a 16 A fuse; voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, and not all combinations of these deviations are permitted, so supply quality should be checked before commissioning.

TC100E-F Immersion Circulator Datasheet, Specifications
Mounting
Probe needs 900 mm immersion plus 1150 mm flexible routing

The 13 mm diameter, 900 mm long probe and 1150 mm flexible connection constrain vessel opening, immersion depth and cable routing; the 61 kg unit also needs floor space and handling provision for its 295 × 500 × 570 mm footprint.

TC100E-F Immersion Circulator Datasheet, Specifications
Electrical
Country-specific plug must be specified at order

Single/two-phase devices in the 100-240 V class are supplied with a power cable and a country-specific plug that has to be stated when ordering; without that specification the site may need to rework the connection.

TC100E-F Immersion Circulator Datasheet, Installation
Process Limits
Second stage uses flammable R-1150 (A3)

The second-stage refrigerant R-1150 is classified A3/H220, so installation location, ventilation and service access must account for a flammable refrigerant; the first stage R-452A (A1, H280) carries 0.6 t CO2 equivalent at 0.26 kg charge.

TC100E-F Immersion Circulator Datasheet, Safety
Mounting
Ambient must stay within 5...40 °C

Operation outside the 5...40 °C ambient band degrades or prevents correct operation, and since cooling capacity is rated at 20 °C room temperature, a hot plant room effectively derates the cold-end performance.

TC100E-F Immersion Circulator Datasheet, Specifications
Control & I/O
EMC Class A equipment for non-domestic environments

The unit is classified Class A, Group 1 to EN55011, intended for non-domestic environments; in residential or light-industrial settings it may cause interference, which affects where the unit can be placed relative to sensitive electronics.

TC100E-F Immersion Circulator Datasheet, Compliance

Selection & Integration Q&A

When should I choose the TC100E-F rather than the TC100-F or the CT50 Single OLÉ?

Pick the TC100E-F when you need the -100 °C cold end plus an external Pt100 sensor input for measuring at the cold trap or bath rather than at the unit. The TC100-F offers the same cold-end class without that external sensor interface; the CT50 Single OLÉ is a smaller single-stage sibling at 59 dB(A).

What supply and protection does the TC100E-F need at the installation site?

A 110-120 V 1~ 60 Hz supply rated for 11 A maximum with a 16 A fuse. Voltage tolerance is ±5% with simultaneous frequency tolerance ±2%, and not all combinations are permitted. The country-specific plug must be specified at order, as the cable is supplied with the unit.

What ambient and duty-cycle conditions apply to the TC100E-F?

Ambient must remain between 5 and 40 °C. Cooling capacity is rated at 20 °C room temperature and drops as ambient rises, so the cold-end load must be sized against actual room conditions. The refrigeration unit runs continuously, with temperature regulated by a solenoid valve.

What is not included in base delivery and what must be planned for service?

The power cable is supplied but the country-specific plug must be specified at order. No pump is fitted, so any circulation must be provided externally. Service planning must account for the flammable R-1150 second-stage charge (0.056 kg) and the R-452A first-stage charge (0.26 kg, 0.6 t CO2e).

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
TC100E-F Immersion Circulator Datasheet
Datasheet ·108,6 КБ ·2 pp.
View PDF