Peter Huber Kältemaschinenbau Unichiller 100T
Peter Huber Kältemaschinenbau Unichiller 100T — Recirculating chillers.
The Unichiller 100T is an air-cooled recirculating chiller for externally closed loops, operating from -20 to 40 °C with a C3 pump and G1 1/4 male connections. It is the 100-size model in the Unichiller series, distinguished from smaller 030/040 siblings by its larger capacity and 400 V three-phase supply. It serves as the primary cooling source for closed process circuits requiring stable temperature control.
Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection
| Model | Noise level, dB | Pump connection | Pump type | Safety class | Temperature range, °C | Power supply |
|---|---|---|---|---|---|---|
| Unichiller 100T Current | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 030T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 460V 3~ 60Hz |
| Unichiller 030Tw | 61 | G3/4 AG male | B | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 040T | 69 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040T-H | 69 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 040Tw | 70 | G3/4 AG male | B | I / NFL | -10–40 | 400V 3~ 50Hz |
| Unichiller 040Tw-H | 70 | G3/4 AG male | B | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 050-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 050w-H | 62 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 055T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 055T-H | 74 | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 055Tw | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T | — | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 060T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 060Tw-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 070T-H | 66 | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 070Tw-H | — | G1 1/4 AG male | C3 | III / FL | -10–100 | 400V 3~ 50Hz |
| Unichiller 075w | — | G1 1/4 AG male | E1 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller 075w-H | — | G1 1/4 AG male | E1 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100-H | 58 | G1 1/4 AG male | E1 | III / FL | -20–100 | 460V 3~ 60Hz |
| Unichiller 100T-H | — | G1 1/4 AG male | C3 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller 100Tw | 65 | G1 1/4 AG male | C3 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller GL 035w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 050w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075 | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 075w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller GL 100w | — | G1 1/4 AG male | Pressure pump, speed not | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050 | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w | 62 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P050w-H | 62 | G1 1/4 AG male | E2 | III / FL | -20–100 | — |
| Unichiller P075 | 61 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P075w-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100 | — | G1 1/4 AG male | E2 | I / NFL | -20–40 | 460V 3~ 60Hz |
| Unichiller P100-H | — | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
| Unichiller P100w | 58 | G1 1/4 AG male | E2 | I / NFL | -20–40 | 400V 3~ 50Hz |
| Unichiller P100w-H | 58 | G1 1/4 AG male | E2 | III / FL | -20–100 | 400V 3~ 50Hz |
Full specifications
| Expansion volume | 7 l |
| Pump connection | G1 1/4 AG male |
| Pump speed adjustable | no |
| Pump type | C3 |
| External sensor connection | Pt100 |
| Temperature range | -20...40 °C |
| Weight | 271 kg |
| Current consumption | 13 A (max) |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Ambient temperature limit | 5...40 °C |
| Protection rating | IP20 |
| Safety class | I / NFL |
Description
Construction and series positioning
- Air-cooled refrigeration circuit using R-449A (A1, H280), CFC- and HCFC-free; no cooling water connection required.
- C3 pump with fixed speed (not adjustable) and G1 1/4 AG male pump connections.
- 7 l expansion volume and 8.36 l minimum filling capacity define the hydraulic envelope.
- Within the Unichiller series, the 100T is the larger-capacity variant; siblings 030T, 030Tw, 030Tw-H and 040T are smaller models. The 100T requires a 400 V 3~ 50 Hz supply with 13 A max current, whereas smaller siblings may differ in electrical configuration.
- Controller: Pilot ONE with Ethernet, USB (Host and Device) and RS232 interfaces; external Pt100 sensor input available.
Behavior under load and media
- Temperature range -20...40 °C; cooling capacity is rated at 20 °C room temperature and may drop as ambient temperature rises toward the 40 °C limit.
- Intended only for externally closed applications; not for open-loop or once-through cooling.
- Safety class I / NFL per DIN 12876; classification depends on whether the thermal fluid is combustible.
- Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached, rather than switching off all poles — relevant for exothermic processes.
Installation, connection and control
- Requires three-phase 400 V 3~ 50 Hz supply with 3x16 A fuse; voltage tolerance ±5% and frequency ±2%, not simultaneously in the unfavorable combination.
- Three-phase devices below 63 A are supplied with cable but without plug; a suitable plug must be provided on site.
- IP20 protection: indoor dry location only, protected from water and dust; illuminance at least 300 lx; altitude up to 2000 m.
- Wall clearance 50–200 cm per side for maintenance on free-standing models; leveling feet or parking brakes must be activated before operation.
- Power-failure response configurable via Pilot ONE: OFF/Standby, ON, or Power Failure Auto.
- EN55011 Class A Group 1 — industrial EMC environment; may need separation from sensitive equipment.
Integration & Operating Notes
Free-standing installation requires a wall clearance of 50 to 200 cm on either side to allow maintenance access to the refrigeration circuit and pump. Insufficient clearance blocks service work and may impede air flow on this air-cooled unit.
The unit draws up to 13 A from a 400 V 3~ 50 Hz supply and requires a 3x16 A fuse. Because three-phase devices below 63 A ship with cable but without plug, the installer must procure and fit a matching plug; verify voltage ±5% and frequency ±2% before commissioning.
The chiller is designed exclusively for externally closed applications. The external loop plus expansion volume must accommodate at least 8.36 l filling capacity, while the 7 l expansion volume limits how much thermal expansion the system can absorb without overfilling.
Published cooling capacity is valid at 20 °C room temperature. If the installation room runs warmer — up to the 40 °C ambient limit — available cooling capacity drops, so sizing margins must reflect actual site conditions rather than the datasheet rating.
The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON, or Power Failure Auto. For processes where loss of cooling is hazardous, configure Process Safety mode so maximum cooling is released at the overtemperature trigger instead of all poles switching off.
Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The R-449A charge has a CO2 equivalent of 5.6 t, which may trigger leak-check and reporting obligations; plan service access accordingly.
The unit is not explosion-protected and must not be installed or started up inside an ATEX zone. It is rated IP20 and for indoor use only, at illuminance of at least 300 lx and altitude up to 2000 m; check refrigerant charge against room volume to stay within the permissible limit per m³.
Selection & Integration Q&A
Choose the 100T when the closed-loop heat load and hydraulic volume exceed the capacity of the 030/040 models. It provides a larger expansion volume (7 l) and minimum fill of 8.36 l, and requires a 400 V three-phase 3x16 A supply — confirm your site has this before selecting it over single-phase siblings.
The external circuit must mate with G1 1/4 AG male pump connections and be a closed loop holding at least 8.36 l. Electrically, provide a 400 V 3~ 50 Hz supply with 3x16 A fuse and a site-fitted plug (cable supplied, plug not). The Pilot ONE offers Ethernet, USB and RS232 for controller integration.
Refrigerant-circuit work must be done by approved refrigeration contractors; the R-449A charge has a 5.6 t CO2 equivalent, which may impose leak-check intervals. The unit ships with a power cable but no plug, and no hoses or adapters for the G1 1/4 connections — these must be sourced separately. Shut off cooling water even for overnight downtime if water-cooled auxiliaries are present.