Goldfisch Fat Extraction Apparatus Manual (1984-Present)
Technical Reference Summary
Covered Models & Document Scope
Covers the Labconco Goldfisch Fat Extraction Apparatus, 6-unit model 35001 (catalog model 3500100). Document part # 3518100, Rev. L. Accessories noted: stainless steel sample tube container 35067 and tall-form beaker 35046 are not furnished as standard equipment.
Section-by-Section Technical Breakdown
Lists the carton contents for the 6-unit Goldfisch apparatus, including the apparatus itself, glassware sets of six, gaskets, wave washers, retainer rings, heater covers and the instruction manual. Users are instructed to check for transit damage before discarding packing material.
- Apparatus part number 35001.
- Beakers 100 cc (12 pcs, 35051).
- Sample tubes (6, 35049), reclaiming tubes (6, 35048), extraction thimbles (6, 35050).
- Condenser gaskets (6, 35052), wave washer–beaker seals (6, 35191), beaker retainer rings (6, 35054).
- Heater cover assemblies (6, 35185) and instruction manual 35181.
Describes the continuous extraction principle for single or multiple fat extractions per AOAC methods, noting reduced reflux time versus Soxhlet. Identifies the manifold, pressure release valves, condenser head system and spring-loaded heater system.
- Manifold distributes cooling water to condenser heads; water exits through the control valve on the upper left.
- Pressure release valves at the top of the manifold above each condenser vent excess gaseous pressure and reset automatically.
- Condenser head system condenses solvent vapor on a water-cooled metal cone surface; condensate drips through the sample in the ceramic thimble.
- Each heater is 100 W, adjustable and spring loaded, with coiled heater wires embedded in ceramic; variable heat control switch at the front.
Covers site preparation, location requirements, utility connections and initial start-up. The unit is shipped fully assembled; operating parts are installed after placement. Requires a 115 VAC outlet, 3/8" NPT cold water supply and 3/4" NPT drain, and operation inside an explosion-proof hood.
- Location must have a convenient 115 VAC outlet, unrestricted 3/8" NPT cold water supply and drain facilities of at least 3/4" NPT.
- Operate inside an explosion-proof hood with proper airflow for flammable solvents.
- Connect cold water to the 3/8" NPT control valve fitting on the upper left.
- Connect the 3/8" NPT control valve on the upper right to an open drain of 3/4" IPS or larger.
- Cooling water minimum 0.75 gal/min (2.8 L/min); 55 °F (13 °C) or less preferred.
- Initial check: turn main power and variable heat control switches on, verify all heating elements heat properly, wait two to three minutes.
Describes assembling the gasket, wave washer, retaining ring and glassware per Appendix A, hand-tightening retaining rings to achieve a vapor seal, and verifying even water flow.
- Operating parts package is taped to the unit.
- Assemble gasket, wave washer, retaining ring and glassware per Appendix A assembly drawing.
- Hand tighten retaining rings to achieve vapor seal.
- Turn water on and confirm even flow leaving the unit.
Step-by-step routine extraction procedure: cooling water, sample preparation, thimble and sample tube loading, solvent charging, beaker assembly, heater elevation, heat rate selection, pre-heating, reflux regulation, solvent reclamation and extract drying.
- Extraction time reduced from 16 hours to about four hours by the refluxing system.
- 25 to 35 ml of solvent is normally sufficient for an estimated 3 to 4 hours of refluxing.
- Heat control 'HIGH' provides approximately 100 W to each heater; position '5' provides approximately 35 W each.
- Panel switches each control two heaters; if only one extraction is run, load the companion heater with a beaker of solvent or water.
- Pre-heat heaters about 8 to 10 minutes before starting samples to reduce waiting time for reflux.
- Excessive reduction of condenser water flow causes solvent loss; a condenser without flow permits solvent to escape through the pressure relief valves.
- After extraction, install heater covers, replace sample tubes with reclaiming tubes and remove covers to transfer solvent into the reclaiming tube, leaving oil in the beaker.
- Swing beaker holder behind each heater inclines the beaker in warm air for drying; finish in a drying oven and compare to tare weight.
States that all operating-parts-package items are routine maintenance items and recommends keeping spares. Gaskets and beakers wear out or break and are ordered by Labconco part number through a dealer or Labconco.
- All parts in the operating parts package are routine maintenance items.
- Keep additional supplies of gaskets and beakers in stock.
- Order by Labconco part number; when contacting Labconco directly, furnish the dealer name.
Appendix A lists replacement part numbers and quantities for electrical components, glassware, gaskets, seals and heater/condenser assemblies. Items marked with an asterisk are accessories not furnished as standard equipment.
