Unistat 620w Technical Data Sheet

Contents

  1. Overview p. 1
    Product introduction and general description of the Unistat 620w unit
  2. Spec pp. 2-3
    Technical data according to DIN 12876 including performance and dimensions
  3. p. 2 table, 30 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C760 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions Temperature Control Unit WxDxH918x963x1771 mm
Dimensions of External Transformer WxDxH364x740x682 mm
Net weight616 kg
Weight External Transformer200 kg
sound pressure level +/- 4 dB(A)68 dB(A)
Power Supply for Temperature Control Unit460V 3~ 60Hz
Input Voltage External Transformer208V 3~ 60Hz
Output Voltage External Transformer440V 3~ 60Hz
Max. current temperature control unit25,4 A
Max. current temp. control unit with external transformer55 A
Fuse temperature control unit3X32 A
Fuse input voltage external transformer3X60 A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S607684

Delivery at 1,0 bar; Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Technical data according to DIN 12876 — Unistat 620w (Order-No.: 5016.0104.01)

ParameterValue
Operating temperature range-60...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C12 kW
Cooling power with Ethanol at -20°C12 kW
Cooling power with Ethanol at -40°C5,6 kW
Cooling power with Ethanol at -60°C1,4 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
UN-numberUN 2857
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a water-cooled cooling machine, so a cooling water supply with adequate differential pressure must be planned at the installation site.

Determines whether cooling water infrastructure is needed before selecting/installing the unit.

Source
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”
Installation

Cooling water inlet temperature must stay between 5°C and 20°C, and pipes must be insulated if water can drop below the condensation point.

Defines cooling water supply limits and condensation/insulation requirements for installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops as cooling water temperature rises, so the specified 12 kW figures are only valid at the stated reference conditions.

Prevents overestimating cooling capacity when site cooling water is warmer than 15°C.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and material compatibility decisions for the cooling loop.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance

Voltage and frequency tolerances are ±5% and ±2% respectively, but they may not be combined in opposite directions.

Defines acceptable mains quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

Leakage current exceeds 3.5 mA, which affects protective earthing and RCD selection.

Critical for electrical installation and choosing appropriate residual current protection.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so a plug must be sourced separately.

Affects electrical installation planning and procurement of the mains connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The temperature control unit is fed from an external transformer, so both units and their combined current must be accommodated.

Requires space and electrical planning for the transformer in addition to the main unit.

Source
“the required voltage is provided using a external transformer (supplied as standard)”
Selection

The unit is classified as pressure equipment category I, which matters for compliance documentation.

Relevant for regulatory compliance assessment during selection.

Source
“Pressure equipment category”
Operation

Ambient temperature must remain between 5°C and 40°C for correct operation.

Defines the allowable room conditions at the installation site.

Source
“max. ambient temperature”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.

Restricts installation location and environmental protection requirements.

Source
“Degree of Protection”
Maintenance

The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 7.7 t, which creates leak-checking and reporting obligations.

Affects maintenance obligations and environmental compliance for the refrigeration circuit.

Source
“Global Warming Potential (GWP)”
Installation

Pump connections are M38x1.5 male and cooling water connections are G3/4 male, so matching hoses and fittings must be ordered.

Needed to specify compatible hoses and adapters before installation.

Source
“Pump connection”
Selection

The unit needs a minimum filling capacity of 8.9 l and has a 19.9 l expansion tank, which sets the heat transfer fluid volume to prepare.

Determines the required heat transfer fluid quantity for commissioning.

Source
“min. filling capacity”