The pump must be kept within -20 to +60 °C and 0–95% non-condensing RH during transport and storage.
Sets the environmental limits for shipping and warehousing before installation.
Source
“Temperatur: -20 °C bis +60 °C (-4 °F bis 140 °F)”
| Model | IDP-10 Dry Scroll Single Hermetic Vacuum Pump |
|---|---|
| Interface dimensions | H=297 mm x W=260 mm x L= 420mm |
| Peak pumping speed | 170L/min (10.2 m3/h) |
| Media | No corrosive, explosive or particulate forming gases |
| Ultimate pressure* | 1.5 x 10-2 Torr (2.0 x 10-2 mbar) |
| Maximum recommended continuous inlet pressure | 300 Torr (400 mbar) |
| Maximum inlet pressure at nominal voltage | 1.0 atmosphere (0 psig) |
| Maximum outlet pressure | 115V range: 5 psig, 115 VAC 50/60HZ or greater, derate linearly to 0 psig at 100VAC below 115VAC 230V range: 5 pisg, 200 VAC 50/60HZ or greater |
| Inlet connection | NW25 |
| Exhaust connection | Female 1/4" National Pipe Thread (NW16) |
| Gas ballast | Female 1/8" National Pipe Thread, (20 micron sintered plug provided) |
| Ambient operating temperature | 5 °C to 40 °C (41 °F to 104 °F) |
| Storage temperature | –20 °C to 60 °C (–4 °F to 140 °F) |
| Motor rating | 1/2 HP (370 W) |
| Operating voltages** | Operates with inverter - functions with several power inputs: 1 phase/100-120VAC 50/60Hz, 9 A typical; 200-240 50/60Hz, 5 A typical |
| Motor thermal protection | Automatic thermal protection |
| Operating speed | Full speed is 1900 RPM; RPM can be adjusted between 1350 to 1900 RPM. Factory setting for X3810-64000 and X3810-64010 is 1900 rpm, there is optional default setting at 1800 rpm available. |
| Cooling system | Air-cooled |
| Weight | Pump only: 24.7 kg (55 lbs) Shipping weight: 28.1 kg (62 lbs) |
*Ultimate pressure; **Operating voltages
| Country | Power Cord Specification | Order |
|---|---|---|
| Europe | 10 A / 220-230 VAC, 2.5 m | 656494220 |
| Denmark | 10 A / 220-230 VAC, 2.5 m | 656494225 |
| Switzerland | 10 A / 230 VAC, 2.5 m | 656494235 |
| UK/Ireland | 13 A / 230 VAC, 2.5 m | 656494250 |
| India | 10 A / 220-250 VAC, 2.5 m | 656494245 |
| Israel | 10 A / 230 VAC, 2.5 m | 656494230 |
| Japan | 12 A / 100 VAC, 2.3 m | 656494240 |
| North America | 15 A / 125 VAC, 2.0 m | 656458203 |
| 10 A / 230 VAC, 2.5 m | 656494255 |
| Factory Service Plans | Model Number |
|---|---|
| IDP-10 Exchange Pump | X3810-69000 |
| IDP-10 Exchange with Inlet Valve | X3810-69010 |
| IDP-10 Full Pump Rebuild | X3810-69000-RP |
| IDP-10 Full Pump (with Inlet Valve) Rebuild | X3810-69010-RP |
| Pin # | Function |
|---|---|
| 1 | Analog speed control input voltage 0-10 VDC |
| 2 | External pump On/Off input, active LO |
| 3 | NC |
| 4 | NC |
| 5 | Ground |
| 6 | NC |
| 7 | +10VDC output |
| 8 | Reserved for Factory use |
| 9 | Ground |
| Раздел | Страница |
|---|---|
| Description of the IDP-10 Dry Scroll Vacuum Pump | 184 |
| Technical Specifications | 185 |
| IDP-10 Dry Scroll Vacuum Pump Outline Drawing | 187 |
| Electrical Connections | 189 |
| Power Cord | 189 |
| Grounding Instructions | 189 |
| Extension Cords | 190 |
| Inrush Current | 191 |
| Mechanical Connections | 191 |
| Isolation Valve | 191 |
| Pump Inlet | 191 |
| Pump Exhaust | 191 |
| Gas Ballast | 192 |
| Pump Operations | 193 |
| Mounting of the IDP-10 Scroll Vacuum Pump | 193 |
| Protection Features | 194 |
| Troubleshooting | 195 |
| Kits and Service Options | 196 |
| Tip Seal Replacement | 197 |
| Characteristic | Value |
|---|---|
| Noise Level (per ISO 11201) | 53 +/- 2 dB(A) |
| Vibration level at inlet (per ISO 10816-1) | Class 1B, 7 mm/s |
| Fuse type | M5 x 20 mm, 10 A, Slo-Blo, Littelfuse Model 215010 or equivalent, not user replaceable |
| Hour meter | Integral hour meter provided, displays running time to closest 1 hour |
| Conformance standards | Conforms to EN61010 3rd edition, CSA 22.2 for US and Canada, CE marked. |
