Table 10 Unscheduled Maintenance and When to Perform
Unscheduled Maintenance
When to Perform
Debubble
• The CV of the data is larger than normal. • QC test failed. • No events showed up after sample collection started for certain period of time. • Instrument was not used for over one week.
Clean the External Surface of the Instrument
Keep the external surface of the instrument clean.
Decontamination
When suspecting contamination in the system.
Prime Fluidics System before Long-Term Storage
No plan to run the instrument for more than two weeks.
Prime Fluidics System after Long-Term Storage
Instrument has not been used for more than two weeks.
Purge Fluidics System before Shipment
The instrument needs to be shipped.
Calibrate the NovoCyte Fluidics Station II or NovoCyte Fluidics Cart
Abnormal warning about the liquid level in the reagent containers is observed.
Clean the Optical Filters
The optical filters get dirty.
Clean Fluidics System
• The CV of the data is larger than normal. • QC test failed. • Excessive debris in display.
No fluidic tubing disassembly or re-connection required
Fluid Level Sensing
Active sensing using weight sensors with automated warnings when any fluid level is out of specified range
Fluid Container Capacity
3 L sheath, 3 L waste, 500 mL cleaning, 500 mL rinse with Fluidics Station II. Optional large container for sheath (15 L) and waste (15 L) with Fluidics Cart.
Carryover
< 0.1%
Sample Injection Probe (SIP) Rinse
Automated flying collar wash of inner and outer SIP surface after each sampling
Fluidics System Monitoring
In-line pressure sensor monitors the pressure in real time. Automated system recovery when pressure is out of range due to clogging
Fluidics Maintenance
Automated startup and shutdown with fluidic system cleaning. Automated user executable cleaning, debubble, rinse, unclog, priming. Automated execution of user definable maintenance sequence.
Data Acquisition Software
NovoExpress
Parameters
Height and Area for FSC, SSC and all Fluorescent Channels, Width off FSC and Time
Dynamic Range
24 bits; 7.2 decades logarithmic scale
Fluorescence Photodetector Gain Control
0-1000 User adjustable, optimized default gain setting for each individual channel
Compensation
Full inter-beam matrix, during or post acquisition
File Export
FCS 3.0, FCS 3.1; CSV; Batch PDF report
Data Report
Automatic report. Customizable report. Batch PDF report
Usage Monitor
Comprehensive Transaction Log and System Log
User Management
Administrative creation of individual user accounts and user groups with privilege control. Configurable roles for individual users. User operation time tracking
1Specification and performance claims would be based on certain specific experiment conditions.
Table 6 NovoCyte Fluidics Station II LED Status Indicator Information
Name
Color
Meaning
NovoCyte Fluidics Station II LED Status Indicator
Green
Reagent volume is above the designated threshold and is sufficient to run experiments (NovoFlow > 0.9 L; NovoRinse > 150 mL; NovoClean > 150 mL), and the waste container is not full (< 2.8 L).
Engineering Facts
14 facts · every claim linked to a page
Installation
The instrument weighs over 23 kg, so plan for two-person lifting and proper technique during installation.
Determines manpower and rigging needed for safe placement during installation.
Source
“This product is large and the weight is more than 51 lb (23 kg).”
Installation
Moving the instrument after professional installation is discouraged and requires Agilent personnel to re-install.
Affects site planning and layout decisions to avoid costly relocation later.
Source
“moving the instrument is not recommended. If you need to move the instrument, be sure to contact Agilent Technologies professionals.”
Safety
The unit is a Class I laser product with lasers fully enclosed, so no special work area is needed except during service.
Confirms laser safety classification for lab siting and EHS approval.
Source
“Agilent flow cytometer is a Class I laser product according to IEC/EN 60825-1:2014.”
Safety
Only Agilent-supplied power cables may be used to maintain safety and EMC compliance.
Constrains power connection hardware during installation.
Source
“Never use cables other than the ones supplied by Agilent Technologies to ensure proper functionality and compliance with safety or EMC regulations.”
Safety
A reliable protective earth on a three-pin grounded outlet is mandatory to avoid shock hazard.
Requires verifying grounded outlets at the installation site.
Source
“the power cord must be inserted in the three-pin socket with a protective ground connection.”
Operation
Priming is mandatory after adding reagents, otherwise air in the tubing degrades instrument performance.
Defines a required operational step to avoid bad data after reagent refill.
Source
“It's necessary to run Priming after adding instrument reagent.”
Operation
Reagent containers must be color-coded and correctly matched to fluidics ports to prevent malfunction.
Guides correct plumbing hookup during setup and reagent changes.
Source
“Incorrect connection will cause the instrument malfunction. The tubing and containers are color coded to prevent such errors.”
Operation
Sample flow rate is selectable from 5 to 120 µL/min with three presets, and Slow is advised for high-resolution work.
Informs assay setup and throughput versus resolution trade-offs.
Source
“You can also manually set a desired flow rate within the range of 5 µL/min to 120 µL/min by adjusting the sliding bar.”
Performance
The ADC digitizes pulses at 80 million samples per second with 24-bit resolution.
Defines signal digitization performance relevant to data quality expectations.
Source
“A high-accuracy high-speed 24-bit analog-to-digital convertor is used to process the pulse in a speed of 80 million times per second.”
Maintenance
Fluidic system consumables must be replaced every two months or when software prompts.
Sets a recurring maintenance schedule and consumable budget.
Source
“Replacing Fluidic System Consumables Every two months or when prompted by NovoExpress Software.”
Maintenance
Preventive maintenance including SIP cleaning and flow cell unclogging is required monthly.
Establishes monthly service workload for reliable operation.
Source
“Every month.”
Operation
Software warns when sheath, rinse, or clean volumes drop below thresholds and when waste exceeds 2.8 L.
Defines fluid capacity limits that drive lab logistics and waste handling.
Source
“when the waste volume exceeds a set threshold (2.8 L), the software will give a warning to remind the user to empty the waste container.”
Compliance
The instrument is Class A Group 1 ISM equipment, which may cause radio interference in domestic environments.
Affects EMC planning and placement in non-residential lab environments.
Source
“This product complies with the requirement of CISPR 11, Group 1, Class A as radiation professional equipment.”
Safety
Data is stored on the workstation hard drive and is unrecoverable if deleted or corrupted, so regular backups are required.
Drives IT backup provisioning decisions at installation.
Source
“Please regularly backup the data to a server or other mobile storage devices.”