Product introduction and general description of the Unistat P515w unit
Specpp. 1-2
Technical data and performance specifications according to DIN 12876
p. 2table, 21 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 1,5 bar
74 l/min
Delivery at 2,0 bar
55 l/min
Delivery at 2,5 bar
33 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
504 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
4,1 l
Filling capacity expansion tank
3,7 l
Overall dimensions WxDxH **
460x554x1453 mm
Net weight
181 kg
Power supply requirement
400V 3~ 50Hz
max. current
14,5 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S429820
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P515w — Technical data according to DIN 12876
Parameter
Value
Order-No.
1071.0004.01
Operating temperature range
-50...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
7 kW
Cooling power with Thermooil at 200°C
7 kW
Cooling power with Thermooil at 100°C
7 kW
Cooling power with Thermooil at 20°C
7 kW
Cooling power with Ethanol at 0°C
5,3 kW
Cooling power with Ethanol at -20°C
2,8 kW
Cooling power with Ethanol at -40°C
0,9 kW
Cooling power with Ethanol at -50°C
0,25 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,45 kg
CO2 equivalent
3,1 t
Circulation pump:
MK pump
max. delivery
119 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
106 l/min
Delivery at 1,0 bar
91 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable and protection device needed before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.
Cooling water supply pressure must be verified to avoid insufficient cooling or overpressure damage.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.
Defines the permissible cooling water conditions and condensation protection required at the site.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure, so site conditions affect achievable performance.
Performance figures must be corrected if site cooling water or ambient conditions differ from the rating basis.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and -”
Performance
Higher cooling water temperature reduces cooling power and may require a higher cooling water flow rate.
Prevents overestimating cooling capacity when the available cooling water is warmer than the rating basis.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides selection of cooling water quality and treatment to avoid corrosion or deposits.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Compliance
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Defines acceptable supply quality limits for reliable operation and warranty compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety
The equipment has a leakage current greater than 3.5 mA, which affects protective earthing and RCD selection.
Requires suitable earthing and residual current protection design for safe installation.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure a matching plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections use M30x1.5 male threads with bore shape Y (60°) per DIN 3863, and hoses must use ball socket and sleeve nut per DIN 3870.
Ensures correct selection of hoses and fittings for leak-free connection.
Source
“Pump connection M30x1,5 male”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.
Limits the choice of heat transfer fluid and operating temperature range to protect the pump.
Source
“max. permissible kin. viscosity 50 mm²/s”
Installation
The unit must be operated within an ambient temperature range of 5 °C to 40 °C.
Defines the required room conditions and space planning for reliable operation.
Source
“min. ambient temperature 5 °C”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III/FL.
Confirms built-in safety protection relevant for risk assessment and compliance.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection
The P models use circulation pumps with high discharge pressure for applications with high pressure drops.
Helps confirm the pump is suitable for high-pressure-drop external circuits.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops”