The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.
Defines the usable process temperature envelope during selection.
Source
“Operating temperature range -10...40 °C”
| Characteristic | Value |
|---|---|
| Operating temperature range | -10...40 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Temperature stability at -10°C | 0,5 K |
| Safety classification | I / NFL |
| Cooling power at 15°C | 16 kW |
| Cooling power at 0°C | 8,8 kW |
| Cooling power at -10°C | 4 kW |
| Refrigeration machine | air-cooled, CFC- and HCFC-free |
| Refrigerant (ASHRAE, GHS) | R-449A (A1, H280) |
| Global Warming Potential (GWP) | 1397 |
| Circulation pump: | C3 |
| max. delivery | 99 l/min |
| max. delivery pressure | 5,9 bar |
| Delivery at 1,0 bar | 91 l/min |
| Delivery at 2,0 bar | 81 l/min |
| Delivery at 3,0 bar | 70 l/min |
| Delivery at 4,0 bar | 57 l/min |
| Delivery at 5,0 bar | 39 l/min |
| Pump connection | G1 1/4 male |
| min. filling capacity | 14 l |
| Volume of expansion | 39 l |
| Overall dimensions WxDxH ** | 904x1582x1902 mm |
| Net weight | 433 kg |
| Power supply (3 Phase) | 400V 3~ 50Hz |
| Pressure equipment category | I |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | 1.0/24 |
** Overall dimensions WxDxH
The chiller is rated for an operating temperature range of -10 to 40 °C, so process requirements outside this window cannot be met.
Defines the usable process temperature envelope during selection.
“Operating temperature range -10...40 °C”
Cooling capacity drops from 16 kW at 15 °C to 4 kW at -10 °C, so sizing must be based on the actual operating temperature.
Prevents undersizing when the process runs at low temperatures.
“Cooling power at 15°C 16 kW at 0°C 8,8 kW at -10°C 4 kW”
The unit requires a 400 V three-phase 50 Hz supply, so site power must be verified before installation.
Determines electrical infrastructure and cable/plug requirements.
“Power supply (3 Phase) 400V 3~ 50Hz”
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug.
Affects electrical installation planning and procurement.
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions, so supply quality must be checked.
Avoids invalid supply conditions that could damage the compressor or void compliance.
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts installation environment and enclosure requirements.
“Degree of Protection IP20”
Ambient temperature must stay between 5 and 40 °C, and cooling capacity may drop above 20 °C room temperature.
Ensures the room ventilation and heat rejection are adequate for rated performance.
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
The refrigerant is R-449A with GWP 1397 and safety group A1/H280, so leak detection and refrigerant handling procedures apply.
Affects safety classification, service training, and environmental compliance.
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.
Prevents mismatch of process piping and accessories.
“Pump connection G1 1/4 male”
The pump delivers 99 l/min at zero head but only 39 l/min at 5.0 bar, so system pressure drop must be checked against the pump curve.
Ensures adequate flow through the external circuit at actual pressure loss.
“max. delivery 99 l/min max. delivery pressure 5,9 bar”
Minimum filling capacity is 14 l and expansion volume is 39 l, so the external loop volume must be considered for proper expansion tank operation.
Avoids overflow or insufficient expansion volume in closed external circuits.
“min. filling capacity 14 l Volume of expansion 39 l”
The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial environments.
Restricts use to industrial locations and informs EMC planning.
“Classification (disturbance) to EN55011: Class A, Group 1”
The unit is compliant with US-SNAP for industrial process refrigeration, so it can be applied in that end-use under US regulations.
Confirms refrigerant acceptability for US industrial process refrigeration applications.
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”
The 4-year warranty requires registration, so the commissioning paperwork must be completed to preserve coverage.
Protects the investment by ensuring warranty validity.
“4-year warranty - registration required.”