Induztr
Engineering Digest Document Viewer Related Models

Unistat 1005w Technical Data

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat 1005w unit
  2. Spec pp. 2-3
    Technical data according to DIN 12876 including performance and dimensions
  3. p. 2 table, 30 rows
  4. p. 1 table, 31 rows

Technical data according to DIN 12876

ParameterValue
Refrigerant 3rd stage (ASHRAE, GHS)R-14 (A1, H280)
Global Warming Potential (GWP)7380
Refrigerant quantity 3rd stage0,6 kg
CO2 equivalent 3rd stage4,4 t
UN-numberUN 3358
Circulation pump:
max. delivery30 l/min
max. delivery pressure0,9 bar
Delivery at 0,2 bar27 l/min
Delivery at 0,4 bar23 l/min
Delivery at 0,6 bar15 l/min
Delivery at 0,8 bar8 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C300 l/h
min. cooling water differential pressure3 bar
max. cooling water pressure6 bar
min. filling capacity3,6 l
Filling capacity expansion tank8 l
Overall dimensions WxDxH **700x804x1520 mm
Net weight340 kg
Power supply requirement400V 3~ 50Hz
max. current16,5 A
Fuse3x20A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S585552

** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Technical data according to DIN 12876 — Unistat 1005w (Order-No.: 1062.0002.01)

CharacteristicValue
Operating temperature range-120...100 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 100°C1,5 kW
Cooling power with Thermooil at 50°C1,5 kW
Cooling power with Thermooil at 20°C1,5 kW
Cooling power with Kryothermal S at 0°C1,5 kW
Cooling power with Kryothermal S at -20°C1,5 kW
Cooling power with Kryothermal S at -40°C1,5 kW
Cooling power with Kryothermal S at -60°C1,4 kW
Cooling power with Kryothermal S at -80°C1,4 kW
Cooling power with Kryothermal S at -100°C1 kW
Cooling power with Kryothermal S at -120°C0,2 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,25 kg
CO2 equivalent2,7 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,785 kg
CO2 equivalent 2nd stage11,6 t

Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Kryothermal S' в оригинале разбиты на подстроки по температурам.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 16.5 A maximum current and 3x20 A fusing, so the site electrical infrastructure must be sized accordingly.

Determines the electrical supply, cable and breaker rating needed at the installation site.

Source
“Power supply requirement 400V 3~ 50Hz max. current 16,5 A Fuse 3x20A”
Installation

The cooling water circuit needs at least 3 bar differential pressure and must not exceed 6 bar, with G1/2 male connections.

Defines the required cooling water supply pressure and connection type for proper installation.

Source
“min. cooling water differential pressure 3 bar max. cooling water pressure 6 bar”
Installation

The unit is designed for cooling water intake between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.

Prevents condensation and ensures the specified cooling performance is achievable.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects achievable cooling capacity and cooling water supply sizing at the site.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Avoids surprises during commissioning because the customer must supply the mains plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Compliance

Voltage and frequency tolerances are only valid in combination: ±5% voltage with ±2% frequency, and opposite-sign deviations are not allowed.

Ensures the supply quality at the installation site is within the permitted operating envelope.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant is R-452A (A1, H280) with a GWP of 2141 and a CO2 equivalent of 2.7 t, which matters for refrigerant handling and environmental reporting.

Relevant for refrigerant safety classification, leak response and CO2-equivalent reporting.

Source
“R-452A (A1, H280)”
Safety

The second and third refrigerant stages use R-23 and R-14 respectively, both classified A1 with H280, adding to the total CO2 equivalent.

Needed for full refrigerant inventory, safety classification and environmental compliance.

Source
“Refrigerant 2nd stage (ASHRAE, GHS) R-23 (A1, H280)”
Operation

The pump delivers 30 l/min maximum and 0.9 bar maximum pressure, with flow falling to 8 l/min at 0.8 bar, so external circuit pressure drop must be checked.

Determines whether the pump can supply the external application's flow and pressure requirements.

Source
“max. delivery 30 l/min max. delivery pressure 0,9 bar”
Operation

The minimum filling capacity is 3.6 l and the expansion tank holds 8 l, which sets the heat transfer fluid charge and expansion allowance.

Guides heat transfer fluid selection and filling volume during commissioning.

Source
“min. filling capacity 3,6 l Filling capacity expansion tank 8 l”
Installation

The unit weighs 340 kg and measures 700x804x1520 mm, so floor loading and access clearance must be planned.

Needed for transport, rigging and installation space planning.

Source
“Overall dimensions WxDxH ** 700x804x1520 mm Net weight 340 kg”
Operation

The ambient temperature must stay between 5 °C and 40 °C, and the unit is rated IP20, so it must be installed in a dry, protected location.

Defines the permissible room conditions and protection against dust and water ingress.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Compliance

The equipment is classified Class A, Group 1 for electromagnetic disturbance to EN55011, which has implications for the installation environment.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The unit is a water-cooled refrigerated heating circulator with a stainless steel pump and cooled shaft seal, suitable for externally closed and open applications.

Confirms the application suitability and cooling medium requirement at the selection stage.

Source
“Refrigerated Heating Circulator with water-cooled cooling machine and optical level indicator.”