Peter Huber Kältemaschinenbau Unichiller 075w
Peter Huber Kältemaschinenbau Unichiller 075w — Recirculating chillers.
The Unichiller 075w is a water-cooled recirculating chiller with a 48 l/min pump and a -20…40 °C working range, distinguished within the series by its water-cooled refrigeration circuit and G1/2 cooling-water connection. It serves externally closed process loops requiring stable temperature control without air-side heat rejection.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 075w Current | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Expansion volume | 48 l |
| Maximum delivery flow rate | 48 l/min (also 3, 4 bar pump pressure) |
| Minimum fill volume | 18 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | E1 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 300 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 240 l/h (cooling water 15°C, flow -20°C) cooling water 15°C, flow -10°C: 270 l/h · cooling water 15°C, flow 0°C: 330 l/h |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
The Unichiller 075w is a water-cooled refrigerating machine (CFC- and HCFC-free, refrigerant R-449A) built for externally closed applications only. Unlike air-cooled siblings such as the 030T/030Tw variants, this model rejects heat to a cooling-water circuit via G1/2 male connections, so it needs no large air-side clearance but does need a cooling-water supply. The pump is a fixed-speed E1 type (not speed-adjustable) with G1 1/4 AG male connections, delivering up to 48 l/min at a maximum pressure of 3.4 bar. The 48 l expansion volume and 18 l minimum fill define the hydraulic buffer available to the loop.
Behavior under load and media
Cooling-water consumption depends on the required flow temperature at a 15 °C cooling-water intake: 270 l/h at -10 °C, 330 l/h at 0 °C, and 240 l/h at -20 °C. Cooling power falls and cooling-water flow demand may rise as the cooling-water temperature increases. The cooling-water circuit materials (copper, stainless steel 1.4401, MS, PA, PPE, PTFE, EPDM) constrain acceptable water quality. Pump flow decreases as external pressure drop rises, so the external circuit must be sized against the 3.4 bar maximum.
Installation, control and operation
The Pilot ONE controller provides Ethernet, USB (host and device) and RS232 interfaces, plus a Pt100 external sensor connection. Power-failure response is configurable as OFF/Standby, ON, or Power Failure Auto. Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached rather than switching off all poles. The unit is IP20, safety class I/NFL, and must not be installed in an ATEX zone.
Integration & Operating Notes
The chiller is designed for cooling-water intake between 5 °C and 20 °C with a minimum differential pressure of 0.5 bar and a maximum of 6 bar. If intake is below the condensation point, internal cooling-water pipes must be insulated to prevent condensate.
Free-standing models require a wall clearance of 50 to 200 cm on either side to permit maintenance access. The unit must be used indoors at an illuminance of at least 300 lx and at altitudes up to 2000 m.
The 400 V 3~ 50 Hz supply must respect ±5% voltage with simultaneous ±2% frequency tolerance, but the combination of -5% voltage and +2% frequency is not permitted. Three-phase devices drawing less than 63 A are supplied with a cable but no plug, so the installer must provide a suitable plug and connection.
The unit is designed exclusively for externally closed applications; open baths or vented circuits are outside its intended use. Models with 'P' in the model name are the ones suited to high-pressure-drop applications, so verify loop pressure drop against the 3.4 bar pump maximum.
The cooling-water supply must be closed even for short downtimes such as overnight to prevent water damage and corrosion. Before startup, activate parking brakes on casters and/or unscrew leveling feet so the unit cannot move during operation.
In Process Safety mode the controller releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles. Choose this mode where cutting cooling would cause runaway temperatures; otherwise the power-failure response (OFF/Standby, ON, or Power Failure Auto) governs restart behavior.
Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a GWP of 1397 and a CO2 equivalent of 1.3 t, which must be considered for leak reporting and environmental compliance.
Selection & Integration Q&A
Pick the 075w when a cooling-water circuit is available and air-side heat rejection or fan noise is undesirable; it is water-cooled with G1/2 connections. Choose an air-cooled 030T/030Tw when no cooling water is available, accepting their 69 dB(A) or 61 dB(A) noise levels.
Provide cooling water at 5–20 °C intake, 0.5 bar minimum differential and 6 bar maximum, through G1/2 male connections. Size the external loop against the pump's 48 l/min and 3.4 bar maximum, and use water compatible with copper, 1.4401 stainless, PA, PPE, PTFE and EPDM.
Install indoors only, at ≥300 lx and up to 2000 m altitude, with 50–200 cm wall clearance per side for maintenance. Ambient must stay within 5–40 °C. Shut off cooling water even overnight, and never install in an ATEX zone.
Three-phase units under 63 A ship with a cable but no plug, so the installer supplies the plug. Refrigerant-circuit work is restricted to approved refrigeration contractors. The 18 l minimum fill and 48 l expansion volume must be respected; no accessories or consumables are listed in the verified facts.