No photo available yet
Peter Huber Kältemaschinenbau Unistat P515w Dynamisches
Peter Huber Kältemaschinenbau Unistat P515w Dynamisches — Temperature control units.
The Unistat P515w Dynamisches is a water-cooled, high-discharge-pressure circulating temperature control unit with a -50 to 250 °C working range and 6 kW heating capacity. Its magnetically coupled stainless-steel pump and M30x1.5 connections distinguish it from air-cooled and standard-pressure siblings in the Unistat 5xx series. It serves processes with high pressure-drop external circuits requiring precise, dynamic temperature control.
Peter Huber Kältemaschinenbau Unistat Series — Model Comparison & Selection
| Model | Noise level, dB | Display resolution | Heating capacity, W | Pump connection | Pump type | Temperature range, °C |
|---|---|---|---|---|---|---|
| Unistat P515w Dynamisches Current | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 515w Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat 520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 527w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 530w Dynamisches | 59 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat 610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | Pressure pump, variable | -60–200 |
| Unistat 620w Dynamisches | 68 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 640w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat 645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 650w Dynamisches | 73 | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat 680w Dynamisches | — | 0, 01 | 96000 | DN50 PN16 (DIN EN 1092-1 | magnetically coupled, st | -60–200 |
| Unistat GL 530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL 570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P530w Dynamisches | 61 | 0, 01 | 6000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P535w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P540w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P550w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P560w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat GL P570w Dynamisches | — | 0, 01 | 48000 | M38x1, 5 AG male | magnetically coupled, st | -50–200 |
| Unistat P510 Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P510w Dynamisches | 60 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -50–250 |
| Unistat P520 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P520w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525 Dynamisches | 56 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P525w Dynamisches | 61 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P527w Dynamisches | 57 | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P530w Dynamisches | 62 | 0.01 | 12000 | M38x1,5 AG male | magnetically coupled, st | -55–250 |
| Unistat P540w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P545w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -55–250 |
| Unistat P610 Dynamisches | — | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P610w Dynamisches | 62 | 0, 01 | 6000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615 Dynamisches | — | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P615w Dynamisches | 62 | 0, 01 | 12000 | M30x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P620w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P625w Dynamisches | — | 0, 01 | 12000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P630w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P635w Dynamisches | — | 0, 01 | 24000 | M38x1, 5 AG male | magnetically coupled, st | -60–200 |
| Unistat P645w Dynamisches | — | 0, 01 | 36000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
| Unistat P650w Dynamisches | — | 0, 01 | 48000 | flange DN32 PN10/40 EN10 | magnetically coupled, st | -60–200 |
Full specifications
| Control output | POKO ONE |
| Heating capacity | 6 kW |
| Pump connection | M30x1, 5 AG male |
| Pump speed adjustable | Yes |
| Pump type | magnetically coupled, stainless steel |
| External sensor connection | Pt100 |
| Temperature range | -50...250 °C |
| Weight | 181 kg |
| Power supply | 400V 3~ 50Hz |
| Controller | Pilot ONE |
| Display resolution | 0, 01 K |
| Digital interface | Ethernet, USB (Host u. Device), RS232 |
| Maximum viscosity | 50 mm²/s |
| Ambient temperature limit | 5...40 °C |
| Cooling water connection | G1/2 male |
| Cooling water consumption | 504 l/h (at 15°C, flow 0°C) |
| Cooling water differential pressure | 1 bar (min) |
| Cooling Water Pressure | 6 bar (max) |
| Protection rating | IP20 |
| Safety class | III / FL |
Description
Construction and series positioning
- Water-cooled (w) free-standing Unistat 5xx platform with Pilot ONE controller and POKO ONE control output.
- The 'P' designation denotes a circulation pump with high discharge pressure for applications with high pressure drops; the 'Dynamisches' suffix indicates the dynamic temperature control variant.
- Sibling models in the series (510w, 515w, 520w, 525) differ in cooling capacity and noise level; this P515w is the high-pressure-drop modification.
