Technical Reference Summary

Covered Models & Document Scope

Covers Labconco Purifier® HEPA Filtered Enclosures and Purifier® Class I Filtered Enclosures in 2', 3' and 4' widths. Listed part numbers: 3980200, 3980201, 3980202, 3980203, 3980220, 3980221, 3980222, 3980223, 3980300, 3980301, 3980302, 3980303, 3980320, 3980321, 3980322, 3980323, 3980400, 3980401, 3980402, 3980403, 3980420, 3980421, 3980422, 3980423. Options/accessories covered: HEPA filters, Odor Control carbon filters (organic vapor, formaldehyde, ammonia/amines, acid sulfur), thimble exhaust connections, Guardian™ Airflow Monitor, UV light (Class I only), work surfaces, utility shelves, remote blowers and adjustable dampers.

Section-by-Section Technical Breakdown

pp. 6-7
Exhaust and Blower Requirements

The enclosures use an integral motorized impeller to draw room air past the operator and push it through the HEPA filter; filtered exhaust is discharged from the top and may be recirculated or ducted outside via a thimble exhaust connection and remote blower. Exhaust volume, noise and static pressure data are tabulated for face velocities of 60, 75, 80, 90, 100 and 105 fpm for each width. If connected to a house exhaust system, an adjustable damper or valve must be installed to control airflow.

  • 2' Class I: 60 fpm = 85 CFM, 48-53 dB(A), max 400 ft equivalent 6" duct, 0.20" external static pressure
  • 2' HEPA Filtered: 75 fpm = 110 CFM, 49-55 dB(A), 220 ft, 0.19"; 105 fpm = 155 CFM, 54-59 dB(A), 70 ft, 0.12"
  • 3' Class I: 60 fpm = 130 CFM, 48-53 dB(A), 130 ft, 0.15"; 3' HEPA Filtered: 75 fpm = 165 CFM, 52-56 dB(A), 65 ft, 0.12"; 105 fpm = 230 CFM, 60-64 dB(A), 20 ft, 0.07"
  • 4' Class I: 60 fpm = 175 CFM, 49-54 dB(A), 100 ft, 0.20"; 4' HEPA Filtered: 75 fpm = 220 CFM, 54-58 dB(A), 60 ft, 0.19"; 105 fpm = 310 CFM, 64-67 dB(A), 20 ft, 0.12"
  • For outside exhaust the enclosure must be connected to a dedicated blower or house exhaust system; an adjustable damper/valve is mandatory when connected directly to house exhaust, especially when one house system serves multiple enclosures
pp. 9-11, 17
Installing the Filtered Enclosure on a Supporting Structure and Work Surface

The enclosure must be placed on a level, structurally adequate base cabinet and a smooth, rigid, chemical-resistant work surface at least as wide as the enclosure, with a dished recess for primary spills. The front of the enclosure must align within 0.36" of the front of the work surface. Four mounting screws secure the enclosure; a minimum 1/8" clearance under the screw heads allows the rear and front air-foil flanges to slide into place.

  • Work surface must be smooth, rigid, non-porous and resistant to biohazards, powders, solvents and chemicals; epoxy resin is given as an example
  • Front of enclosure aligned within 0.36" of front of work surface
  • Four mounting screws with minimum 1/8" clearance under the head
  • A stand or mobile bench is not recommended for precise powder weighing
  • Seal enclosure to work surface with a bead of silicone sealant after ducting and wiring
pp. 12-13
Connecting to the Exhaust System (Optional)

Thimble exhaust connections are available for ducting the enclosure to a house exhaust system or to a FilterMate Portable Exhauster. The weight of the exhaust ductwork must be supported independently of the enclosure superstructure, and the duct material must match the enclosure, procedures and chemical applications. Thimble exhaust part numbers are 3924400 (2'), 3924401 (3') and 3924402 (4').

  • 6" connection for house exhaust; 5" connection for FilterMate Portable Exhauster
  • Ductwork weight must be supported independently of the enclosure superstructure
  • Exhaust system should be installed by a qualified HVAC contractor
  • Thimble exhaust part numbers: 3924400 (2'), 3924401 (3'), 3924402 (4')
pp. 14-15
Installation of HEPA Filters and Accessory Odor Control Carbon Filters

HEPA filters ship installed with the gasket on the downstream side and are leak-checked at Labconco; a second leak check is required before use and at least annually thereafter. Carbon filters are gravity-installed with the gasket side down on the exhaust side of the HEPA filter after removing the upper perforated cover. Filter part numbers are listed by type and enclosure width.

