Unichiller P075w-H Datasheet

Contents

  1. Overview p. 1
    Introduces the Unichiller P075w-H chiller unit
  2. Spec p. 2
    Performance and technical data per DIN 12876
  3. p. 2 table, 19 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C348 l/h
Consumption at water 15°C, flow 0°C330 l/h
Consumption at water 15°C, flow -10°C270 l/h
Consumption at water 15°C, flow -20°C240 l/h
min. cooling water differential pressure0,5 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight296 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S618391

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicUnichiller P075w-H
Order-No.3040.0012.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C at 100°C7,5 kW
Cooling power at ambient temperature +20°C at 20°C7,5 kW
Cooling power at ambient temperature +20°C at 0°C5,3 kW
Cooling power at ambient temperature +20°C at -10°C3,3 kW
Cooling power at ambient temperature +20°C at -20°C1,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,95 kg
CO2 equivalent1,3 t
Circulation pumpE2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires a cooling water connection with G1/2 male thread, so the site must provide a matching cooling water supply line.

Determines the piping and fittings needed at the installation site before commissioning.

Source
“Cooling water connection G1/2 male”
Installation

The unit requires a minimum cooling water differential pressure of 0.5 bar and tolerates a maximum cooling water pressure of 6 bar, so the supply must be sized within this window.

Prevents undersized or overpressurized cooling water supply that would impair cooling or damage the circuit.

Source
“min. cooling water differential pressure 0,5 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and if the cooling water falls below the condensation point the cooling water pipes inside the unit must be insulated.

Defines the allowable cooling water temperature range and the need for insulation to avoid condensation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power is rated at 7.5 kW at 20 °C and drops to 1.8 kW at -20 °C, so the cooling capacity must be checked against the actual process temperature.

Ensures the selected chiller can deliver enough cooling at the lowest required process temperature.

Source
“at 20°C 7,5 kW at 0°C 5,3 kW at -10°C 3,3 kW at -20°C 1,8 kW”
Installation

The unit is a three-phase 400 V 3~ 50 Hz device with a maximum current of 14 A and a 3x16 A fuse, and three-phase devices under 63 A are supplied with cable but without a plug.

The electrician must provide a suitable 3-phase supply and a plug or direct connection matching the 3x16 A protection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency is explicitly not allowed.

Warns that the site power quality must satisfy the combined voltage/frequency tolerance or the unit may not operate correctly.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, with 0.95 kg charge equal to 1.3 t CO2 equivalent.

Affects refrigerant handling, leak reporting obligations and environmental compliance at the installation site.

Source
“R-449A (A1, H280)”
Installation

The unit has protection degree IP20 and must be operated within an ambient temperature of 5 °C to 40 °C.

Restricts the installation location to a dry indoor environment within the specified ambient range.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit needs a minimum filling capacity of 18 l and has an expansion volume of 48 l, and it is intended for externally closed applications.

Confirms the external circuit must be closed and sized to accommodate the expansion volume.

Source
“For externally closed applications.”
Performance

The circulation pump delivers a maximum of 130 l/min at up to 5.7 bar, but flow falls to 20 l/min at 5.5 bar, so system pressure drop must be checked against the pump curve.

Ensures the pump can overcome the external circuit pressure drop at the required flow rate.

Source
“max. delivery 130 l/min max. delivery pressure 5,7 bar”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Guides water quality selection to avoid corrosion or fouling of the mixed-material cooling water circuit.

Source
“Please use suitable cooling water.”
Compliance

The equipment is classified for electromagnetic disturbance as EN55011 Class A, Group 1.

Class A equipment may require EMC precautions in residential or light-industrial environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”