| Installation | Indoor use, Installation Category II, Pollution Degree 2 |
| Altitude | 2000 m (6562’) |
| Emissions and Immunity | EN 61236-1:2006 as a Class A device |
| Water Vapor Handling | 120 g/hr |
| Ambient Relative Humidity | 5-95%; non-condensing |
| Safety | EN /IEC 61010 3rd Edition CSA certified for US and Canada |
*If the pump and/or vacuum system is new, or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the system it is recommended to open GB1 for 15 minutes and/or flush with dry nitrogen. Ultimate pressure will varies for different configuration and RPM setting. **Mains supply voltage fluctuations shall not exceed 10% of the nominal voltage.
The pump must be kept within -20 to +60 °C and 0–95% non-condensing RH during transport and storage.
Sets the environmental limits for shipping and warehousing before installation.
“Temperatur: -20 °C bis +60 °C (-4 °F bis 140 °F)”
The IDP-10 alone weighs 24.7 kg, so lifting aids and proper handling techniques are required to avoid injury.
Determines the lifting equipment and manpower needed for safe installation.
“La pompe IDP-10 pèse à elle seule 24,7 kg (55 lb).”
A clear space of at least 60 cm must be maintained around the equipment for ventilation, service access, and main breaker access.
Defines the minimum bench/floor clearance required when laying out the installation.
“Il doit y avoir un espace bien dégagé d’au moins 60 cm (24") sur l’intégralité du pourtour de l’équipement.”
The unit is Class A Group 1 ISM equipment, so EMC compatibility may be difficult to guarantee outside industrial environments.
Restricts where the pump may be installed and warns about interference in residential or sensitive environments.
“Ce dispositif est conforme aux exigences de CISPR11, Groupe 1, Classe A en tant qu’équipement professionnel de radiation.”
Pin 2 is a TTL-compatible logic input where a closed switch to signal ground or a logic LO runs the pump, and the slide switch must be in position 0 when using it.
Specifies the remote start/stop wiring and the mandatory switch setting to avoid misoperation.
“Notez que l'interrupteur coulissant DOIT être dans la position 0 lorsque vous utilisez cette sortie.”
Pin 7 provides a 10 VDC reference for a speed-control potentiometer, and a 5 kOhm potentiometer is recommended.
Gives the correct potentiometer value for manual speed control wiring.
“Un potentiomètre de 5 Kohm est recommandé.”
Pin 8 must be left unconnected, and the shell (Pin 9) is chassis ground and is not connected to the signal ground pins.
Prevents wiring errors and clarifies the grounding scheme for the control connector.
“Broche n°8 : n'effectuez pas de raccordement à cette broche.”
Vacuum pumps and components may be contaminated by toxic, reactive, or radioactive substances, so adequate PPE must be worn during maintenance and reinstallation.
Mandates PPE selection based on process contamination risk during service work.
“Wear adequate protective equipment during maintenance and repairs or in case of reinstallation.”
Any equipment heavier than 20 kg must be moved with a suitable lifting device, and steel-toe safety shoes per EN 347 are required.
Sets the manual-handling threshold and PPE requirement for moving the pump.
“Any equipment heavier than 20 kg should be transported using a suitable lifting device.”
The pump must be fully stopped before servicing or before mounting/dismounting from the system, and hands must not be put inside the high-vacuum flange.
Defines the lockout/standstill prerequisite for safe maintenance and removal.
“Before servicing the pump or before any mounting/ dismounting action of the pump from the system, wait for the complete standstill of the pump.”
Higher-level maintenance must be performed only by specially trained personnel, not by the ordinary user.
Clarifies the boundary between user-level and trained-service maintenance tasks.
“must not be carried out by the user.”