Crude Fiber Digestion Apparatus Manual
Technical Reference Summary
Covered Models & Document Scope
Covers Labconco Crude Fiber Digestion Apparatus Models 3000100 (115 V, 50/60 Hz, 18.3 A, 6-unit) and 3000200 (230 V, 50/60 Hz, 9.11 A, 6-unit). Both are supplied without a power cord; customer must wire the main power switch per local electrical codes. Each unit ships with 12 × 600 ml beakers. Manual part number 3002300, Rev. E.
Section-by-Section Technical Breakdown
Lists the contents of the shipping carton for the Crude Fiber Digestion Apparatus. The carton should contain the apparatus (115 V or 230 V variant), the user's manual, and 12 crude fiber beakers of 600 ml. Packaging should be retained until all components are checked and the unit is installed and tested.
- Part 3000100: Crude Fiber Apparatus 115 V, 50/60 Hz
- Part 3000200: Crude Fiber Apparatus 230 V, 50/60 Hz
- Part 3002300: User's Manual
- Part 1935000: Crude Fiber Beakers – 600 ml (12 pieces)
Describes the main functional components of the apparatus: the manifold assembly with a 6-inch diameter exhaust connection, the condenser head system that condenses and returns vapors to the beakers, the 350 W spring-loaded ceramic-embedded heaters, the variable heat control switches providing 20–100% input, the elevation control knob for beaker positioning, and the main power switch.
- Manifold assembly: molded 6" diameter exhaust connection for low static pressure loss and good transport velocity
- Condenser head system: cone-shaped heads above each heater condense vapors and return them to the digestion medium
- Heater system: each 350 W adjustable spring-loaded heater with coiled heater wires embedded in ceramic material
- Variable heat control switch: provides 20–100% of capacity to each heater
- Elevation control knob: adjusts beaker positions
- Main power switch: turns apparatus on/off without changing variable heat control settings
Provides electrical and plumbing requirements for both models. Model 3000100 requires 115 V, 50/60 Hz, 18.3 A; Model 3000200 requires 230 V, 50/60 Hz, 9.11 A. Neither model includes a power cord; the customer must wire the main power switch according to local electrical codes. Water connections are 3/8" NPT inlet (upper left) and 3/8" NPT outlet (upper right) to an open drain of at least 1/2" NPT. Condensers are Type 304 stainless steel.
- Model 3000100: 6-unit, 115 V, 50/60 Hz, 18.3 A, no power cord supplied
- Model 3000200: 6-unit, 230 V, 50/60 Hz, 9.11 A, no power cord supplied
- Each heater has an individual infinite control switch with pilot light; main ON/OFF switch provided
- Condensers: Type 304 stainless steel
- Water inlet: 3/8" NPT valve on upper left side; water outlet: 3/8" NPT on upper right side
- Drain facilities: at least 1/2" NPT, open drain for visual flow check
- Location must have convenient electrical outlet of required voltage, cold water supply of 3/8" NPT, and drain of at least 1/2" NPT
Instructs that all heater control switches should be turned on to ensure proper performance during start-up and check-out.
- All heater control switches should be turned on to insure proper performance
Outlines the routine digestion procedure: turn on cold water and verify flow at drain, prepare samples per AOAC methods, place sample and digestion medium in beaker, raise beaker to condenser head using the elevation control knob (turn 1/4 turn left while holding knob, do not release until beaker is against condenser head), turn main power switch ON, and follow AOAC method for fiber determination. Heater and water flow rates may be adjusted to vary reflux rates.
- Turn on cold water supply and check visual flow at drain area
- Heater and water flow rates may be adjusted to vary reflux rates according to operator preference
- Prepare samples per standard method for product under examination
- Place required quantity of sample and proper digestion medium into beaker
- Place beaker on heaters; while firmly holding knob, turn 1/4 turn to the left and slowly let beaker rise to condenser head; DO NOT let go of knob until beaker is in position against condenser head
- Turn main power switch (right side) to ON; controls independently regulate one 350 W heater; pilot lights indicate heater ON cycle
- Follow AOAC method and procedure for determination of fiber content
States that if problems develop with the unit, the user should contact the laboratory dealer from whom the unit was purchased. If satisfaction is not obtained, contact Labconco Corporation at 8811 Prospect, Kansas City, Missouri 64132, phone 1-800-821-5525.
- If problems develop, contact the laboratory dealer from whom the unit was purchased
- If satisfaction is not obtained, contact Labconco Corporation, 8811 Prospect, Kansas City, Missouri 64132, Phone 1-800-821-5525
Lists replacement parts and accessory part numbers for the Crude Fiber Digestion Apparatus. Replacement parts include infinite controls, toggle switch, needle valve, knob, heating elements, beakers, heater harness wires, pilot light wire, heater lead wire, condenser assembly, and condenser gasket. Accessories include California Modified Büchner Funnel and Oklahoma State Filter.