- Pump: magnetically coupled, stainless steel, speed-adjustable; pump connection M30x1.5 male with bore shape Y (60°) per DIN 3863.
- Safety classification III/FL with adjustable overtemperature protection per DIN 12876; external Pt100 sensor input.
Working range and media behavior
- Temperature range -50 to 250 °C; heating capacity 6 kW.
- Maximum permissible kinematic viscosity of heat transfer fluid: 50 mm²/s.
- Cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM; only compatible cooling water may be used.
- Cooling capacity is rated at 20 °C room temperature, 15 °C cooling water inlet and 1 bar differential pressure; higher cooling water temperature reduces cooling power and may require increased flow.
- Cooling water consumption 504 l/h at 15 °C, flow 0 °C.
Installation, connection and control
- Power supply 400 V 3~ 50 Hz with 3x16 A fuse; three-phase devices below 63 A are delivered with cable but without plug.
- Voltage tolerance ±5% with simultaneous frequency tolerance ±2%; leakage current >3.5 mA affects protective earthing and RCD selection.
- Cooling water connection G1/2 male; minimum differential pressure 1 bar, maximum pressure 6 bar; intake temperature 5–20 °C; insulate pipes if condensation can occur.
- Ambient temperature 5–40 °C; protection rating IP20; weight 181 kg.
- Interfaces: Ethernet, USB (Host and Device), RS232; display resolution 0.01 K.
- Pilot ONE controller supports Process Safety mode (requires compressor set to Permanently ON), Power Failure Auto restart, and adjustable OT protection.
Integration & Operating Notes
The unit requires a 400 V three-phase 50 Hz supply protected at 3x16 A. Three-phase devices below 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site. Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Cooling water intake temperature must stay between 5 °C and 20 °C, with minimum differential pressure 1 bar and maximum pressure 6 bar. Higher cooling water temperature reduces cooling power and may require increased flow rate. Pipes must be insulated if condensation can occur. The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water may be used.
The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s. For FL-rated (flammable) fluid, the fluid must have a fire point of ≥65 °C; if ethanol is used, only supervised operation is permitted. Thorough venting and correct fill level are mandatory to avoid low-level protection interrupting circulation.
The responsible body must install a condensation water / heat transfer fluid drip tray below the unit; national legislation may prescribe it. The unit must NOT be installed or started up within an ATEX zone; explosion protection requires an Ex px cabinet. Ambient temperature must be 5–40 °C. Transport and store upright only; allow about 2 hours acclimation when moving between cold and warm environments before power-on.
In Process Safety mode, when the set overtemperature trigger value is reached the controller releases maximum cooling capacity instead of switching off all poles. This mode requires the compressor to be set to Permanently ON. Power Failure Auto resumes temperature control automatically after a power outage if it was active; choose power-interruption behavior according to whether unattended restart is acceptable.
Shut off the cooling water supply to the unit even for shorter downtimes (e.g. overnight) to prevent water damage and unnecessary load on the cooling water circuit. Check periodically for leaks and the quality of the hose and hose connections and replace as required. Work on the coolant circuit must be carried out by approved refrigeration/air-conditioning system contractors.
Selection & Integration Q&A
Choose the P515w when the external circuit has high pressure drop requiring the high-discharge-pressure pump. The 510w and 520w differ in cooling capacity and noise level; select those if pressure drop is moderate and their capacity matches the load.
Cooling water: G1/2 male connection, 5–20 °C intake, 1 bar min differential pressure, 6 bar max, 504 l/h at 15 °C. Process loop: M30x1.5 male pump connections with bore shape Y (60°) per DIN 3863; hoses must use ball socket and sleeve nut per DIN 3870.
Periodically check hoses and connections for leaks and quality; shut off cooling water overnight; clean liquefier fins and strainer. A plug for the 400 V cable is not included; a drip tray and compatible heat transfer fluid must be provided on site.