  • HEPA filters: 2' 3707900, 3' 3707901, 4' 3707902; 99.99% removal efficiency for 0.3-micron particles
  • Organic Vapor carbon: 2' 3937200, 3' 3937300, 4' 3937400
  • Formaldehyde carbon: 2' 3937201, 3' 3937301, 4' 3937401
  • Ammonia and Amines carbon: 2' 3937202, 3' 3937302, 4' 3937402
  • Acid Sulfur carbon: 2' 3937203, 3' 3937303, 4' 3937403
  • Carbon filters provide odor control only, no particulate protection
  • All seams upstream of the HEPA filter are jacketed by negative pressure
pp. 15-16
Electrical Supply Connection and Face Velocity Setting

115V models connect via the supplied power cord to the IEC inlet on the rear; at 50 Hz maximum airflow is reduced by 17%. 230V models ship without a plug and require a locally appropriate plug. Face velocity is set with the speed control behind the front panel: 75-105 fpm for biohazardous operations, with containment maximized within this range and quietest operation at the lowest appropriate velocity.

  • 115V models: supplied power cord to IEC inlet; 50 Hz operation reduces maximum airflow by 17%
  • 230V models: shipped without plug; install plug per local codes
  • Face velocity 75-105 fpm for biohazardous operations
  • Speed control located behind front panel; counter-clockwise increases face velocity, clockwise decreases it
  • Do not use a detachable power cord not adequately rated for the unit
pp. 16
Validating the Vented Enclosure

Face velocity is measured across the sash opening in two rows: 3" down from the upper sash foil and 3" up from the work surface, with readings 6" from the ends and every 12". Four readings are taken for the 2' enclosure, six for the 3' and eight for the 4', then averaged. Enclosures were tested per ASHRAE 110-1995 with an as-manufactured rating of less than 0.05 ppm at 4 lpm, and per NSF/ANSI 49 biological containment testing at 75 fpm inflow.

  • Two measurement rows: 3" below upper sash foil and 3" above work surface
  • Readings 6" from ends and evenly spaced every 12"
  • 2' enclosure: 4 readings; 3' enclosure: 6 readings; 4' enclosure: 8 readings
  • ASHRAE 110-1995 as-manufactured rating: AM <0.05 ppm at 4 lpm
  • Biological containment tested at 75 fpm inflow with ~5 x 10^8 spores of Bacillus subtilis var Niger; typically 1 or fewer spores escaped
pp. 18-24
High Performance Features

The enclosure uses aerodynamic air foils, an upper sash foil, an upper dilution air supply, a zoned rear perforated baffle and side-entry air foils to produce horizontal airflow and capture contaminants on the first pass. The plenum and HEPA filter are jacketed by negative pressure so leaks are recaptured and refiltered. Key construction details include a 10-degree ergonomic slope, 23" internal depth, 1/4" tempered safety glass sash, and a vibration-isolated motorized impeller.

  • Upper dilution air supply introduces 5-7% of required air volume
  • Internal depth 23"; ergonomic slope 10 degrees
  • Sash constructed of 1/4 inch tempered safety glass; must be down for normal operation
  • Vibration-isolated, dynamically balanced motorized impeller for precision weighing
  • Filter system pressure gauge indicates total system pressure across HEPA filter and plenum
  • Guardian™ Airflow Monitor available as an option on all models
pp. 33-49
Maintaining Your Filtered Enclosure

Chapter 6 covers the routine maintenance schedule, decontamination, determination of when to replace HEPA and carbon filters, HEPA filter leak testing, setting inflow face velocity with the speed control, calibrating and operating the airflow monitor, calculating carbon filter life, initial certification and re-certification, and replacement of fluorescent/UV lamps, motorized impeller and speed control.

  • HEPA filter leak test required before use and at least annually
  • Speed control adjustment used to set inflow face velocity
  • Airflow monitor calibration and operation procedures included
  • Carbon filter life calculation method provided
  • Initial certification and re-certification procedures included
  • Replacement procedures for fluorescent lamp, UV lamp, motorized impeller and speed control
pp. 56-59
Troubleshooting

Chapter 8 provides a table of problems, probable causes and corrective actions for containment, blower, lighting, UV and airflow monitor issues. It also includes a service operating record log for documenting maintenance.

