Induztr
Engineering Digest Document Viewer Related Models

Unichiller 210T-H Datasheet

Contents

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 3 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/19

Unichiller 210T-H — Technical data according to DIN 12876

ParameterValue
Order-No.3020.0030.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,2 K
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C21 kW
Cooling power at 0°C21 kW
Cooling power at -10°C13,5 kW
Cooling power at -20°C7,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:D3
max. delivery210 l/min
max. delivery pressure4,7 bar
Delivery at 1,1 bar202 l/min
Delivery at 1,5 bar193 l/min
Delivery at 2,0 bar179 l/min
Delivery at 2,5 bar159 l/min
Delivery at 3,0 bar140 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar77 l/min
Delivery at 4,5 bar31 l/min
Pump connectionG1 1/4 male
min. filling capacity14 l
Volume of expansion39 l
Overall dimensions WxDxH **905x2172x1900 mm
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20

** см. сноску к обозначению габаритов (Overall dimensions WxDxH) в оригинале.

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 100 °C, which defines the process window it can serve.

Confirms whether the unit covers the required process temperatures before selection.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity drops from 21 kW at 0 °C to 7.5 kW at -20 °C, so low-temperature duty needs a larger unit or derating check.

Prevents undersizing when the process runs near the cold end of the range.

Source
“at -20°C 7,5 kW”
Installation

The unit requires a 400 V three-phase 50 Hz supply, so site power must be verified before installation.

Determines electrical infrastructure and whether a suitable three-phase feed is available.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the final connection.

Affects electrical installation planning and the need to procure a suitable plug or isolator.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in opposite directions, so supply quality must be checked.

Avoids invalid supply conditions that could damage the compressor or void compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Installation

Ambient temperature must stay between 5 and 40 °C, and cooling capacity may fall above 20 °C room temperature.

Guides room ventilation and placement to avoid capacity loss or overheating.

Source
“max. ambient temperature 40 °C”
Installation

The unit is air-cooled and CFC/HCFC-free with refrigerant R-449A (GHS H280), so ventilation and refrigerant handling rules apply.

Informs room ventilation, leak response and service safety planning.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280)”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Ensures correct hydraulic connection hardware is ordered and installed.

Source
“Pump connection G1 1/4 male”
Operation

The pump delivers 210 l/min at zero head but only 31 l/min at 4.5 bar, so system pressure drop must be matched to the pump curve.

Prevents selecting a unit whose flow collapses under the actual circuit resistance.

Source
“max. delivery 210 l/min”
Operation

Minimum filling capacity is 14 l and expansion volume is 39 l, so the external loop volume must be considered for proper operation.

Avoids underfilling or overflow issues in the closed external circuit.

Source
“min. filling capacity 14 l”
Safety

The unit has adjustable overtemperature protection per DIN 12876 and safety classification III/FL, so it can be set to protect the process.

Supports safe operation and compliance with laboratory chiller safety requirements.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Compliance

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and needs appropriate installation.

Affects placement and EMC planning, especially near sensitive electronics.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Overall dimensions are 905x2172x1900 mm and space requirements must be respected, so the installation footprint must be planned.

Ensures sufficient floor space and access for installation and service.

Source
“Overall dimensions WxDxH ** 905x2172x1900 mm”