Peter Huber Kältemaschinenbau Unistat P430 Dynamisches

Peter Huber Kältemaschinenbau Unistat P430 Dynamisches — Temperature control units with Noise level 65 dB.

The Unistat P430 Dynamisches is a free-standing, air-cooled temperature control unit with a magnetically coupled stainless-steel pump and Pilot ONE controller, covering -40 to 250 °C with 4 kW heating. It is the dynamic-pressure variant (P) of the Unistat 4xx series, intended for externally closed circuits with high pressure drop, delivering 30 l/min at 2.5 bar.

Noise level
65 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat P430 Dynamisches Current 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430w Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Noise level 65 dB(A)
Control output POKO ONE
Filling volume 4 l
Heating capacity 4 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -40...250 °C
Dimensions, Weight & Mechanical Design
Weight 197 kg
Electrical, Power & Drive Specifications
Current consumption 15 A
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Thermal and hydraulic behavior

Installation and operation

Integration & Operating Notes

Electrical
Supply must be 400 V 3~ 50 Hz with 3×16 A fuse

The unit is factory-configured for three-phase 400 V 50 Hz and requires a 3×16 A fuse. Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide and fit a suitable plug. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.

Datasheet, page 2
Electrical
Leakage current exceeds 3.5 mA — plan RCD accordingly

The datasheet states leakage current >3.5 mA. Site electrical design must account for this with a suitable residual current device or protective earthing arrangement; standard 30 mA RCDs may not be appropriate without further assessment.

Datasheet, page 2
Mounting
Never install or start inside an ATEX zone

The manual explicitly prohibits installation or start-up within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number stated on the Ex px cabinet rating plate is suitable. Site selection must exclude hazardous areas.

Operation Manual, Section 1.3.3
Mounting
Transport upright and empty; acclimatize 2 h before power-on

Only transport the unit upright and emptied; horizontal transport damages the compressor. Casters are not for transport and carry only 25% of total mass symmetrically — use an industrial truck. After moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults.

Operation Manual, page 22 and page 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON to keep cooling active at the OT trigger. Configure accordingly for exothermic process protection.

Operation Manual, Section 1.7.2 and page 19
Maintenance
Shut off cooling water even for overnight downtimes

The manual directs shutting off the cooling water supply even for shorter downtimes such as overnight, to prevent water damage and unnecessary load on the cooling water circuit. Hoses and connections must be periodically checked for leaks and quality and replaced as needed.

Operation Manual, page 29
Maintenance
Install a drip tray below the unit; local rules may mandate it

The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit; national legislation may prescribe it. Installation planning must include the drip tray and verification of local regulations.

Operation Manual, Section 1.4

Selection & Integration Q&A

When should I choose the Unistat P430 Dynamisches instead of a sibling Grande Fleur variant?

Choose the P430 when your externally closed circuit has high pressure drop and you need dynamic temperature changes; the 'P' pump delivers 30 l/min at 2.5 bar. Sibling Grande Fleur variants differ in cooling/refrigerant configuration — select those only if their specific cooling or media configuration matches your process, not as a substitute for high-pressure capability.

What adjacent system parts and connections must I provide for the P430?

Provide a 400 V 3~ 50 Hz supply with 3×16 A fuse and a suitable plug (cable supplied, plug not). Pump connections are M30×1.5 male with Y-bore and ball-socket per DIN 3863/3870 — order matching hoses and fittings. Ensure RCD/protective earthing handles >3.5 mA leakage. External Pt100 sensor and Ethernet/USB/RS232 interfaces are available for control integration.

What are the mounting, environment and duty-cycle limits for this unit?

Install indoors, dry, IP20, ambient 5…40 °C, never in an ATEX zone. Transport upright and empty; use an industrial truck, not casters. Allow 2 h acclimation after temperature changes. Leveling feet must be activated and parking brakes engaged. Cooling capacity drops above 20 °C ambient and at low temperatures; size against the curve.

What maintenance schedule and accessories/consumables are required but not included in base delivery?

Base delivery excludes plug, mating hoses/fittings, drip tray, and external Pt100 sensor. Periodically check hoses and connections for leaks and quality and replace as needed; shut off cooling water overnight. Clean liquefier fins and hat-type strainer/dirt trap; inspect mechanical seal and plug contacts. Heat transfer fluid must be inspected/replaced per manual; refrigerant work only by approved contractors.

Dimensional Drawings & Technical Documentation

Technical documentation

PDF
Unistat P430 Temperature Control System Datasheet
Datasheet ·31,1 КБ ·2 pp.
View PDF
PDF
Unistat® Operation Manual — Unistat® 5, Unistat® Pilot ONE®
Manual ·3,6 МБ ·100 pp.
View PDF
PDF
Unistat Operation Manual
Manual ·3,7 МБ ·102 pp.
View PDF