Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and protection device rating needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit needs at least 1 bar differential pressure and must not exceed 6 bar, so the building water supply must be checked.
Insufficient or excessive cooling water pressure will prevent proper operation or damage the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Out-of-range cooling water temperature reduces cooling power and risks condensation damage.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops sharply at low temperatures with ethanol, from 3.5 kW at 0 °C to only 0.15 kW at -40 °C.
Applications near the lower temperature limit must account for very low available cooling capacity.
Source
“at -40°C 0,15 kW”
Selection
The pump delivers 97 l/min at zero head but only 30 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve matching.
Ensures the selected pump can meet the flow demand of the external circuit at its actual pressure drop.
Source
“max. delivery 97 l/min”
Safety
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, which affects site safety classification and ventilation requirements.
Flammable refrigerant requires appropriate room ventilation and ignition-source control during installation and operation.
Source
“R-1270 (A3, H220)”
Safety
Leakage current exceeds 3.5 mA, so protective earthing and residual current device selection must be verified.
High leakage current affects RCD type/rating and earthing design for safe operation.
Source
“Attention: leakage current > 3,5mA”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not combine in the unfavorable direction.
Site power quality must be checked to avoid operating outside the allowed supply envelope.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.
Prevents delays at commissioning because the customer must supply the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.
Defines the permissible installation location and protection against dust and water ingress.
Source
“Degree of Protection IP20”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Incompatible water chemistry can corrode or foul the cooling circuit and reduce performance.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits fluid choice at low temperatures.
Selecting a fluid that is too viscous at low temperature will overload the pump and reduce flow.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, so matching hoses and fittings must be ordered.
Ensures correct mating fittings and avoids leaks or delays during plumbing.
Source
“Pump connection M30x1,5 male”
Compliance
The equipment is EMC class A group 1, which is intended for industrial environments and may cause interference in residential areas.
Affects where the unit may be installed and what EMC measures may be needed.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”