Unichiller 075w — Technical Data Sheet

Contents

  1. Overview p. 1
    Introduces the Unichiller 075w model and its main features
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data according to DIN 12876
  3. p. 2 table, 12 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight300 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)7,5 A
Fuse (3 phase)3x10A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S430098

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller 075w (Order-No.: 3040.0009.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C7,5 kW
at 0°C6,1 kW
at -10°C4 kW
at -20°C2,4 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,95 kg
CO2 equivalent1,3 t
Circulation pumpE1
max. delivery48 l/min
max. delivery pressure3,4 bar
Delivery at 0,3 bar46 l/min
Delivery at 0,5 bar44 l/min
Delivery at 1,0 bar40 l/min
Delivery at 1,5 bar33 l/min
Delivery at 2,0 bar27 l/min
Delivery at 2,5 bar20 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C348 l/h
Consumption at water 15°C, flow 0°C330 l/h
Consumption at water 15°C, flow -10°C270 l/h
Consumption at water 15°C, flow -20°C240 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 0.5 bar and a maximum pressure of 6 bar.

Ensures the cooling water circuit can provide adequate flow and pressure without overpressurizing the unit.

Source
“min. cooling water differential pressure 0,5 bar”
Installation

The unit must be operated with cooling water intake temperature between 5°C and 20°C, and if below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and ensures proper cooling performance and reliability.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.

Helps in sizing the cooling water supply and predicting performance under varying conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so only suitable cooling water should be used.

Guides water quality requirements to avoid corrosion or damage to the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage tolerance is ±5% with simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.

Ensures the power supply meets the unit's requirements and avoids invalid tolerance combinations.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug.

Requires the installer to provide a suitable plug and connection, affecting installation planning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant R-449A has a Global Warming Potential of 1397 and a CO2 equivalent of 1.3 t, which must be considered for environmental compliance.

Relevant for refrigerant handling, leak reporting, and environmental regulations.

Source
“Global Warming Potential (GWP) 1397”
Operation

The unit has a minimum filling capacity of 18 l and an expansion volume of 48 l, which must be respected for proper operation.

Ensures the system is filled correctly to avoid pump cavitation or overflow.

Source
“min. filling capacity 18 l”
Installation

The unit requires a 400V 3-phase 50Hz power supply with a maximum current of 7.5A and fuses of 3x10A.

Essential for electrical installation and ensuring the supply matches the unit's requirements.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Operation

The maximum ambient temperature for operation is 40°C and the minimum is 5°C.

Ensures the installation location maintains suitable ambient conditions for reliable operation.

Source
“max. ambient temperature 40 °C”
Maintenance

The unit uses a water-cooled refrigerating machine that is CFC- and HCFC-free, with refrigerant R-449A.

Indicates environmental compliance and maintenance requirements for the refrigerant circuit.

Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”
Performance

The circulation pump delivers a maximum flow of 48 l/min and a maximum pressure of 3.4 bar, with flow decreasing at higher pressures.

Helps in designing the external circuit to match pump performance and avoid excessive pressure drop.

Source
“max. delivery 48 l/min”