Product introduction and general description of the Unistat P545w unit
Specp. 2
Technical data according to DIN 12876 including performance parameters
p. 2table, 21 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
Delivery at 2,0 bar
154 l/min
Delivery at 3,0 bar
124 l/min
Delivery at 4,0 bar
87 l/min
Delivery at 5,0 bar
42 l/min
Pump connection
M38x1,5 male
max. permissible kin. viscosity
100 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
1630 l/h
min. cooling water differential pressure
1 bar
max. cooling water differential pressure
4 bar
max. cooling water pressure
6 bar
max. System pressure
6 bar
min. filling capacity
9,9 l
Filling capacity expansion tank
19,9 l
Overall dimensions WxDxH **
730x804x1738 mm
Power supply requirement
400V 3~ 50Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
1.0/23
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876
Характеристика
Unistat P545w
Order-No.
5012.0002.01
Operating temperature range
-55...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Interface analog
optional
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
24 kW
Cooling power with Thermooil at 250°C
35 kW
Cooling power with Thermooil at 200°C
35 kW
Cooling power with Thermooil at 100°C
35 kW
Cooling power with Thermooil at 20°C
35 kW
Cooling power with Ethanol at 0°C
32 kW
Cooling power with Ethanol at -20°C
16 kW
Cooling power with Ethanol at -40°C
4 kW
Cooling power with Ethanol at -50°C
1,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
3,6 kg
CO2 equivalent
7,7 t
Circulation pump:
MK pump
max. delivery
201 l/min
max. delivery pressure
5,3 bar
Delivery at 1,0 bar
179 l/min
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a water-cooled refrigerated heating circulator with 24 kW heating and up to 35 kW cooling, suitable for externally closed and open applications.
Confirms the unit type and that it can serve both closed and open external circuits when selecting for a process.
Source
“Refrigerated Heating Circulator with water-cooled cooling machine and electronical level indicator.”
Selection
The 'P' model pump is designed for high pressure drops, making it suitable for flow-through chemistry or semiconductor applications.
Guides selection toward this model when the external circuit has high pressure drop.
Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
Cooling water must be supplied between 5 °C and 20 °C, and pipes must be insulated if water temperature can fall below the condensation point.
Defines the required cooling water supply conditions and condensation protection for installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation
The cooling water circuit requires a differential pressure between 1 bar and 4 bar, with a maximum cooling water pressure of 6 bar.
Ensures the site cooling water supply meets the minimum and maximum pressure limits for correct operation.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit requires a 400 V three-phase 50 Hz supply and draws high power, so the electrical supply must be sized accordingly.
Critical for electrical planning and ensuring the site provides the correct three-phase supply.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Prevents installation on a supply where the combined voltage/frequency deviation exceeds the permitted envelope.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices with current consumption less than 63 A are delivered with cable but without a plug.
The installer must provide and fit a suitable plug or connection for this three-phase unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
Cooling power drops as cooling water temperature rises, and cooling water flow rate may need to increase.
Operators must account for reduced cooling capacity if the site cooling water is warmer than the rated 15 °C.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Operation
The pump delivers 201 l/min maximum and 5.3 bar maximum pressure, with delivery falling as pressure rises.
Allows checking whether the pump curve suits the external circuit's flow and pressure drop requirements.
Source
“max. delivery 201 l/min”
Operation
The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.
Prevents selecting a fluid too viscous for the pump and circuit at operating temperature.
Source
“max. permissible kin. viscosity 100 mm²/s”
Safety
The refrigerant R-452A is classified A1 with GHS hazard H280, and the unit contains 3.6 kg with a CO2 equivalent of 7.7 t.
Needed for refrigerant handling, leak reporting and environmental compliance at the installation site.
Source
“R-452A (A1, H280)”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms the safety device required for unattended operation and for protecting the external application.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation
The unit must be operated within an ambient temperature range of 5 °C to 40 °C.
Ensures the installation room stays within the allowed ambient conditions for reliable operation.
Source
“max. ambient temperature 40 °C”
Compliance
The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.
Class A equipment may require EMC precautions in residential or light-industrial environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”