  • Problems covered: contaminants outside enclosure, blower won't operate, low face velocity or poor containment, blower and lights won't operate, light does not work, airflow monitor malfunction
  • Corrective actions reference face velocity ranges of 60-100 fpm for weighing and 75-105 fpm for biohazardous procedures
  • Service operating record log for date, comments, customer name, model number, date installed, serial number and unit location

Key Integration, Operating & Safety Requirements

Specification / Condition Engineering Implication Applies To
HEPA filters are rated at least 99.99% efficient for 0.3-micron particles. p.5 ↗ Suitable for containment of toxic, noxious and biohazardous particulates; verify filter part number by enclosure width. All Purifier HEPA Filtered and Class I Enclosures
Face velocity for biohazardous operations should be 75-105 fpm; for weighing procedures 60-100 fpm. p.15 ↗ Set speed control within these ranges; lowest appropriate velocity gives quietest operation. All models
At 50 Hz operation, blower maximum airflow is reduced by 17%. p.15 ↗ Account for reduced airflow when sizing exhaust and setting face velocity on 50 Hz supplies. 115V models
2' HEPA Filtered Enclosure at 75 fpm: 110 CFM, 49-55 dB(A), max 220 ft equivalent 6" duct, 0.19" external static pressure. p.7 ↗ Use these values for blower selection and duct sizing for 2' models. 2' Purifier HEPA Filtered Enclosure
3' HEPA Filtered Enclosure at 75 fpm: 165 CFM, 52-56 dB(A), max 65 ft equivalent 6" duct, 0.12" external static pressure. p.7 ↗ Use these values for blower selection and duct sizing for 3' models. 3' Purifier HEPA Filtered Enclosure
4' HEPA Filtered Enclosure at 75 fpm: 220 CFM, 54-58 dB(A), max 60 ft equivalent 6" duct, 0.19" external static pressure. p.7 ↗ Use these values for blower selection and duct sizing for 4' models. 4' Purifier HEPA Filtered Enclosure
An adjustable damper or valve must be installed when the enclosure is connected directly to a house exhaust system. p.7 ↗ Required to control airflow properly, especially when one house system serves multiple enclosures. All models with house exhaust connection
Enclosure weight is between 125 and 195 lbs (55 to 85 kg) depending on model. p.8 ↗ Use mechanical lift truck or floor jack; do not lift by the front air foil; do not tilt onto a hand truck. All models
Front of enclosure must be aligned within 0.36" of the front of the work surface. p.9 ↗ Ensures proper support and sealing; work surface must be at least as wide as the enclosure. All models
Mounting screws require a minimum of 1/8" clearance under the head for positioning the enclosure. p.10 ↗ Allows rear flange and front air foil flanges to slide under screw heads during installation. All models
Thimble exhaust connections: 6" for house exhaust and 5" for FilterMate Portable Exhauster. p.12 ↗ Select duct and blower accordingly; ductwork weight must be supported independently of the enclosure. All models with thimble exhaust
HEPA filter leak check must be performed before use and at least annually thereafter. p.14 ↗ Schedule leak testing as part of commissioning and routine maintenance. All models
Carbon filters are gravity-installed with the gasket side down on the exhaust side of the HEPA filter. p.14 ↗ Correct orientation is essential for odor control; carbon filters provide no particulate protection. All models with accessory carbon filters
Upper dilution air supply introduces 5-7% of the required air volume. p.22 ↗ Part of the containment design; do not block the upper dilution air supply. All models
Internal depth is 23 inches and the ergonomic slope is 10 degrees. p.22 ↗ Accommodates modern balances and auxiliary equipment without extending outside the enclosure. All models
Sash is constructed of 1/4 inch tempered safety glass and must be down for normal operation. p.22 ↗ Operating with sash open compromises containment; use spring-loaded latch only for loading and cleaning. All models
ASHRAE 110-1995 as-manufactured rating is less than 0.05 ppm at 4 lpm (AM <0.05). p.16 ↗ Provides a benchmark for field certification and containment performance. All models
Face velocity measurement: two rows, 3" below upper sash foil and 3" above work surface; readings 6" from ends and every 12"; 4 readings for 2', 6 for 3', 8 for 4'. p.16 ↗ Standardized method for validating average face velocity. All models
UV lamp is found only on the Purifier Class I Enclosure and will not penetrate the plane of the sash and lower airfoil. p.23 ↗ UV is for surface decontamination when the enclosure is not in use; UV closure panel must be in place for the lamp to work. Purifier Class I Enclosure
Carbon filter concentrations in the work area must not exceed the IDLH for the chemical and exhaust must not exceed the TWA. p.14 ↗ Carbon filters are for trace nuisance odors, not continuous fume generation; verify chemical suitability with a Labconco product specialist. All models with accessory carbon filters

