** Please respect space requirements. See operating conditions at www.huber-online.com
Unichiller P015w — Technical data according to DIN 12876
Parameter
Value
Order-No.
3051.0025.01
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at ambient temperature +20°C
at 15°C
1,5 kW
at 0°C
1 kW
at -10°C
0,7 kW
at -20°C
0,3 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
B
max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 0,5 bar
21 l/min
Delivery at 1,0 bar
17 l/min
Delivery at 1,5 bar
11 l/min
Delivery at 2,0 bar
6 l/min
Pump connection
G3/4 male
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 15°C
68 l/h
Consumption at water 15°C, flow 0°C
60 l/h
Consumption at water 15°C, flow -10°C
48 l/h
Consumption at water 15°C, flow -20°C
36 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,8 l
Cooling power values at 15°C, 0°C, -10°C, -20°C are sub-rows of 'Cooling power at ambient temperature +20°C'.
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar, so the site water supply must be verified against these limits before installation.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the water circuit.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and if the cooling water temperature falls below the condensation point the cooling water pipes in the unit must be insulated.
Operating outside the intake temperature range or omitting insulation causes condensation and possible damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so the water supply must be sized for the actual site conditions.
Sizing the cooling water supply only for nominal conditions can lead to insufficient cooling capacity in practice.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
The plastic pump is not resistant to acetone or polyglycols, so these media must not be used; for freeze protection a water mixture with glysantine or ethylene glycol is recommended.
Using incompatible media will damage the pump and void reliable operation.
Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used to avoid corrosion.
Aggressive or untreated cooling water can corrode the mixed-material water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is explicitly not allowed.
The site power supply must be checked against this combined tolerance to avoid invalid operating conditions.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%
Example
-5% voltage and +2% frequency -> not allowed!”
Installation
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering, while three-phase devices below 63A come with cable but without plug.
The plug configuration must be specified at order time and the site must provide the matching receptacle or a plug for three-phase units.
Source
“1. Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)
2. Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit has IP20 protection and requires ambient temperature between 5°C and 40°C, so it must be installed in a dry indoor location within this temperature range.
IP20 offers no water protection and the ambient limits define the permissible installation environment.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Selection
The refrigeration machine is water-cooled and uses R-449A refrigerant with GWP 1397, which matters for refrigerant regulations and environmental compliance at the site.
The refrigerant type and GWP affect regulatory obligations and environmental reporting for the installation.
Source
“Refrigeration machine
water-cooled, CFC- and
HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397”
Selection
The pump delivers 25 l/min maximum and only 6 l/min at 2.0 bar, so applications with high pressure drop must be checked against the pump curve.
High-pressure-drop circuits may receive insufficient flow, so the pump curve must be matched to the application.
Source
“max. delivery
25 l/min
max. delivery pressure
2,5 bar
Delivery at 2,0 bar
6 l/min”
Installation
The minimum filling capacity is 3.8 l and the expansion vessel volume is 1.7 l, which determines the minimum heat transfer fluid charge and expansion allowance.
Underfilling below the minimum capacity prevents proper operation and expansion control.
Source
“min. filling capacity
3,8 l”
Operation
The unit is factory-set for 220-240V 1~/2~ 50Hz with maximum current 5.5 A and requires a fuse between 10A and 16A.
The electrical supply and protective device must be sized within these limits for safe connection.
Source
“Power supply requirement
220-240V 1~/2~ 50Hz
max. current
5,5 A
min. Fuse
10A
max. Fuse
16A”
Compliance
The equipment is classified for electromagnetic disturbance as EN55011 Class A Group 1, which restricts its use to industrial environments.
Class A equipment may cause interference in residential areas and is intended for industrial use only.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”