- Pilot light 12726; needle water valve 13516; mercury toggle switch 13159 (115 V, 15 A); pressure relief valve 13564.
- Heating element 35022 (115 V, 110 W); infinite control switch 35163; wiring harness 35167 (6-unit).
- Beaker 35046 (tall form, accessory) and 35053 (100 cc); extraction thimble 35059 (22 x 80 mm).
- Sample tube container 35061 (glass) and 35067 (stainless steel, accessory); reclaiming tube 35062.
- Condenser gasket 35160 (upper) and 35094 (outer shell); wave washer 35191 (6 required); heater cover 35185.
- Spring 35120 (compression heater push rod); condenser assembly 35172.
- Specify the equipment serial number when ordering replacement parts.
Appendix B provides overall dimensions of the apparatus.
- Overall dimensions shown: 35.7 in, 26.9 in, 9 in.
Appendix C gives electrical ratings and standard features for the 6-unit model 35001.
- Model 35001 operates on 115 V, 50/60 Hz, maximum 5.2 A.
- Shipped with power cord and grounded 3-wire plug.
- Main power on-off; full-range variable heat control switch, each switch controlling two heaters.
- All stainless steel (Type 304) condenser assembly; pressure relief valve for each condenser.
Appendix D contains the wiring diagram for the apparatus.
- Wiring diagram provided in Appendix D.
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Applies To |
|---|---|---|
| Model 35001 operates on 115 V, 50/60 Hz at maximum 5.2 A and is supplied with a grounded 3-wire plug. p.15 ↗ | Provide a dedicated 115 VAC grounded outlet; verify circuit capacity for 5.2 A. | 6-unit model 35001 |
| Cold water supply connection is 3/8" NPT; drain must be 3/4" IPS or larger. p.7 ↗ | Size plumbing and drain accordingly before installation. | whole series |
| Cooling water flow minimum is 0.75 gal/min (2.8 L/min); 55 °F (13 °C) or less is preferable. p.7 ↗ | Insufficient flow or warm water degrades condensation and risks solvent loss. | whole series |
| The extractor must be operated inside an explosion-proof hood with proper airflow. p.7 ↗ | Site selection must include an explosion-proof hood due to extremely flammable solvents. | whole series |
| Each heater is 100 W; 'HIGH' provides approximately 100 W per heater and control position '5' provides approximately 35 W per heater. p.10 ↗ | Heat input can be selected over a wide range; each panel switch controls two heaters. | 6-unit model 35001 |
| A solvent charge of 25 to 35 ml is normally sufficient for an estimated 3 to 4 hours of refluxing. p.10 ↗ | Use this as the standard solvent volume for routine extractions. | whole series |
| Pre-heating the heaters for about 8 to 10 minutes before starting samples reduces waiting time for reflux. p.10 ↗ | Include a pre-heat step in the operating procedure. | whole series |
| If only one extraction is made, the companion heater on the same panel switch should be loaded with a beaker of solvent or water. p.10 ↗ | Loading the companion heater increases heater service life. | whole series |
| Reducing condenser water flow too much causes solvent loss in the beakers; a condenser without flow permits solvent to escape through the pressure relief valves. p.10 ↗ | Maintain adequate condenser water flow at all times during operation. | whole series |
| Retaining rings are hand tightened to achieve a vapor seal. p.8 ↗ | Do not over-tighten; hand tightening is sufficient for sealing. | whole series |
| Extraction time is reduced from 16 hours to about four hours by the refluxing system. p.9 ↗ | Plan batch durations around approximately four hours of reflux. | whole series |
| Heater covers must be used when removing or attaching beakers to condenser heads to prevent solvent contact with the hot heater. p.11 ↗ | Include heater cover use in the operating procedure for safety and to avoid ignition. | whole series |
| The apparatus is suitable for the indirect (Crude Fat) method: moisture determination followed by extraction of dried material, with loss reported as ether extract. p.11 ↗ | Method selection must follow official AOAC procedures. | whole series |
| Replacement heating element is 115 V, 110 W (part 35022); mercury toggle switch is rated 115 V, 15 A (part 13159). p.12 ↗ | Use exact replacement part numbers for electrical service. | 6-unit model 35001 |
| Extraction thimble replacement is 22 x 80 mm (part 35059); beaker replacement is 100 cc (part 35053). p.12 ↗ | Order consumables by these part numbers. | whole series |
| Overall dimensions are 35.7 in, 26.9 in and 9 in. p.14 ↗ | Verify bench and hood space against these dimensions. | 6-unit model 35001 |