Faults & Troubleshooting

Code / Symptom Possible Cause Remedy Note
Contaminants outside of enclosure. p.60 ↗ Improper user techniques for the enclosure. See “Certifying the Enclosure” Chapter 3 and “Safety Precautions” Chapter 4 sections in the manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.60 ↗ Restriction of the baffle air slots or blockage of the exhaust outlet. Remove obstruction to ensure that all air slots and the exhaust outlet are unobstructed. Chapter 8
Contaminants outside of enclosure. p.60 ↗ External factors are disrupting the enclosure airflow patterns or acting as a source of contamination. See “Location Requirements” Chapter 2, “Certifying the Enclosure” Chapter 3, and “Safety Precautions” Chapter 4 sections of this manual. (Ref. Appendix D). Chapter 8
Contaminants outside of enclosure. p.60 ↗ Enclosure has improper face velocity. Have enclosure certified and check exhaust system. Check HEPA filters for loading. Adjust enclosure speed control. Enclosure should have an average face velocity of 60-100 fpm for weighing procedures and 75-105 fpm for biohazardous procedures. Chapter 8
Blower won’t operate. p.61 ↗ Unit not plugged into outlet. Plug the enclosure into appropriate electrical service. Chapter 8
Blower won’t operate. p.61 ↗ Circuit breaker(s) or Ground Fault Interrupter. Reset circuit breaker. Chapter 8
Blower won’t operate. p.61 ↗ Blower wiring is disconnected. Inspect blower wiring. Chapter 8
Blower won’t operate. p.61 ↗ Blower switch is defective. Replace switch. Chapter 8
Blower won’t operate. p.61 ↗ Motorized impeller or blower is defective. Replace motorized impeller or blower. See Chapter 6. Chapter 8
Low face velocity or poor containment of contaminants. p.61 ↗ Enclosure sash not closed. Close sash to the lowest position. Chapter 8
Low face velocity or poor containment of contaminants. p.61 ↗ HEPA filter clogged. Replace HEPA filter or increase speed. Chapter 8
Low face velocity or poor containment of contaminants. p.61 ↗ Carbon filter loaded with chemicals. Replace carbon filter. Chapter 8
Blower and lights won’t operate. p.61 ↗ Unit not plugged into outlet. Plug enclosure into appropriate electrical service. Chapter 8
Blower and lights won’t operate. p.61 ↗ Circuit breaker(s) tripped. Reset or replace circuit breaker. Chapter 8
Light does not work. p.61 ↗ Lamp not installed properly. Inspect lamp installation. Chapter 8
Light does not work. p.61 ↗ Lamp wiring disconnected. Inspect lamp wiring. Chapter 8
Light does not work. p.61 ↗ Defective lamp. Replace lamp. Chapter 8
Light does not work. p.61 ↗ Light switch is defective. Replace light switch. Chapter 8
Light does not work. p.61 ↗ Defective electronic ballast. Replace ballast. Chapter 8
Light does not work. p.61 ↗ UV closure panel not installed. Install UV closure panel (UV lights only). Chapter 8
Light does not work. p.61 ↗ Magnetic switch for UV closure panel defective. Replace switch (UV lights only). Chapter 8
Light does not work. p.61 ↗ UV timer circuit board defective. Replace circuit board (UV lights only). Chapter 8
Airflow Monitor Malfunction p.62 ↗ No power. No lights. No display. Power cable to circuit board is disconnected and needs to be connected. Sensor cable needs to be connected. Verify that all airflow monitor interface cables are connected. Check fuses on enclosure. Chapter 8
Airflow Monitor Malfunction p.62 ↗ No audible alarm. Alarm has been temporarily silenced using “SILENCE ALARM”. Wrong alarm set point. Airflow monitor was not properly adjusted. Repeat calibration steps outlined in this manual in Chapter 6. Chapter 8
Airflow Monitor Malfunction p.62 ↗ Constant audible alarm. Check airflow and calibration of airflow monitor. See Chapter 6. Chapter 8
Airflow Monitor Malfunction p.62 ↗ Continuous alarm. Check the face velocity of the enclosure as the airflow of the system may have changed. If incorrect, adjust the speed control to increase face velocity. The HEPA filter may have become loaded. If face velocity is correct, calibrate the airflow monitor outlined in this manual in Chapter 6. Chapter 8
Airflow Monitor Malfunction p.62 ↗ Monitor alarms; air way to airflow monitor sensor is blocked by insects, dust or debris. Lightly clean the airway with clean air. Be careful not to touch sensitive electrical components. Chapter 8
Airflow Monitor Malfunction p.62 ↗ Audible disable will not stay operational. An alarm condition must be continuously present before the audible alarm can be silenced. If flow conditions fluctuate near the alarm set point, the airflow monitor will automatically reset it. Action should be taken to bring the enclosure airflow into proper operating parameters or adjust the alarm set point lower. Chapter 8