The chiller is rated for an operating temperature range of -10 to 40 °C, so it must be selected only for processes within this window.
Defines the valid process temperature envelope for selection.
Source
“Operating temperature range -10...40 °C”
| Parameter | Value |
|---|---|
| Power supply requirement | 220-240V 1~ 50Hz |
| max. current | 8,5 A |
| Fuse | 16 A |
| Pressure equipment category | Art. 4.3 PED |
| Degree of Protection | IP20 |
| min. ambient temperature | 5 °C |
| max. ambient temperature | 40 °C |
| from Serial-No.: | S585478 |
| Parameter | Value |
|---|---|
| Operating temperature range | -10...40 °C |
| temperature set point / display | 5,7" colour Touchscreen |
| Internal temperature sensor | Pt100 |
| Sensor external connection | Pt100 |
| Resolution of display | 0,1 K |
| Temperature stability at -10°C | 0,5 K |
| Interface digital | Ethernet, USB (Host u. Device), RS232 |
| Safety classification | I / NFL |
| Cooling power at ambient temperature +20°C | |
| at 15°C | 2,2 kW |
| at 0°C | 1,6 kW |
| at -10°C | 1 kW |
| Refrigeration machine | air-cooled, natural refrigerant |
| Refrigerant (ASHRAE, GHS) | R-290 (A3, H220) |
| Global Warming Potential (GWP) | 0,02 |
| Refrigerant quantity | 0,25 kg |
| CO2 equivalent | 0,0 t |
| UN-number | UN 3358 |
| Circulation pump: | B |
| max. delivery | 25 l/min |
| max. delivery pressure | 2,5 bar |
| Delivery at 0,5 bar | 21 l/min |
| Delivery at 1,0 bar | 17 l/min |
| Delivery at 1,5 bar | 11 l/min |
| Delivery at 2,0 bar | 6 l/min |
| Pump connection | G3/4 male |
| min. filling capacity | 3,8 l |
| Volume of expansion | 1,7 l |
| Overall dimensions WxDxH ** | 460x590x743 mm |
| Net weight | 81 kg |
| sound pressure level +/- 4 dB(A) | 65 dB(A) |
** см. сноску в оригинале (Overall dimensions WxDxH **)
The chiller is rated for an operating temperature range of -10 to 40 °C, so it must be selected only for processes within this window.
Defines the valid process temperature envelope for selection.
“Operating temperature range -10...40 °C”
Cooling capacity drops with falling temperature (2.2 kW at 15 °C, 1.6 kW at 0 °C, 1 kW at -10 °C), so sizing must use the capacity at the actual working temperature.
Prevents undersizing when the process runs at low temperatures.
“at 15°C 2,2 kW at 0°C 1,6 kW at -10°C 1 kW”
The unit uses flammable refrigerant R-290 (A3, H220), which imposes specific handling and ventilation requirements during installation and service.
Flammable refrigerant classification affects site safety and service procedures.
“R-290 (A3, H220)”
The pump connection is G3/4 male, so mating hoses and fittings must be selected accordingly.
Determines the connection hardware needed for plumbing the loop.
“Pump connection G3/4 male”
The minimum filling capacity is 3.8 l with 1.7 l expansion volume, so the loop must be filled and sized within these limits.
Sets the required fluid volume and expansion allowance for the circuit.
“min. filling capacity 3,8 l Volume of expansion 1,7 l”
The unit requires a 220-240 V single-phase 50 Hz supply with max current 8.5 A and a 16 A fuse, so the electrical supply must be sized accordingly.
Defines the electrical feed and protection rating for installation.
“220-240V 1~ 50Hz max. current 8,5 A Fuse 16 A”
The degree of protection is IP20, so the unit must be installed in a dry, protected environment.
Restricts installation location regarding moisture and dust ingress.
“Degree of Protection IP20”
The ambient temperature must stay between 5 °C and 40 °C, and cooling capacity may drop if ambient rises above 20 °C.
Ensures the installation room stays within the allowed ambient window.
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant pump ordered.
Prevents chemical attack on the pump when choosing the heat transfer fluid.
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
For freeze protection, water should be mixed with glysantine or ethylene glycol rather than relying on pure water.
Guides correct fluid preparation to avoid freezing damage.
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
Constrains the acceptable power quality at the installation site.
“-5% voltage and +2% frequency -> not allowed!”
The unit is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which affects EMC planning at the site.
Class A equipment may require EMC measures in residential-like environments.
“Classification (disturbance) to EN55011: Class A, Group 1”
Single/two-phase devices are delivered with a power cable and country-specific plug that must be specified at ordering.
Ensures the correct plug type is ordered for the installation country.
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”