Peter Huber Kältemaschinenbau Unistat 620w Dynamisches
Peter Huber Kältemaschinenbau ·Made in Germany · In Active Production

Peter Huber Kältemaschinenbau Unistat 620w Dynamisches

Peter Huber Kältemaschinenbau Unistat 620w Dynamisches — Temperature control units with Noise level 68 dB.

The Unistat 620w Dynamisches is a water-cooled refrigerated heating circulator with a magnetically coupled stainless-steel pump, covering -60 to 200 °C with 12 kW heating capacity. It is the largest water-cooled dynamic modification in the Unistat 5xx/6xx series, distinguished by its external transformer and G3/4 cooling-water connection. It serves as a high-capacity temperature source for externally closed process loops requiring precise Pt100 control via the Pilot ONE controller.

Noise level
68 dB

Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionHeating capacity, WPump connectionPump typeTemperature range, °C
Unistat 620w Dynamisches Current 680, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat 520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat 527w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 530w Dynamisches 590, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat 610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat 615w Dynamisches 620, 0112000M30x1, 5 AG malePressure pump, variable -60–200
Unistat 625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 640w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat 645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 650w Dynamisches 730, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat 680w Dynamisches —0, 0196000DN50 PN16 (DIN EN 1092-1magnetically coupled, st-60–200
Unistat GL 530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL 570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P530w Dynamisches 610, 016000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P535w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P540w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P550w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P560w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat GL P570w Dynamisches —0, 0148000M38x1, 5 AG malemagnetically coupled, st-50–200
Unistat P510 Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P510w Dynamisches 600, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P515w Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-50–250
Unistat P520 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P520w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525 Dynamisches 560, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P525w Dynamisches 610, 016000M30x1, 5 AG malemagnetically coupled, st-55–250
Unistat P527w Dynamisches 570, 0112000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P530w Dynamisches 620.0112000M38x1,5 AG malemagnetically coupled, st-55–250
Unistat P540w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P545w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-55–250
Unistat P610 Dynamisches —0, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P610w Dynamisches 620, 016000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615 Dynamisches —0, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P615w Dynamisches 620, 0112000M30x1, 5 AG malemagnetically coupled, st-60–200
Unistat P620w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P625w Dynamisches —0, 0112000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P630w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P635w Dynamisches —0, 0124000M38x1, 5 AG malemagnetically coupled, st-60–200
Unistat P645w Dynamisches —0, 0136000flange DN32 PN10/40 EN10magnetically coupled, st-60–200
Unistat P650w Dynamisches —0, 0148000flange DN32 PN10/40 EN10magnetically coupled, st-60–200

Full specifications

Core Performance & Functional Specifications
Noise level 68 dB(A)
Control output POKO ONE
Heating capacity 12 kW
Pump connection M38x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -60...200 °C
Dimensions, Weight & Mechanical Design
Weight 616 kg
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 100 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G3/4 male
Cooling water consumption 760 l/h (at 15°C, flow 0°C)
Cooling water differential pressure 1 bar
Cooling Water Pressure 6 bar
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Mounting
Install only in a dry indoor area outside ATEX zones

The IP20 rating and the manual's prohibition of ATEX-zone installation mean the unit must be placed in a dry, protected indoor environment. Explosion protection would require a separate Ex px cabinet with its own ATEX annex and rating plate; the standard unit is not protected against explosion.

Operation Manual, Section 1.3.3 / page 15-16; Datasheet, Degree of Protection
Electrical
Leakage current exceeds 3.5 mA; plan RCD and earthing accordingly

The datasheet states leakage current >3.5 mA, which affects protective earthing and residual current device selection. The unit is fed from an external transformer supplied as standard, so both the transformer and the unit must be accommodated and their combined current considered. Three-phase devices below 63 A are delivered with a cable but without a plug.

Datasheet, page 2 (leakage current, external transformer, cable/plug
Process Limits
Cooling water must stay between 5 and 20 °C at the inlet

The datasheet specifies cooling water intake temperature of 5–20 °C. Cooling power drops as cooling water temperature rises above the 15 °C reference, and flow rate may need to increase. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.

Datasheet, page 2 (cooling water temperature, materials, performance note
Maintenance
Shut off cooling water even for overnight downtime

The manual directs shutting off the cooling water supply even for short downtimes such as overnight to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed. Refrigerant-circuit work must be performed by approved refrigeration/air-conditioning contractors.

Operation Manual, page 29 (cooling water shutoff, hose inspection); Section 1.4.2 / page 19

Selection & Integration Q&A

When should I choose the Unistat 620w Dynamisches over the smaller 510w or 520w water-cooled siblings?

Choose the 620w when you need the largest water-cooled dynamic capacity in the series: 12 kW heating, -60 to 200 °C, 616 kg mass and 19.9 l expansion tank. The 510w and 520w have lower noise levels (60/61 dB(A)) but smaller capacity; select them for lower heat loads or tighter acoustic limits.

What cooling water supply and connections must be provided for this unit?

Provide cooling water at 5–20 °C inlet, max 6 bar pressure, 1 bar differential pressure, with G3/4 male connections. Consumption is 760 l/h at 15 °C water and 0 °C flow. Insulate pipes if below condensation point; use only water compatible with copper, 1.4401, MS, PA, PPE, PTFE and EPDM.

What are the ambient and duty-cycle limits for installing this unit?

Ambient must be 5–40 °C, IP20 dry indoor, outside ATEX zones. Transport upright only and allow ~2 h acclimation after temperature changes before power-on. A drip tray is required below the unit. Medical-device and direct-food applications are not permissible.

What maintenance items and accessories are not included in the base delivery?

The base delivery excludes a mains plug for three-phase units below 63 A, hoses and fittings for M38x1.5 pump and G3/4 cooling water connections, and the drip tray. Heat transfer fluid (min 8.9 l) must be supplied separately. Periodic hose inspection and cooling water shutoff are operator tasks; refrigerant work requires certified contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat 620w Technical Data Sheet
Datasheet ·20,0 КБ ·3 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF