  1. Overview p. 1
    Product introduction and main features of the Unichiller 022-H
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S407048

Technical data according to DIN 12876 — Unichiller 022-H (Order-No.: 3010.0101.01)

ParameterValue
Operating temperature range-10...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Resolution of display0,1 K
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at ambient temperature +20°C
at 15°C2,2 kW
at 0°C1,6 kW
at -10°C1 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,58 kg
CO2 equivalent0,8 t
UN-numberUN 2857
Circulation pump:Immersion pump
max. delivery29 l/min
max. delivery pressure1 bar
Delivery at 0,5 bar21 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **460x590x743 mm
Net weight77 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply requirement220-240V 1~ 50Hz
max. current13 A
Fuse16 A

** Overall dimensions WxDxH — сноска в оригинале.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room conditions.

Cooling capacity is rated at 20°C ambient; higher ambient reduces available cooling and may require a larger model.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit requires a 220–240 V single-phase 50 Hz supply with a 16 A fuse and draws up to 13 A.

Electrical supply and protection must be sized accordingly to avoid nuisance tripping or undersized wiring.

Source
“Power supply requirement 220-240V 1~ 50Hz max. current 13 A Fuse 16 A”
Installation

Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency is not allowed.

Ensures the site power quality is within the allowed combined tolerance before commissioning.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are delivered with a country-specific plug that must be specified at ordering.

Prevents delivery of the wrong plug type and avoids rework at installation.

Source
“with power cable and country-specific plug (please specify when ordering)”
Operation

The plastic pump is not resistant to acetone or polyglycols, so these fluids must be avoided or a more resistant pump specified.

Using incompatible fluids can damage the pump and cause downtime; select alternative pump material if needed.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Prevents freezing and damage when operating near or below 0°C.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 0.58 kg (0.8 t CO2e).

Informs refrigerant handling, leak response, and environmental reporting requirements.

Source
“R-449A (A1, H280)”
Safety

The unit is classified IP20, so it must be installed in a dry location protected from water ingress.

IP20 offers no water protection; installation in wet or washdown areas is not permitted.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must remain between 5°C and 40°C for proper operation.

Operating outside this range may affect performance, reliability, and warranty.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

Space requirements around the unit must be respected per the operating conditions.

Air-cooled units need adequate clearance for heat rejection; insufficient space reduces cooling capacity.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Performance

The circulation pump delivers up to 29 l/min at 1 bar, with 21 l/min at 0.5 bar.

Determines whether the pump can overcome the pressure drop of the external circuit at the required flow.

Source
“max. delivery 29 l/min max. delivery pressure 1 bar Delivery at 0,5 bar 21 l/min”
Performance

Cooling capacity is 2.2 kW at 15°C, 1.6 kW at 0°C, and 1 kW at -10°C (at 20°C ambient).

Allows matching the chiller to the actual heat load at the intended operating temperature.

Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Compliance

The equipment is EMC Class A Group 1, which may require additional filtering in residential environments.

Class A equipment is intended for industrial use and may cause interference in residential areas.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigerant charge is 0.58 kg with a GWP of 1397, giving 0.8 t CO2 equivalent.

Needed for leak-check intervals, refrigerant accounting, and environmental compliance.

Source
“Refrigerant quantity 0,58 kg CO2 equivalent 0,8 t”