Product introduction and main features of the chiller
Specpp. 1-2
Technical data, performance and operating parameters
Specp. 2
Detailed values according to DIN 12876 standard
p. 2table, 10 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Overall dimensions WxDxH **
600x790x1614 mm
Net weight
284 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
16 A
Fuse (3 phase)
3x25A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S479767
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,2 K
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
10 kW
Cooling power at 0°C
9 kW
Cooling power at -10°C
6,5 kW
Cooling power at -20°C
3 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
4 kg
CO2 equivalent
5,6 t
Circulation pump
C3
max. delivery
96 l/min
max. delivery pressure
5,6 bar
Delivery at 0,5 bar
92 l/min
Delivery at 1,0 bar
86 l/min
Delivery at 1,5 bar
81 l/min
Delivery at 2,0 bar
75 l/min
Delivery at 2,5 bar
68 l/min
Delivery at 3,0 bar
62 l/min
Delivery at 3,5 bar
54 l/min
Delivery at 4,0 bar
46 l/min
Delivery at 4,5 bar
37 l/min
Delivery at 5,0 bar
24 l/min
Pump connection
G1 1/4 male
min. filling capacity
8,36 l
Volume of expansion
7 l
Order-No.: 3017.0030.01; Unichiller 100T-H
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a 400 V three-phase 50 Hz supply with a 16 A maximum current and 3x25 A fusing, so the electrical installation must be sized accordingly.
Determines the site power supply, cable cross-section and breaker rating needed before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug on site.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances are only valid in combination, and a -5% voltage with +2% frequency is explicitly not allowed.
Prevents connecting the unit to a marginal grid where combined tolerances could damage the compressor or electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is rated IP20, so it must be installed in a dry, protected location and not exposed to water or dust ingress.
Restricts the installation environment and rules out washdown or outdoor siting.
Source
“Degree of Protection IP20”
Installation
Ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop if the ambient temperature rises above the rated 20 °C.
Needs to be checked against the room ventilation and heat load at the installation site to avoid derating.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Selection
Cooling capacity falls from 10 kW at 15 °C to 3 kW at -20 °C, so the required duty point must be checked against the capacity curve, not just the headline 10 kW.
Prevents oversizing or undersizing when the process runs at low temperatures.
Source
“at 15°C 10 kW”
Selection
The circulation pump delivers 96 l/min at zero head but only 24 l/min at 5.0 bar, so the system pressure drop must be matched to the pump curve.
Ensures the external circuit resistance does not starve flow at the required duty point.
Source
“max. delivery pressure 5,6 bar”
Installation
The pump connection is G1 1/4 male, so hoses and adapters must match this thread size.
Determines the fittings and hoses to procure for the external circuit.
Source
“Pump connection G1 1/4 male”
Operation
The minimum filling capacity is 8.36 l and the expansion volume is 7 l, so the external loop plus unit must be charged with sufficient heat transfer fluid.
Avoids cavitation and low-level alarms by ensuring adequate fluid volume at commissioning.
Source
“min. filling capacity 8,36 l”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, giving a CO2 equivalent of 5.6 t that must be considered for leak detection and F-gas reporting.
Triggers F-gas obligations, room ventilation requirements and leak-check intervals.
Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL, which must be set for the application.
The protection limit must be configured to protect the process and the fluid from overheating.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance
The equipment is EMC class A group 1, so it is intended for industrial environments and may cause interference in residential areas.
Affects siting and shielding decisions when sensitive electronics are nearby.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance
The refrigeration machine is air-cooled, so the condenser needs unobstructed airflow and periodic cleaning.
Poor airflow causes capacity loss and high-pressure trips; maintenance access must be planned.
Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation
The overall dimensions are 600x790x1614 mm and the net weight is 284 kg, so floor loading and access routes must be verified.
Determines rigging, doorway clearance and floor point loading at the installation site.