Peter Huber Kältemaschinenbau Unichiller P022
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unichiller P022

Peter Huber Kältemaschinenbau Unichiller P022 — Recirculating chillers with Noise level 65 dB.

The Unichiller P022 is a recirculating chiller with a plastic pump (type B) and G3/4 male connections, rated for -10 to 40 °C and intended for externally closed circuits with high pressure drop. Its cooling capacity falls from 2.2 kW at 15 °C to 1.0 kW at -10 °C, so sizing must use the capacity at the actual working temperature. It uses flammable R-290 refrigerant and requires a 220–240 V single-phase 50 Hz supply.

Noise level
65 dB

Peter Huber Kältemaschinenbau Unichiller Series — Model Comparison & Selection

Model Noise level, dBControllerSafety classTemperature range, °CInterfacePower supply
Unichiller P022 Current 65Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller 015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 015w-H OLÉ 54OLÉI / NFL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 017T 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017T-H 64Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 017Tw-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020T —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020T-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller 020Tw 60Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 020Tw-H 60Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 022-H 65Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 022w-H —Pilot ONEIII / FL-10–100——
Unichiller 022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller 025T 74Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025T-H 74Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025Tw-H —Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D230V 1~ 50Hz
Unichiller 025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller 025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller 025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P007 58Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208-240V 1~/2~ 50/60Hz
Unichiller P007-H —Pilot ONEIII / FL-20–100——
Unichiller P007-H OLÉ 58OLÉIII / FL-20–100USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P007 OLÉ 58OLÉI / NFL-20–40USB (Device), RS232 Inte208-240V 1~/2~ 50/60Hz
Unichiller P010 57Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P010-H 57Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010-H OLÉ 57OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P010 OLÉ 57OLÉI / NFL-20–40USB (Device), RS232 Inte110-120V 1~ 50/60Hz
Unichiller P010w —Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P010w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012 59Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D208V 2~ 60Hz
Unichiller P012-H 59Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012-H OLÉ 59OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012 OLÉ 59OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P012w-H —Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~/2~ 50/60Hz
Unichiller P012w-H OLÉ —OLÉIII / FL-20–100USB (Device), RS232 Inte208V 2~ 60Hz
Unichiller P012w OLÉ —OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~/2~ 50/60Hz
Unichiller P015 63Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015-H 63Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P015-H OLÉ 63OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015 OLÉ 63OLÉI / NFL-20–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P015w 54Pilot ONEI / NFL-20–40Ethernet, USB (Host u. D220-240V 1~/2~ 50Hz
Unichiller P015w-H 54Pilot ONEIII / FL-20–100Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P015w-H OLÉ 54OLÉIII / FL-20–100USB (Device), RS232 Inte220-240V 1~/2~ 50Hz
Unichiller P022-H 65Pilot ONEI / NFL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022-H OLÉ 65OLÉI / NFL-10–100USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P022 OLÉ 65OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P022w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P022w-H 60Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P022w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz
Unichiller P025 64Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H 64Pilot ONEIII / FL-10–100Ethernet, USB (Host u. D220-240V 1~ 50Hz
Unichiller P025-H OLÉ 64OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025 OLÉ 64OLÉI / NFL-10–40USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w 60Pilot ONEI / NFL-10–40Ethernet, USB (Host u. D208-230V 1~/2~ 60Hz
Unichiller P025w-H OLÉ 60OLÉIII / FL-10–100USB (Device), RS232 Inte220-240V 1~ 50Hz
Unichiller P025w OLÉ 60OLÉI / NFL-10–40USB (Device), RS232 Inte208-230V 1~/2~ 60Hz

Full specifications

Core Performance & Functional Specifications
Noise level 65 dB(A)
Pump connection G3/4 AG male
Pump speed adjustable no
Pump type B
External sensor connection Pt100
Temperature range -10...40 °C
Dimensions, Weight & Mechanical Design
Weight 81 kg
Electrical, Power & Drive Specifications
Fuse rating 16 A
Power supply 220-240V 1~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 1 K
Digital interface Ethernet, USB (Host u. Device), RS232
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class I / NFL

Description

Construction and series positioning

Thermal behavior and media

Installation, control and operation

Integration & Operating Notes

Process Limits
Cooling capacity must be sized at the actual working temperature

Capacity falls from 2.2 kW at 15 °C to 1.6 kW at 0 °C and 1.0 kW at -10 °C, and further drops if ambient exceeds 20 °C. Selecting the unit on its 15 °C rating for a -10 °C process will undersize the chiller and risk overtemperature trips.

