** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Characteristic
Unichiller P075w-H
Order-No.
3040.0012.01
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at ambient temperature +20°C at 100°C
7,5 kW
Cooling power at ambient temperature +20°C at 20°C
7,5 kW
Cooling power at ambient temperature +20°C at 0°C
5,3 kW
Cooling power at ambient temperature +20°C at -10°C
3,3 kW
Cooling power at ambient temperature +20°C at -20°C
1,8 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
0,95 kg
CO2 equivalent
1,3 t
Circulation pump
E2
max. delivery
130 l/min
max. delivery pressure
5,7 bar
Delivery at 2,4 bar
116 l/min
Delivery at 2,5 bar
115 l/min
Delivery at 3,0 bar
108 l/min
Delivery at 3,5 bar
100 l/min
Delivery at 4,0 bar
92 l/min
Delivery at 4,5 bar
80 l/min
Delivery at 5,0 bar
59 l/min
Delivery at 5,5 bar
20 l/min
Pump connection
G1 1/4 male
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water connection with G1/2 male thread, so the site must provide a matching cooling water supply line.
Determines the piping and fittings needed at the installation site before commissioning.
Source
“Cooling water connection
G1/2 male”
Installation
The unit requires a minimum cooling water differential pressure of 0.5 bar and tolerates a maximum cooling water pressure of 6 bar, so the supply must be sized within this window.
Prevents undersized or overpressurized cooling water supply that would impair cooling or damage the circuit.
Source
“min. cooling water differential pressure
0,5 bar
max. cooling water pressure
6 bar”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, and if the cooling water falls below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the allowable cooling water temperature range and the need for insulation to avoid condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power is rated at 7.5 kW at 20 °C and drops to 1.8 kW at -20 °C, so the cooling capacity must be checked against the actual process temperature.
Ensures the selected chiller can deliver enough cooling at the lowest required process temperature.
Source
“at 20°C
7,5 kW
at 0°C
5,3 kW
at -10°C
3,3 kW
at -20°C
1,8 kW”
Installation
The unit is a three-phase 400 V 3~ 50 Hz device with a maximum current of 14 A and a 3x16 A fuse, and three-phase devices under 63 A are supplied with cable but without a plug.
The electrician must provide a suitable 3-phase supply and a plug or direct connection matching the 3x16 A protection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and a -5% voltage with +2% frequency is explicitly not allowed.
Warns that the site power quality must satisfy the combined voltage/frequency tolerance or the unit may not operate correctly.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant is R-449A with GHS classification H280 and a GWP of 1397, with 0.95 kg charge equal to 1.3 t CO2 equivalent.
Affects refrigerant handling, leak reporting obligations and environmental compliance at the installation site.
Source
“R-449A (A1, H280)”
Installation
The unit has protection degree IP20 and must be operated within an ambient temperature of 5 °C to 40 °C.
Restricts the installation location to a dry indoor environment within the specified ambient range.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit needs a minimum filling capacity of 18 l and has an expansion volume of 48 l, and it is intended for externally closed applications.
Confirms the external circuit must be closed and sized to accommodate the expansion volume.
Source
“For externally closed applications.”
Performance
The circulation pump delivers a maximum of 130 l/min at up to 5.7 bar, but flow falls to 20 l/min at 5.5 bar, so system pressure drop must be checked against the pump curve.
Ensures the pump can overcome the external circuit pressure drop at the required flow rate.