Peter Huber Kältemaschinenbau Unichiller 160T
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller 160T

Peter Huber Kältemaschinenbau Unichiller 160T — Recirculating chillers.

The Unichiller 160T is a free-standing, air-cooled recirculating chiller with a non-variable C3 pressure pump and Pilot ONE controller, rated -10 to 40 °C. Within the Unichiller series it is the 160-size, 400 V three-phase modification, distinguished from the 110T siblings by capacity rather than by a 'P' high-pressure-drop or water-cooled variant. It serves externally closed industrial process loops requiring up to 16 kW of cooling at 15 °C.

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Pump typeSafety classTemperature range, °CPower supplyExpansion volume, m³Minimum fill volume, m³
Unichiller 160T Current C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 110T C3I / NFL-10–40400V 3~ 50Hz0.007—
Unichiller 110T-H C3III / FL-10–100400V 3~ 50Hz0.007—
Unichiller 110Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 130T C3I / NFL-10–40400V 3~ 50Hz0.0390.014
Unichiller 130T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 130Tw C3I / NFL-20–40400V 3~ 50Hz——
Unichiller 160T-H C3III / FL-10–100400V 3~ 50Hz0.0390.014
Unichiller 160Tw C3I / NFL-20–40400V 3~ 50Hz0.01—
Unichiller 180-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 180T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 180T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 180Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 180Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 180w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 180w-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 200T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 200Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 210T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 210T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 210Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 230-H E1III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller 230w E1I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller 230w-H E1III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller 250T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 250T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 250Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 260T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 260T-H D3III / FL-20–100400V 3~ 50Hz0.0390.014
Unichiller 260Tw D3I / NFL-20–40400V 3~ 50Hz0.0220.015
Unichiller 260Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 300T D3I / NFL-20–40400V 3~ 50Hz0.0390.014
Unichiller 300T-H D3III / FL-20–100460V 3~ 60Hz0.0390.014
Unichiller 300Tw D3I / NFL-20–40460V 3~ 60Hz0.0220.015
Unichiller 300Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 350T D3I / NFL-20–40460V 3~ 60Hz0.0390.014
Unichiller 350Tw-H D3III / FL-20–100400V 3~ 50Hz0.0220.015
Unichiller 500Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 600Tw-H D3III / FL-20–100400V 3~ 50Hz0.007—
Unichiller 700Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller 800Tw D3I / NFL-20–40400V 3~ 50Hz0.007—
Unichiller GL 1000w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.05—
Unichiller GL 200w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 250w Pressure pump, speed notI / NFL-20–40460V 3~ 60Hz0.03—
Unichiller GL 300w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 400w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.03—
Unichiller GL 750w Pressure pump, speed notI / NFL-20–40400V 3~ 50Hz0.05—
Unichiller P180 E2I / NFL-20–40—0.0480.018
Unichiller P180-H E2III / FL-20–100400V 3~ 50Hz0.0480.018
Unichiller P180w E2I / NFL-20–40400V 3~ 50Hz0.0480.018
Unichiller P180w-H E2III / FL-20–100460V 3~ 60Hz0.0480.018
Unichiller P230 E2I / NFL-20–40460V 3~ 60Hz0.0480.018
Unichiller P230-H E2III / FL-20–100460V 3~ 60Hz0.0480.018

Full specifications

Core Performance & Functional Specifications
Expansion volume 39 l
Minimum fill volume 14 l
Pump connection G1 1/4 AG male
Pump speed adjustable no
Pump type C3
External sensor connection Pt100
Temperature range -10...40 °C
Dimensions, Weight & Mechanical Design
Weight 433 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Externally closed circuits only; no open bath

The manual restricts the unit to externally closed applications, so the hydraulic circuit must be designed as a sealed loop with an expansion volume of 39 l and a minimum fill of 14 l. Open-tank or atmospheric processes are outside the intended use and invalidate the thermal design assumptions.

Unichiller Operation Manual, Function Description / Selection, p. 34
Mounting
Reserve 50–200 cm clearance per side for service access

Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance. The unit must also stand indoors at illuminance of at least 300 lx and at altitudes up to 2000 m; installation inside an ATEX zone is prohibited because the unit is not explosion-protected.

Unichiller Operation Manual, Installation, p. 25
Process Limits
Size cooling capacity at the actual process temperature

Cooling capacity falls from 16 kW at 15 °C to 8.8 kW at 0 °C and 4 kW at -10 °C, and rated data assume 20 °C room temperature. Above 20 °C ambient the capacity may drop further. Selection must therefore use the lowest expected process temperature and the real ambient, not the headline 15 °C figure.

Unichiller 160T Technical Data, p. 1–2
Cooling & Media
Match external loop pressure drop to the fixed-speed pump curve

The C3 pump is not speed-adjustable and delivers 99 l/min at zero head but only 39 l/min at 5.0 bar. The external circuit pressure drop must be calculated against this curve; a high-drop loop will reduce flow below the process requirement, and the 'P' high-pressure-drop variant is the appropriate series alternative in that case.

Unichiller 160T Technical Data, p. 1
Maintenance
Refrigerant circuit work is restricted to approved contractors

The unit uses R-449A (A1, H280; GWP 1397). Any work on the refrigerant circuit must be carried out by approved refrigeration/air-conditioning system contractors. Leak detection and refrigerant handling procedures apply, and the refrigerant charge versus room size must be checked at installation.

Unichiller Operation Manual, Maintenance, p. 17; Technical Data, p. 1
Control & I/O
Configure power-failure response to match process restart policy

The Pilot ONE allows the power-failure response to be set to OFF/Standby, ON or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger is reached instead of switching off all poles. The chosen mode must be agreed with the process safety concept before commissioning.

Unichiller Operation Manual, Normal Operation, p. 18–19
Maintenance
Register the unit to activate the four-year warranty

The four-year warranty requires registration. Commissioning paperwork must be completed and submitted so that coverage is preserved; otherwise the warranty period may not apply to the delivered unit.

Unichiller 160T Technical Data, p. 1

Selection & Integration Q&A

When should I choose the Unichiller 160T rather than a 110T or a 'P' variant from the same series?

Pick the 160T when the process needs the larger 160-size capacity at 400 V three-phase and a standard-pressure-drop loop. Choose a 110T sibling for lower capacity; choose a 'P' model if the external circuit has high pressure drop, since the 160T's fixed-speed pump falls to 39 l/min at 5.0 bar.

What must I provide on the process side to connect the 160T hydraulically and electrically?

The pump connection is G1 1/4 male, so mating hoses and fittings must match that thread and be counterheld at the maximum tightening torque. Electrically, the site must supply 400 V 3~ 50 Hz within ±5% voltage and ±2% frequency and provide a plug, since three-phase units below 63 A ship with cable but no plug.

What installation environment and duty-cycle limits apply to the 160T?

Install indoors only, at 5–40 °C ambient, illuminance at least 300 lx, altitude up to 2000 m, with 50–200 cm clearance per side. It is IP20 and not ATEX-rated. Cooling capacity is rated at 20 °C room temperature and may drop above that; shut off cooling water even for overnight downtimes.

What maintenance and consumables are required, and what is not included in the base delivery?

Refrigerant-circuit work on the R-449A system must be done by approved refrigeration contractors. The unit ships with cable but without a plug for three-phase supply below 63 A, and the four-year warranty requires registration. External hoses, fittings and the mating plug are customer-supplied.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF
PDF
Unichiller 160T Technical Data
Datasheet ·185,9 КБ ·2 pp.
View PDF