Datasheet, Specifications
Cooling & Media
Plastic pump incompatible with acetone and polyglycols

The standard plastic pump is not resistant to acetone or polyglycols (depending on manufacturer). For freeze protection, use water mixed with glysantine or ethylene glycol; verify fluid compatibility before filling the 3.8 l minimum circuit with 1.7 l expansion volume.

Datasheet, Specifications
Electrical
Voltage and frequency tolerances are coupled and restricted

The supply must be 220–240 V 1~ 50 Hz with max current 8.5 A and a 16 A fuse. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed; verify site power quality before commissioning.

Datasheet, Specifications
Mounting
Refrigerant charge vs room volume must be checked on site

The unit uses flammable R-290 (A3, H220). The refrigerant charge (see data sheet/rating plate) and the room size must be checked at installation to avoid exceeding the refrigerant limit per m³ of ambient air; work on the refrigerant circuit must be performed by approved refrigeration contractors.

Operation Manual, Installation
Mounting
Maintain 50–200 cm wall clearance per side for service access

Free-standing models require a wall clearance of 50 to 200 cm on either side for maintenance. The unit must be installed indoors at illuminance of at least 300 lx and up to 2000 m altitude, and must not be placed in an ATEX zone.

Operation Manual, Installation
Process Limits
Only externally closed circuits are permitted

The P022 is designed for externally closed applications; open baths or circuits that expose the fluid to atmosphere are outside the intended use. The "P" designation additionally targets loops with high pressure drop, so verify the process pressure drop against the pump curve.

Operation Manual, Function Description
Control & I/O
Configure power-failure response to match process risk

The Pilot ONE allows power-failure response OFF/Standby, ON, or Power Failure Auto. For exothermic processes, Process Safety mode releases maximum cooling capacity when the overtemperature trigger value is reached instead of switching off all poles; select the mode that matches the process hazard analysis.

Operation Manual, Normal Operation
Maintenance
Refrigerant circuit work restricted to approved contractors

Any work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors. The unit is classified to DIN 12876 with identification depending on whether the thermal fluid is combustible; confirm the safety class (I/NFL) against the fluid used.

Operation Manual, Maintenance
Mounting
Secure casters and leveling feet before start-up

Parking brakes must be activated at the casters (if any) and/or leveling feet unscrewed/activated before the unit is put into operation. Pump connections must be counterheld and maximum tightening torques must not be exceeded to avoid leaks.

Operation Manual, Commissioning

Selection & Integration Q&A

When should I choose the P022 over the Unichiller 015 or 015-H variants?

Choose the P022 when the process loop has high pressure drop, since the "P" model is specifically suited to such applications. The 015/015-H siblings are rated 63 dB(A) versus 65 dB(A) for the P022; select a sibling if noise is the dominant constraint and pressure drop is low.

What mating hardware and fluid circuit preparation does the P022 require?

The pump connection is G3/4 male, so hoses and fittings must match; counterhold connections and do not exceed maximum tightening torques. Fill to at least 3.8 l with 1.7 l expansion allowance, using water with glysantine or ethylene glycol for freeze protection; avoid acetone and polyglycols with the plastic pump.

What installation environment and duty limits apply to the P022?

Install indoors only, at 5–40 °C ambient, illuminance at least 300 lx, altitude up to 2000 m, and never in an ATEX zone. Maintain 50–200 cm wall clearance per side, protect electrical lines below the unit from condensate, and shut off cooling water even for overnight downtimes.

What maintenance and accessories are not included in the base delivery?

The base unit ships without a country-specific plug unless specified at ordering; single-phase devices are delivered with a power cable but the plug must be requested. Refrigerant circuit service must be performed by approved contractors; no consumables or spare parts are listed in the provided documents.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unichiller P022 Recirculating Chiller Datasheet
Datasheet ·160,0 КБ ·2 pp.
View PDF
PDF
Unichiller Operation Manual
Manual ·2,8 МБ ·82 pp.
View PDF