Unichiller P022 Recirculating Chiller Datasheet

Contents

  1. Overview p. 1
    Product introduction and main features of the Unichiller P022
  2. Spec pp. 1-2
    Technical data, performance and dimensions per DIN 12876
  3. p. 2 table, 8 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Power supply requirement220-240V 1~ 50Hz
max. current8,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S585478

Unichiller P022 — Technical data according to DIN 12876

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Resolution of display0,1 K
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 15°C2,2 kW
at 0°C1,6 kW
at -10°C1 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-290 (A3, H220)
Global Warming Potential (GWP)0,02
Refrigerant quantity0,25 kg
CO2 equivalent0,0 t
UN-numberUN 3358
Circulation pump:B
max. delivery25 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar21 l/min
Delivery at 1,0 bar17 l/min
Delivery at 1,5 bar11 l/min
Delivery at 2,0 bar6 l/min
Pump connectionG3/4 male
min. filling capacity3,8 l
Volume of expansion1,7 l
Overall dimensions WxDxH **460x590x743 mm
Net weight81 kg
sound pressure level +/- 4 dB(A)65 dB(A)

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

13 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -10 to 40 °C, so it must be selected only for processes within this window.

Defines the valid process temperature envelope for selection.

Source
“Operating temperature range -10...40 °C”
Selection

Cooling capacity drops with falling temperature (2.2 kW at 15 °C, 1.6 kW at 0 °C, 1 kW at -10 °C), so sizing must use the capacity at the actual working temperature.

Prevents undersizing when the process runs at low temperatures.

Source
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
Safety

The unit uses flammable refrigerant R-290 (A3, H220), which imposes specific handling and ventilation requirements during installation and service.

Flammable refrigerant classification affects site safety and service procedures.

Source
“R-290 (A3, H220)”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must be selected accordingly.

Determines the connection hardware needed for plumbing the loop.

Source
“Pump connection G3/4 male”
Installation

The minimum filling capacity is 3.8 l with 1.7 l expansion volume, so the loop must be filled and sized within these limits.

Sets the required fluid volume and expansion allowance for the circuit.

Source
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
Installation

The unit requires a 220-240 V single-phase 50 Hz supply with max current 8.5 A and a 16 A fuse, so the electrical supply must be sized accordingly.

Defines the electrical feed and protection rating for installation.

Source
“220-240V 1~ 50Hz max. current 8,5 A Fuse 16 A”
Installation

The degree of protection is IP20, so the unit must be installed in a dry, protected environment.

Restricts installation location regarding moisture and dust ingress.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop if ambient rises above 20 °C.

Ensures the installation room stays within the allowed ambient window.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant pump ordered.

Prevents chemical attack on the pump when choosing the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.

Guides correct fluid preparation to avoid freezing damage.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.

Constrains the acceptable power quality at the installation site.

Source
“-5% voltage and +2% frequency -> not allowed!”
Compliance

The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which affects EMC planning at the site.

Class A equipment may require EMC measures in residential-like environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.

Ensures the correct plug type is ordered for the installation country.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”