Induztr
Engineering Digest Document Viewer Related Models

Unichiller 100T Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unichiller 100T
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)13 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S516014

Technical data according to DIN 12876 — Unichiller 100T (Order-No.: 3017.0029.01)

ParameterValue
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C10 kW
Cooling power at 0°C9 kW
Cooling power at -10°C6,5 kW
Cooling power at -20°C3 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity4 kg
CO2 equivalent5,6 t
Circulation pump:C3
max. delivery96 l/min
max. delivery pressure5,6 bar
Delivery at 0,5 bar92 l/min
Delivery at 1,0 bar86 l/min
Delivery at 1,5 bar81 l/min
Delivery at 2,0 bar75 l/min
Delivery at 2,5 bar68 l/min
Delivery at 3,0 bar62 l/min
Delivery at 3,5 bar54 l/min
Delivery at 4,0 bar46 l/min
Delivery at 4,5 bar37 l/min
Delivery at 5,0 bar24 l/min
Pump connectionG1 1/4 male
min. filling capacity8,36 l
Volume of expansion7 l
Overall dimensions WxDxH **600x790x1614 mm
Net weight271 kg

** см. сноску в оригинале (в тексте страницы расшифровка не приведена).

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation scope and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit must be installed where ambient temperature stays between 5 °C and 40 °C.

Room conditions must be checked during site selection to avoid operating outside the allowed range.

Source
“min. ambient temperature 5 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so capacity margins must account for the actual room conditions.

Prevents undersizing the chiller when the installation room is warmer than 20 °C.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is air-cooled, so adequate air flow and space around the refrigerating unit must be provided.

Air-cooled chillers need sufficient ventilation and clearance at the installation site.

Source
“air-cooled, CFC- and HCFC-free”
Installation

The pump connection is G1 1/4 male, so mating hoses or adapters must match this thread.

Determines the connection fittings needed for the external circuit.

Source
“Pump connection G1 1/4 male”
Operation

The minimum filling capacity is 8.36 l, so the external loop and expansion volume must be planned to meet this minimum.

Ensures the system has enough heat transfer fluid for safe pump operation.

Source
“min. filling capacity 8,36 l”
Operation

The expansion volume is 7 l, which limits how much thermal expansion the system can accommodate.

Needed to size the external circuit and avoid overfilling the expansion tank.

Source
“Volume of expansion 7 l”
Safety

The refrigerant is R-449A with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

Informs safety measures and labeling requirements for the refrigerant circuit.

Source
“R-449A (A1, H280)”
Compliance

The CO2 equivalent of the refrigerant charge is 5.6 t, which may trigger refrigerant reporting or leak-check obligations.

Relevant for environmental compliance and maintenance leak-check intervals.

Source
“CO2 equivalent 5,6 t”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts installation environment and prevents placing the unit where moisture or dust is present.

Source
“Degree of Protection IP20”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but not simultaneously in the unfavorable combination, so supply quality must be verified.

Ensures the site power quality is within the allowed envelope before commissioning.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is classified EN55011 Class A Group 1, so it is intended for industrial electromagnetic environments and may need separation from sensitive equipment.

Affects EMC planning and placement relative to sensitive electronics.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The unit is intended for externally closed applications, so the external circuit must be designed as a closed loop.

Prevents selecting the chiller for open-loop or once-through cooling arrangements.

Source
“For externally closed applications.”