Unistat P430 Temperature Control System Datasheet

Contents

  1. Overview p. 1
    Product introduction and general description of Unistat P430
  2. Spec pp. 1-2
    Technical data, performance and operating specifications per DIN 12876
  3. p. 2 table, 18 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar63 l/min
Delivery at 2,0 bar48 l/min
Delivery at 2,5 bar30 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity4 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x570x1629 mm
Net weight197 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)15 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S611262

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P430 Technical data according to DIN 12876

ParameterValue
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power4 kW
Cooling power with Thermooil at 250°C3,5 kW
Cooling power with Thermooil at 200°C3,5 kW
Cooling power with Thermooil at 100°C3,5 kW
Cooling power with Thermooil at 20°C3,5 kW
Cooling power with Ethanol at 0°C3,5 kW
Cooling power with Ethanol at -20°C2 kW
Cooling power with Ethanol at -40°C0,15 kW
Refrigeration machineair-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity0,75 kg
CO2-Äquvivalent0,0 t
UN-numberUN 3358
Circulation pump: MK pump max. delivery97 l/min
Circulation pump: MK pump max. delivery pressure3 bar
Circulation pump: MK pump Delivery at 0,5 bar86 l/min
Circulation pump: MK pump Delivery at 1,0 bar75 l/min

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.

Affects installation planning and the bill of materials for the electrical connection.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which may require a suitable residual current device or protective earthing arrangement.

Important for electrical safety design and RCD selection at the installation site.

Source
“Attention: leakage current > 3,5mA”
Installation

The pump connections use M30x1.5 male threads with Y-bore and ball-socket fittings per DIN 3863/3870, so matching hoses and fittings must be ordered.

Ensures correct mating of process pipework and flexible tempering hoses.

Source
“Pump connection M30x1,5 male”
Operation

The unit must be operated between 5 °C and 40 °C ambient, and cooling capacity may drop above 20 °C room temperature.

Affects site ventilation and expected performance in warm rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Performance

Cooling capacity falls sharply at low temperature with ethanol, from 3.5 kW at 0 °C to only 0.15 kW at -40 °C.

Critical for sizing the unit for low-temperature applications near the lower limit.

Source
“at -40°C 0,15 kW”
Selection

The pump delivers 97 l/min maximum but only 30 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve checking.

Determines whether the circulation pump can supply the required flow through restrictive external circuits.

Source
“Delivery at 2,5 bar 30 l/min”
Selection

The refrigerant is R-1270 (A3, H220) with GWP 0 and 0.75 kg charge, which is flammable and requires appropriate site precautions.

Flammable refrigerant classification affects room safety, ventilation and insurance requirements.

Source
“R-1270 (A3, H220)”
Installation

The unit is IP20 and weighs 197 kg with dimensions 460x570x1629 mm, so it needs a dry indoor location and adequate floor space and handling equipment.

Needed for site layout, access routes and floor loading assessment.

Source
“Overall dimensions WxDxH ** 460x570x1629 mm”
Operation

Minimum filling capacity is 4 l and the expansion tank holds 3.7 l, so the external loop volume must be checked against these limits.

Prevents underfilling or overflow of the expansion tank in external circuits.

Source
“min. filling capacity 4 l”
Compliance

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Must be verified against the site supply quality to avoid voiding performance or warranty.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.

Relevant for protecting temperature-sensitive process fluids and meeting safety requirements.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The controller supports external Pt100 sensors and internal/process control modes, enabling control at the external process point.

Allows accurate control at the application rather than at the circulator outlet.

Source
“selectable temperature control mode (Internal/Process)”