- 1315400: Infinite Control, 120 Volt, 15 Amp
- 1315500: Infinite Control, 230 Volt, 15 Amp
- 1315900: Toggle Switch, 120 Volt/230 Volt, 15 Amp
- 1351600: Valve, needle, water
- 1870200: Knob for infinite control
- 3003100: Heating element, 115 Volt, 350 Watt
- 3003200: Heating element, 230 Volt, 350 Watt
- 3004300: Beaker, 600 ml, Package of 6
- 3013500: No. 14AF – White heater harness wire, stranded, 15' length
- 3013600: No. 14AF – Black heater harness wire, stranded, 15' length
- 3013700: No. 12AF – White heater harness wire, stranded, 15' length
- 3013800: No. 18AF – Pilot light wire, stranded, 15' length
- 3013900: No. 18 Heater lead wire, stranded nickel, glass braid, 15' length
- 3020800: Condenser Assembly
- 3022500: Condenser Gasket
- 5510000: California Modified Büchner Funnel, complete; standard polyethylene two-piece funnel with 200 mesh stainless steel screen
- 5520100: Replacement screw only, for #5521000
- 5520200: Back-up plate only for #5521000
- 5521000: Oklahoma State Filter, complete; stainless steel, one-piece with threaded cap, 200 mesh screw with perforated back-up plate
Appendix A contains dimensional information for the apparatus, presented in Figure 2. The extracted text does not include numeric dimensions.
- Dimensional drawing provided in Figure 2 (Appendix A)
Appendix B contains the wiring diagram for the apparatus, presented in Figure 3. The extracted text does not include the diagram details.
- Wiring diagram provided in Figure 3 (Appendix B)
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Model 3000100 requires 115 V, 50/60 Hz, 18.3 A; Model 3000200 requires 230 V, 50/60 Hz, 9.11 A. p.7 ↗ | Electrical supply and branch circuit protection must be sized accordingly; no power cord is supplied, so field wiring to the main power switch is required per local codes. | 3000100 and 3000200 |
| Each heater is rated 350 W and is adjustable from 20% to 100% of capacity via individual infinite control switches. p.5 ↗ | Heater output can be tuned for different reflux rates; each of the 6 positions is independently controlled. | 3000100 and 3000200 |
| Water inlet is a 3/8" NPT valve located on the upper left side; water outlet is 3/8" NPT on the upper right side. p.7 ↗ | Plumbing must provide a 3/8" NPT cold water supply and route the outlet to drain. | 3000100 and 3000200 |
| Drain facilities must be at least 1/2" NPT and should be an open drain to allow visual check of water flow. p.7 ↗ | An open drain is required for operator verification of condenser water flow. | 3000100 and 3000200 |
| Condensers are Type 304 stainless steel. p.7 ↗ | Material compatibility is limited to substances compatible with Type 304 stainless steel. | 3000100 and 3000200 |
| The manifold exhaust connection is 6" diameter, sized for minimum static pressure loss while maintaining good transport velocity. p.5 ↗ | Exhaust ducting should be sized to match the 6" connection to preserve hood performance. | 3000100 and 3000200 |
| The unit is supplied with 12 × 600 ml beakers. p.5 ↗ | Spare beakers can be ordered as Part 3004300 (package of 6). | 3000100 and 3000200 |
| During operation, the elevation control knob must be turned 1/4 turn to the left while firmly held; do not release until the beaker is in position against the condenser head. p.9 ↗ | Prevents beaker misalignment or spillage during raising to the condenser head. | 3000100 and 3000200 |
| All heater control switches should be turned on to insure proper performance during start-up and check-out. p.8 ↗ | Verifies all heater circuits and pilot lights function before routine use. | 3000100 and 3000200 |
| Replacement heating elements: 3003100 (115 V, 350 W) and 3003200 (230 V, 350 W). p.11 ↗ | Select the correct element voltage matching the unit model. | 3000100 and 3000200 |
| Replacement infinite controls: 1315400 (120 V, 15 A) and 1315500 (230 V, 15 A); toggle switch 1315900 (120/230 V, 15 A). p.11 ↗ | Control components are voltage-specific; order the correct variant for the unit. | 3000100 and 3000200 |
| Water flow is controlled by a needle valve (Part 1351600). p.11 ↗ | Replacement valve is available for flow regulation maintenance. | 3000100 and 3000200 |
| Condenser assembly Part 3020800 and condenser gasket Part 3022500 are available as replacement parts. p.11 ↗ | Gaskets are maintenance items and may need periodic replacement. | 3000100 and 3000200 |
| Accessory California Modified Büchner Funnel (5510000) uses a 200 mesh stainless steel screen; Oklahoma State Filter (5521000) is stainless steel one-piece with threaded cap and 200 mesh screw with perforated back-up plate. p.12 ↗ | Filtration accessories are available for post-digestion separation. | 3000100 and 3000200 |