Peter Huber Kältemaschinenbau Unistat P430w Dynamisches

Peter Huber Kältemaschinenbau Unistat P430w Dynamisches — Temperature control units with Noise level 65 dB.

The Unistat P430w Dynamisches is a water-cooled, high-pressure-drop circulating temperature control unit with a magnetically coupled stainless-steel pump, covering -40 to 250 °C with 4 kW heating. Its 'P' designation targets externally closed loops with high pressure drop, where the pump delivers 30 l/min at 2.5 bar. It serves precise temperature control of reactors, calorimeters and pilot-scale process equipment.

Noise level
65 dB

Peter Huber Kältemaschinenbau Unistate Bis 45C Series — Model Comparison & Selection

Model Noise level, dBDisplay resolutionPower supplyPump connectionPump typeTemperature range, °C
Unistat P430w Dynamisches Current 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Grande Fleur Dynamisches ——————
Grande Fleur w Dynamisches —0, 01208-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -40–200
Grande Fleur w-eo Dynamisches ——————
Petite Fleur-eo Dynamisches 580, 01 K220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Petite Fleur w Dynamisches —0, 01220-240V 1~/2~ 50/60HzM16x1 AG malePressure pump, variable -40–200
Unistat 410w Dynamisches 610, 01 K220-240V 1~/2~ 50/60HzM24x1, 5 AG malePressure pump, variable -45–250
Unistat 425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat 430w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P405w Dynamisches —0, 01460V 3~ 60HzM30x1, 5 AG malemagnetically coupled, st-45–250
Unistat P425 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P425w Dynamisches 610, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250
Unistat P430 Dynamisches 650, 01400V 3~ 50HzM30x1, 5 AG malemagnetically coupled, st-40–250

Full specifications

Core Performance & Functional Specifications
Noise level 65 dB(A)
Alarm output optic, acoustic, relay
Control output POKO ONE
Filling volume 4 l (min.)
Heating capacity 4 kW
Pump connection M30x1, 5 AG male
Pump speed adjustable Yes
Pump type magnetically coupled, stainless steel
External sensor connection Pt100
Temperature range -40...250 °C
Dimensions, Weight & Mechanical Design
Weight 182 kg
Electrical, Power & Drive Specifications
Power supply 400V 3~ 50Hz
Interfaces, Connections & Control
Controller Pilot ONE
Display resolution 0, 01 K
Digital interface Ethernet, USB (Host u. Device), RS232
Materials, Media & Process Compatibility
Maximum viscosity 50 mm²/s
Operating Conditions, Environment & Protection
Ambient temperature limit 5...40 °C
Cooling water connection G1/2 male
Cooling water differential pressure 1 bar (min.)
Cooling Water Pressure 6 bar (max.)
Protection rating IP20
Standards, Safety & General Data
Safety class III / FL

Description

Construction and series positioning

Behavior under load and media

Installation, control and operation

Integration & Operating Notes

Electrical
Supply must be 400 V three-phase with a 3x16 A fuse

The electrical installation must be sized for a 400 V 3~ 50 Hz supply protected at 3x16 A. Voltage tolerance is ±5% and frequency ±2%, but the two must not combine in the unfavorable direction; verify site power quality before connection.

Unistat P430w Technical Data, page 2
Electrical
Provide a mating plug on site; cable is supplied without one

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must supply and fit a suitable plug. Leakage current exceeds 3.5 mA, so protective earthing and RCD selection must be verified by the installing electrician.

Unistat P430w Technical Data, page 2
Cooling & Media
Cooling water must be 5–20 °C at 1–6 bar differential

The cooling water circuit requires at least 1 bar differential pressure and must not exceed 6 bar, with inlet temperature between 5 and 20 °C. Out-of-range water temperature reduces cooling power and risks condensation, so pipes must be insulated where condensation can occur.

Unistat P430w Technical Data, page 2
Process Limits
Fluid viscosity above 50 mm²/s overloads the pump

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s; selecting a fluid too viscous at low temperature will reduce flow and load the pump. For FL-rated fluid the fire point must be at least 65 °C, and ethanol is permitted only under supervised operation.

Unistat P430w Technical Data, page 2; Operation Manual, Section 1.7
Mounting
Do not install inside an ATEX zone

The unit must not be installed or started within an ATEX zone; explosion protection requires a separate Ex px cabinet and only the model/serial number on the Ex px rating plate is suitable. The refrigerant R-1270 (A3, H220) is flammable, so site ventilation and ignition-source control must be assessed.

Operation Manual, Section 1.3.3
Mounting
Transport upright and allow 2 h acclimation before power-on

Only transport the unit upright and emptied; horizontal transport damages the compressor. When moving between cold and warm environments, allow about 2 hours acclimation before switching on to prevent condensation-induced electrical faults. Leveling feet must be activated and casters are not for transport.

Unistat Operation Manual, page 22 and 29
Control & I/O
Process Safety mode requires compressor set to Permanently ON

In Process Safety mode the controller releases maximum cooling capacity at the overtemperature trigger instead of switching off all poles, but this mode requires the compressor to be set to Permanently ON. Configure the power-failure response (Off/Standby or ON) according to whether unattended restart is acceptable.

Operation Manual, Section 1.7.2 and 1.8.1
Maintenance
Shut off cooling water even for overnight downtime

The cooling water supply should be shut off even for short downtimes such as overnight to prevent water damage and unnecessary load on the circuit. Hoses and connections must be checked periodically for leaks and quality and replaced as needed; refrigerant-circuit work requires approved refrigeration contractors.

Unistat Operation Manual, page 29; Section 1.4.2

Selection & Integration Q&A

When should I choose the P430w Dynamisches instead of a sibling Unistat variant?

Choose the P430w when the external loop has high pressure drop: its pump delivers 30 l/min at 2.5 bar, whereas a standard pump would be throttled. For low-impedance loops or air-cooled sites, a non-P or non-w variant of the series is the better fit.

What must I provide on the cooling water and electrical side to integrate this unit?

Provide cooling water at 5–20 °C with 1–6 bar differential pressure through G1/2 male connections, and a 400 V 3~ 50 Hz supply fused at 3x16 A with a site-fitted plug. Verify RCD selection because leakage current exceeds 3.5 mA.

What mounting and duty-cycle limits apply from the manual?

Install indoors at 5–40 °C ambient, IP20, outside any ATEX zone, upright and level with feet activated. Allow 2 h acclimation after cold/warm moves. The refrigerant is flammable R-1270, so ventilation and ignition-source control are required.

What maintenance and accessories are not included in base delivery?

The three-phase plug, mating hoses/fittings for M30x1.5 pump connections and G1/2 cooling water, heat transfer fluid, and a drip tray are not included. Periodic hose inspection, cooling water shutoff at downtime, and refrigerant work by approved contractors are required.