TC50E-F Immersion Circulator Datasheet

Contents

  1. Overview p. 1
    Product identification and cover page for TC50E-F circulator
  2. Spec p. 2
    Technical specifications according to DIN 12876 standard

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit covers a working range of -50 to 50 °C, so it suits low-temperature cooling tasks but not heating above 50 °C.

Confirms the process temperature window before specifying the immersion cooler.

Source
“Operating temperature range -50...50 °C”
Safety

The refrigerant is R-290 (A3, H220), a flammable natural refrigerant, which drives ventilation, ignition-source and leak-detection requirements.

Flammable A3 refrigerant affects room ventilation, electrical zoning and site safety approval.

Source
“R-290 (A3, H220)”
Installation

The unit must be operated between 5 °C and 40 °C ambient, so the installation location needs adequate room temperature control.

Ambient limits determine where the air-cooled unit can be placed and whether cooling capacity will hold.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is only 0.3 kW at 0 °C and drops to 0.2 kW at -30 °C, so heat loads must be sized against the actual operating point.

Prevents oversizing the process load based on a single nominal capacity figure.

Source
“at 0°C 0,3 kW at -20°C 0,26 kW at -30°C 0,2 kW”
Performance

The stated cooling capacities are valid only at 20 °C room temperature and may fall at higher ambient, so capacity margins must account for the real room.

Capacity derating with ambient temperature must be included in the thermal design.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The supply must be 208-240 V single or two-phase at 50/60 Hz with a maximum current of 3 A, and the circuit must be fused between 10 A and 16 A.

Defines the electrical supply, cable and protective device rating for the installation.

Source
“Power supply requirement 208-240V 1~/2~ 50/60Hz max. current 3 A min. Fuse 10A max. Fuse 16A”
Installation

Voltage and frequency tolerances are coupled: ±5 % voltage is allowed only with ±2 % frequency, and opposite-sign deviations are not permitted.

Prevents commissioning on a supply whose combined voltage/frequency deviation exceeds the EN60034-1 allowance.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.

Class A equipment restricts where it can be installed without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single and two-phase devices are delivered with a power cable and country-specific plug, which must be specified at order.

The plug type must be confirmed at ordering to avoid a non-matching mains connection on site.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The refrigeration unit runs continuously and temperature is controlled by a solenoid valve in the cooling circuit, so the compressor is not cycled for control.

Explains the control principle and that continuous compressor operation is normal, not a fault.

Source
“The refrigeration unit works continuously. The temperature control actuates a solenoid valve in the cooling circuit.”
Installation

The probe is 13 mm in diameter and 900 mm long with a 1150 mm flexible connection, which constrains how it can be inserted into the application.

Probe geometry and cable length determine whether the sensor reaches the measurement point.

Source
“Nominal diameter probe 13 mm Length of probe 900 mm Length flexible connection 1150 mm”
Installation

The unit measures 260 x 330 x 415 mm and weighs 25 kg, and space requirements must be respected at the installation site.

Footprint, weight and clearance requirements must be checked before positioning the air-cooled unit.

Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Safety

The unit is safety class I / NFL according to DIN 12876, which sets the protective-earth and over-temperature protection expectations.

Determines the required earthing and safety concept for the installation.

Source
“Safety classification I / NFL”
Compliance

The equipment is compliant with US-SNAP and applicable EU laws, with industrial process refrigeration as the US-SNAP end use.

Confirms regulatory acceptability of the R-290 refrigerant for the intended industrial use.

Source
“This equipment is compliant to US-SNAP and all applicable EU laws. The US-SNAP end-use for this equipment is the industrial process refrigeration.”